Model CB 56 | CO2 incubators with hot air sterilization and heat sterilizable CO2 sensor
Benefits
- Safe – cell cultures receive maximum protection against contamination
- Reliable – constant well-being conditions for the cells
- Smart – simple routine cleaning and convenient operation
- Economical – efficient operation without consumables
Product description
The BINDER CB series CO₂ incubator is the premium class among the CO₂ incubators. It is suitable for all sensitive incubation applications and ensures optimal cell growth. A BINDER CB series CO₂ incubator can perform complex cultivation experiments as well as specific enviromental conditions.
Important characteristics
- Optimized humidification system with condensation protection
- Tried-and-tested anti-contamination concept with 180°C hot air sterilization
- Hot-air sterilizable CO₂ sensor with single-beam infrared technology
- Seamless inner chamber made of stainless steel with flanges as shelf support system
- Fanless interior with Venturi CO₂ gas mixing nozzle
- Intuitive touchscreen controller
- Internal data recording, can be read out in open format via USB interface
- Fault diagnosis system with visual and acoustic alarm
- Interfaces: Ethernet, USB, zero-voltage alarm contact
- Fail-safe CO₂ system for protecting the pH of cell cultures
Technical specifications
| Data | ||
| Designation | CB056-230V | CB056UL-120V |
| Article Number | 9640-0005 | 9640-0007 |
| Option model | Standard | Standard |
| Performance Data Temperature | ||
| Temperature range | +4 °C above ambient temperature to 50 °C | +4 °C above ambient temperature to 50 °C |
| Temperature uniformity at 37°C | 0.3 ± K | 0.3 ± K |
| Temperature fluctuation at 37°C | 0.1 ± K | 0.1 ± K |
| Recovery time after door was opened for 30 s at 37°C | 5 min | 5 min |
| Performance Data Climate | ||
| Humidity range | 90 ...95 % r.F. | 90 ...95 % r.F. |
| Gas Data | ||
| CO₂ range | 0 ...20 Vol.-% CO₂ | 0 ...20 Vol.-% CO₂ |
| CO₂ measuring technology | IR | IR |
| CO₂ recovery time after door was opened for 30 s at 5 vol. % CO₂ | 5 min | 5 min |
| Electrical data | ||
| Rated Voltage | 200...230 V | 100...120 V |
| Power frequency | 50/60 Hz | 50/60 Hz |
| Nominal power | 0,8 kW | 0,8 kW |
| Unit fuse | 10,0 A | 16,0 A |
| Phase (Nominal voltage) | 1~ | 1~ |
| Dimensions and weights | ||
| Interior volume | 2 cu.ft. | 2 cu.ft. |
| Net weight of the unit (empty) | 117 lbs | 117 lbs |
| Load per rack | 22 lbs | 22 lbs |
| Permitted load | 66 lbs | 66 lbs |
| Wall clearance back | 3,94 in | 3,94 in |
| Wall clearance sidewise | 1,97 in | 1,97 in |
| Housing dimensions not incl. fittings and connections | ||
| Width net | 20.79 in | 20.79 in |
| Height net | 26.54 in | 26.54 in |
| Depth net | 21.46 in | 21.46 in |
| Internal Dimensions | ||
| Interior width | 15.75 in | 15.75 in |
| Interior height | 15.75 in | 15.75 in |
| Interior depth | 12.99 in | 12.99 in |
| Inner doors | 1 | 1 |
| Unit doors | 1 | 1 |
| Environment-specific data | ||
| Sound-pressure level | 41 dB(A) | 41 dB(A) |
| Energy consumption at 37°C | 45 Wh/h | 45 Wh/h |
| Fixtures | ||
| Number of shelves (std./max.) | 2/3 | 2/3 |
Downloads
Options, Accessories, Services
Sorry, no result could be found. Check the spelling or use a new search term.
Stacking adapter - for the thermally isolated stacking of two BINDER CO2 incubators
Base - the base equipped with casters is used for the secure positioning and leveling of a BINDER CO2 incubator
Shelf, perforated - stainless steel, for divided inner door
Shelf, perforated - Stainless steel
APT-COM™ 4 BASIC-Edition - for simple logging and documentation requirements with up to 5 networked units.
