Model MKF 56 | Dynamic climate chambers for rapid temperature changes with humidity control
Benefits
- Homogeneous climate conditions thanks to APT.line™ technology
- Automatic water and wastewater management
- Pressure humidification with fast response times
- Comprehensive programming and data acquisition
- Large heated viewing window
Product description
A BINDER MKF series environmental simulation chamber is ideally suited for any cold and heat testing based on current temperature and climate testing methods according to DIN and IEC standards. The comprehensive standard equipment for this environmental simulation chamber ensures ease of operation.
Important characteristics
- Temperature range: -40 °C to +180 °C
- Humidity range: 10 % to 98 % RH
- 4 zero-voltage relay contacts that can be activated via MCS controller
- APT.line™ preheating chamber technology
- Programmable condensation protection for test material
- Heated viewing window with LED interior lighting
- Humidity regulation with capacitative humidity sensor and vapor humidification
- BINDER Multi Management Software APT-COM™ Basic Edition
- Troubleshooting system with visual and audible alarms
- Intuitive touchscreen controller with time-segment and real-time programming
- Internal data logger, measured values can be read out in open format via USB
- Unit self-test for comprehensive status analysis
- Access port with silicone plug: 50 mm, left
- 4 castors, two with brakes
- Computer interface: Ethernet
- Adjustable ramp function
- Integrated chart recorder
- Real-time clock
- Door heating
- 1 stainless steel rack
- Complete safety connection kit for water supply and drainage, up to 1 m height
- BINDER test confirmation
- Inner chamber made of stainless steel
- Pt 100 temperature sensor
- Thermal insulation using PUR foam
- CFC-free refrigerant R-452A
- Cooling with compressor cooling unit
- Fin evaporator
Technical specifications
Data | ||
Label | MKF056-230V | MKF056-240V |
Article Number | 9020-0378 | 9020-0389 |
Option model | Standard | Standard |
NEU Leistungsdaten Temperatur | ||
Temperature range | -40...180 °C | -40...180 °C |
Temperature uniformity dependent on set value | 0.5...1.5 ± K | 0.5...1.5 ± K |
Temperature fluctuation dependent on set value | 0.1...0.5 ± K | 0.1...0.5 ± K |
Average heating-up rate according to IEC 60068-3-5 | 5.0 K/min | 5.0 K/min |
Cooling down time from 180°C to -40°C | 90 min | 90 min |
Average cooling down time according to IEC 60068-3-5 | 5.0 K/min | 5.0 K/min |
Performance Data Climate | ||
Temperature range with humidity | 10...95 °C | 10...95 °C |
Humidity range | 10...98 % RH | 10...98 % RH |
Temperature fluctuation with humidity dependent on set value | 0.1...0.5 ± K | 0.1...0.5 ± K |
Humidity fluctuation with humidity dependent on set value | ≤2,5 ± % r.F. | ≤2,5 ± % r.F. |
Max. heat compensation at 25°C and 90% RH | 250 W | 250 W |
Electrical data | ||
Rated Voltage | 200...230 V | 200...240 V |
Power frequency | 50 Hz | 60 Hz |
Nominal power | 2,8 kW | 2,8 kW |
Unit fuse | 16,0 A | 16,0 A |
Phase (Nominal voltage) | 1~ | 1~ |
Maße und Gewichte | ||
Interior volume | 60 L | 60 L |
Net weight of the unit (empty) | 175 kg | 175 kg |
Load per rack | 15 kg | 15 kg |
Permitted load | 60 kg | 60 kg |
Wall clearance back | 300 mm | 300 mm |
Wall clearance sidewise | 200 mm | 200 mm |
Housing dimensions not incl. fittings and connections | ||
Width net | 720 mm | 720 mm |
Height net | 1,445 mm | 1,445 mm |
Depth net | 780 mm | 780 mm |
Internal Dimensions | ||
Interior width | 400 mm | 400 mm |
Interior height | 420 mm | 420 mm |
Interior depth | 348 mm | 348 mm |
Unit doors | 1 | 1 |
Environment-specific data | ||
Sound-pressure level | 59 dB(A) | 59 dB(A) |
Energy consumption at 25°C and 60% RH | 800 Wh/h | 800 Wh/h |
Fixtures | ||
Number of shelves (std./max.) | 1/4 | 1/4 |