Model VDL 23 | Vacuum drying chambers for flammable solvents
Benefits
- Fast, gentle drying
- Optimum heat transfer through large thermal conducting plates
- Safe working thanks to ATEX-compliant safety concept
- Also option of modular complete system with vacuum pump and pump chamber
Product description
BINDER VDL series safety vacuum drying chambers ensure maximum safety when drying organic solvents. Explosion-protected unit with classification according to ATEX directive 2014/34/EU: EX II 2/3/- G IIB T3 Gb/Gc/- X.
Important characteristics
- Temperature range: +9°C above ambient temperature to +110 °C
- ATEX conformity for units: EX II 2/3/- G IIB T3 Gb/Gc/- X
- Intuitive touchscreen controller with graphical pressure and temperature display
- Program-controlled drying monitoring with automatic ventilation at end of process
- Internal data logger, measured values can be read out in open format via USB
- Pressure control device for heating activated from < 100 mbar
- 2 aluminum expansion racks, can be custom-positioned
- Universal connection for ventilation with ambient air or inert gas
- Universal access port DN 16 - for the introduction of external measuring devices
- Shatterproof, spring-mounted safety glass panel
- Large viewing window
- Computer interface: Ethernet
Technical specifications
| Data | ||
| Designation | VDL023-230V | VDL023-120V |
| Article Number | 9630-0009 | 9630-0013 |
| Option model | Standard | Standard |
| Performance Data Temperature | ||
| Temperature range | +9 °C above ambient temperature to 110 °C | +9 °C above ambient temperature to 110 °C |
| Temperature uniformity at 100°C | 1.0 ± K | 1.0 ± K |
| Temperature fluctuation at 100°C | 0.1 ± K | 0.1 ± K |
| Heating up time to 100°C | 110 min | 110 min |
| Vacuum Data | ||
| Leak rate | 0,01 bar/h | 0,01 bar/h |
| Pressure fluctuation over time at 50 mbar | 1 ± mbar | 1 ± mbar |
| Pressure fluctuation over time at 400 mbar | 70 ± mbar | 70 ± mbar |
| Electrical data | ||
| Rated Voltage | 200...230 V | 100...120 V |
| Power frequency | 50/60 Hz | 50/60 Hz |
| Nominal power | 0,9 kW | 0,9 kW |
| Unit fuse | 6,3 A | 10,0 A |
| Phase (Nominal voltage) | 1~ | 1~ |
| Connections | ||
| Vacuum connection with small flange | 16 DN mm | 16 DN mm |
| Measuring access port with small flange | 16 DN mm | 16 DN mm |
| Dimensions and weights | ||
| Interior volume | 24 L | 24 L |
| Net weight of the unit (empty) | 72 kg | 72 kg |
| Load per rack | 20 kg | 20 kg |
| Permitted load | 50 kg | 50 kg |
| Wall clearance back | 100 mm | 100 mm |
| Wall clearance sidewise | 70 mm | 70 mm |
| Housing dimensions not incl. fittings and connections | ||
| Width net | 523 mm | 523 mm |
| Height net | 698 mm | 698 mm |
| Depth net | 413 mm | 413 mm |
| Internal Dimensions | ||
| Interior width | 285 mm | 285 mm |
| Interior height | 285 mm | 285 mm |
| Interior depth | 295 mm | 295 mm |
| Unit doors | 1 | 1 |
| Environment-specific data | ||
| Sound-pressure level | 40 dB(A) | 40 dB(A) |
| Energy consumption at 100°C | 140 Wh/h | 140 Wh/h |
| Fixtures | ||
| Number of shelves (std./max.) | 2/4 | 2/4 |
Downloads
Options, Accessories, Services
Sorry, no result could be found. Check the spelling or use a new search term.