APT-COM™ 4 PROFESSIONAL-Edition - convenient unit and user management built on the BASIC edition. Suitable for networking up to 100 units.
APT-COM™ 4 GLP-Edition - for working under GLP-compliant conditions. Measured values are documented in a tamper-proof way in line with the requirements of FDA Regulation 21 CFR 11.
pH-neutral detergent - concentrated, for gentle remove of residual contaminants; 1 kg
Gas tank changer - external, BINDER Gas Supply Service, for connecting 2 gas tanks (CO2, N2 or O2), with audible and visual alarms, as well as zero-voltage alarm output
Sorry, no result could be found. Check the spelling or use a new search term.
Inner door, divided - Cell therapy compartmentalization, consisting of 4 inner doors and two shelf levels with one partition wall each
Analog output 4-20 mA - for temperature and CO2 values (outputs not adjustable)
Access port with silicone plug - 30 mm, back
Access port with silicone plug - 30 mm, left
Access port with silicone plug - 30 mm, right
Electric access port - 8-pin, for low voltage with LEMO socket (coverable) and LEMO plug (max. 24 V – 2 A)
Sorry, no result could be found. Check the spelling or use a new search term.
BRONZE 3-year maintenance contract - Annual Device Inspection Includes Preventive Maintenance with Calibration Certificate. 10% Discount on Spare Parts and Labor for any additional Services needed Wear + tear parts to be purchased or billed to customer. Travel to be quoted as separate.
SILVER 3-year maintenance contract - Annual Device Inspection Includes Preventive Maintenance with Calibration Certificate 15% Discount on Spare Parts and Labor for any additional Services needed One additional courtesy visit per year can be used for service visit or operator training. Wear + tear parts to be purchased or billed to customer. Travel to be quoted separate.
GOLD 3-year maintenance contract - Annual Device Inspection Includes Preventive Maintenance with Calibration Certificate of Temperature. 25% Discount on Spare Parts and Labor for any additional Services needed. One additional courtesy visit per year can be used for service visit or operator. Wear + tear parts to be purchased or billed to customer. Travel to be quoted separate.
Unit installation - Connect the unit to the customer-side connections (electricity, water, wastewater, gas), basic functions check, brief operating instructions. (excl.: unpacking, setup, controller instructions, programming, installation work) Travel to be quoted as separate.
Unit instructions - Instruction regarding operating principle and basic functions of the unit, operation of the control electronics including programming. Travel to be quoted as separate Limited to 3 hours.
Preventive maintenance - One-off maintenance service in accordance with maintenance schedule. Visual inspection of mechanical and electrical components, testing of all key functions. Checking a test temperature specified by the user in center of usable space without certificate. Wear + tear parts need to be purchased or billed to customer. Travel to be quoted as separate.
CO₂ calibration - CO₂ calibration at a specified value, measurements are taken with analyzed test gas at 5%, including certificate. Travel to be quoted separate.
Temperature calibration - Calibration of one (1) test temperature specified by the user in center of chamber, including certificate Travel to be quoted separate.
Temperature calibration - Extension of calibration of one (1) additional test temperature specified by the user in the center of the usable space, including certificate Travel to be quoted separate.
Temperature measurement, 9 measuring points - Temperature measurement with 9 measuring points with a set value specified by the user, including certificate. Travel to be quoted separate. A total of 3 shelvesare required.
Temperature measurement, 18 measuring points - Temperature measurement with 18 measuring points with a set value specified by the user, including certificate. Travel to be quoted separate. A total of 2 shelvesare required.
Temperature measurement, 27 measuring points - Temperature measurement with 27 measuring points with a set value specified by the user, including certificate. Travel to be quoted separate. A total of 3 shelvesare required.
Execution of IQ/OQ - Execution of IQ/OQ in accordance with qualification folder. Travel to be quoted separate.
Execution of IQ/OQ/PQ - Execution of IQ/OQ/PQ in accordance with qualification folder. Travel to be quoted separate.
1-year warranty extension - The warranty is extended by 1 year from the delivery date, wear parts are excluded