Model MKF 720 | Dynamic climate chambers for rapid temperature changes with humidity control
Benefits
- Homogeneous climate conditions thanks to APT.line™ technology
- Automatic water and wastewater management
- Pressure humidification with fast response times
- Comprehensive programming and data acquisition
- Large heated viewing window
Product description
A BINDER MKF series environmental simulation chamber is ideally suited for any cold and heat testing based on current temperature and climate testing methods according to DIN and IEC standards. The comprehensive standard equipment for this environmental simulation chamber ensures ease of operation.
Important characteristics
- Temperature range: -40 °C to +180 °C
- Humidity range: 10 % to 98 % RH
- Integrated water-storage tank, 20 liters
- 4 zero-voltage relay contacts
- APT.line™ preheating chamber technology
- Programmable condensation protection for test material
- Heated viewing window with LED interior lighting
- Humidity regulation with capacitative humidity sensor and vapor humidification
- BINDER Multi Management Software APT-COM™ 4 Basic Edition
- Troubleshooting system with visual and audible alarms
- Intuitive touchscreen controller with time-segment and real-time programming
- Internal data logger, measured values can be read out in open format via USB
- 2 access ports with silicone plugs: 80 mm, left and right
- 4 castors, two with brakes
- Computer interface: Ethernet
- Adjustable ramp function
- Integrated chart recorder
- Real-time clock
- Door heating
- Alarm notification in the event of insufficient water in fresh water tank
- 1 stainless steel rack
- Complete safety connection kit for water supply and drainage, up to 1 m height
- Inner chamber made of stainless steel
- CFC-free refrigerant R-452A
- Cooling with compressor cooling unit
Technical specifications
Data | ||
Label | MKF720-400V | MKF720-480V-C |
Article Number | 9020-0381 | 9020-0359 |
Option model | Standard | Standard |
NEU Leistungsdaten Temperatur | ||
Temperature range | -40...180 °C | -40...180 °C |
Temperature uniformity dependent on set value | 0.1...1.8 ± K | 0.1...1.8 ± K |
Temperature fluctuation dependent on set value | 0.1...0.5 ± K | 0.1...0.5 ± K |
Average heating-up rate according to IEC 60068-3-5 | 4.8 K/min | 4.8 K/min |
Cooling down time from 180°C to -40°C | 120 min | 120 min |
Average cooling down time according to IEC 60068-3-5 | 4.8 K/min | 4.8 K/min |
Performance Data Climate | ||
Temperature range with humidity | 10...95 °C | 10...95 °C |
Humidity range | 10...98 % r.F. | 10...98 % r.F. |
Temperature fluctuation with humidity dependent on set value | 0.2...1.5 ± K | 0.2...1.5 ± K |
Humidity fluctuation with humidity dependent on set value | ≤2,5 ± % r.F. | ≤2,5 ± % r.F. |
Max. heat compensation at 25°C and 90% RH | 1,000 W | 1,000 W |
Electrical data | ||
Nominal Voltage | 400 V | 480 V |
Power frequency | 50 Hz | 60 Hz |
Nominal power | 11 kW | 11 kW |
Unit fuse | 32,0 A | 32,0 A |
Phase (Nominal voltage) | 3~ | 3~ |
Maße und Gewichte | ||
Interior volume | 734 L | 734 L |
Net weight of the unit (empty) | 590 kg | 590 kg |
Load per rack | 40 kg | 40 kg |
permitted load | 160 kg | 160 kg |
Wall clearance back | 300 mm | 300 mm |
Wall clearance sidewise | 200 mm | 200 mm |
Housing dimensions not incl. fittings and connections | ||
Width net | 1,580 mm | 1,580 mm |
Height net | 2,005 mm | 2,005 mm |
Depth net | 1,140 mm | 1,140 mm |
Internal Dimensions | ||
Interior width | 1,200 mm | 1,200 mm |
Interior height | 1,020 mm | 1,020 mm |
Interior depth | 600 mm | 600 mm |
Unit doors | 1 | 1 |
Environment-specific data | ||
Sound-pressure level | 65 dB(A) | 67 dB(A) |
Energy consumption at 25°C and 60% RH | 3,000 Wh/h | 3,000 Wh/h |
Fixtures | ||
Number of shelves (std./max.) | 1/11 | 1/11 |