Expansion racks - aluminum
Expansion racks - stainless steel
Vacuum pump VP5 EX - ATEX chemical membrane pump with nominal pumping speed 1.9 m³/h, final pressure 12 mbar, 230V/50Hz.ATEX conformity: maximum permissible pressure at inlet and 1.1 suction chamber (pumped gases): II 2G Ex h IIC T3 Gb X;exterior with inert gas flushing: II 2G Ex h IIB T4 Gb X;exterior without inert gas flushing: II 3G Ex h IIB T4 Gc X motor: II 2G Ex db IIB T4 Gb
Pump chamber VDL 23 with vacuum pump VP5 EX - 230 V / 50-60 Hz option model, Pump chamber for installation of vacuum pumps. Including ATEX chemical membrane pump with nominal pumping speed 1.9 m³/h, final pressure 12 mbar, 230V/50Hz.ATEX conformity: suction chamber (pumped gases): II 2G Ex h IIC T3 Gb X; exterior with inert gas flushing: II 2G Ex h IIB T4 Gb X; exterior without inert gas flushing: II 3G Ex h IIB T4 Gc X motor: II 2G Ex db IIB T4 Gb
APT-COM™ 4 BASIC-Edition - for simple logging and documentation requirements with up to 5 networked units.
APT-COM™ 4 PROFESSIONAL-Edition - convenient unit and user management built on the BASIC edition. Suitable for networking up to 100 units.
APT-COM™ 4 GLP-Edition - for working under GLP-compliant conditions. Measured values are documented in a tamper-proof way in line with the requirements of FDA Regulation 21 CFR 11.
pH-neutral detergent - concentrated, for gentle remove of residual contaminants; 1 kg
Sorry, no result could be found. Check the spelling or use a new search term.
Expansion racks - stainless steel expansion rack instead of standard aluminum expansion rack, incl. calibration
Door gasket - made of FKM, silicone-free
Temperature measurement of the specimen - with flexible Pt 100 sensor and digital specimen temperature display
Door lock - lockable door handle
Additional universal ventilation - Additional universal connection for ventilation with ambient air or inert gas. Flow limiter thread (RP"), and connection with 3/8" male thread.
Analog output 4-20 mA - for temperature and pressure values (output not adjustable)
Sorry, no result could be found. Check the spelling or use a new search term.
BRONZE 3-year maintenance contract - Maintenance service as contractually agreed, visual inspection of mechanical and electrical components, check of control response, 20% discount on spare parts
Unit instructions - Instruction regarding operating principle and basic functions of the unit, operation of the control electronics including programming
Calibration certificate, temperature - Extension of calibration of one (1) additional test temperature specified by the user in the center of the usable space, including certificate
Temperature measurement, 9 measuring points - Temperature measurement with 9 measuring points with a set value specified by the user, including certificate
Execution of IQ/OQ/PQ - Execution of IQ/OQ/PQ in accordance with qualification folder
1-year warranty extension - The warranty is extended by 1 year from the delivery date, wear parts are excluded
Unit commissioning - Connect the unit to the customer-side connections (electricity, water, wastewater, gas), basic functions check, brief operating instructions. (excl.: unpacking, setup, controller instructions, programming, installation work)
Temperature measurement, 15 measuring points - Temperature measurement with 15 measuring points with a set value specified by the user, including certificate
Calibration digital pressure display - including certificate, measurement at specified pressure, range 20 – 900 mbar
Pressure calibration - Expansion – Digital pressure display calibration at a specified test pressure of 20–900 mbar, including certificate
Execution of IQ/OQ - Execution of IQ/OQ in accordance with qualification folder
VAP 5 vacuum pump maintenance - Maintenance service of the VAP 5 vacuum pump according to the maintenance schedule, visual inspection, replacement of the diaphragms, O-rings and seals, (Service Kit 8500-0160 included in the scope of services) as well as final functional test.
Maintenance - One-off maintenance service in accordance with maintenance schedule. Visual inspection of mechanical and electrical components, testing of all key functions. Calibration of a test temperature specified by the user in center of usable space without certificate
SILVER 3-year maintenance contract - Maintenance service as contractually agreed, visual inspection of mechanical and electrical components, check of control response, 20% discount on spare parts, testing of all key functions, calibration of one test temperature specified by the user in the center of the usable space, without certificate
GOLD 3-year maintenance contract - Maintenance service as contractually agreed, visual inspection of mechanical and electrical components, check of control response, 20% discount on spare parts, testing of all key functions, replacement of wear parts, calibration of one test temperature specified by the user in the center of the usable space, including certificate
Calibration certificate, temperature - Calibration of one (1) test temperature specified by the user in center of chamber, including certificate
Safety test - Testing of ATEX-specific safety systems: inert gas flushing and safety temperature shut-off. Testing of the specified systems also forms part of the more extensive maintenance service (DL20-0520)