Model MK 115 | Dynamic climate chambers for rapid temperature changes
Benefits
- Homogeneous climate conditions thanks to APT.line™ technology
- Comprehensive programming and data acquisition
- Large heated viewing window
Product description
The BINDER MK series chamber is suitable for all heat and cold tests between -40 °C and +180 °C. The MK is a clever alternative to complex individual solutions for cyclical temperature tests.
Important characteristics
- Temperature range: -40 °C to +180 °C
- APT.line™ preheating chamber technology
- Programmable condensation protection for test material
- Heated viewing window with LED interior lighting
- BINDER Multi Management Software APT-COM™ Basic Edition
- Troubleshooting system with visual and audible alarms
- Intuitive touchscreen controller with time-segment and real-time programming
- Internal data logger, measured values can be read out in open format via USB
- Access port with silicone plug, 50 mm, left
- Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
- 4 castors, two with brakes
- Computer interface: Ethernet
- 230 V power socket on the right-side control panel
- Adjustable ramp function
- Integrated chart recorder
- Real-time clock
- Door heating
- 1 stainless steel rack
- Inner chamber made of stainless steel
- CFC-free refrigerant R-452A
- Cooling with compressor cooling unit
Technical specifications
| Data | ||
| Designation | MK115-400V | MK115-480V-C |
| Article Number | 9020-0375 | 9020-0303 |
| Option model | Standard | with voltage and frequency converter |
| Performance Data Temperature | ||
| Temperature range | -40...180 °C | -40...180 °C |
| Temperature uniformity dependent on set value | 0.1...2.0 ± K | 0.1...2.0 ± K |
| Temperature fluctuation dependent on set value | 0.1...0.5 ± K | 0.1...0.5 ± K |
| Average heating-up rate according to IEC 60068-3-5 | 5.3 K/min | 5.3 K/min |
| Cooling down time from 180°C to -40°C | 90 min | 90 min |
| Average cooling down time according to IEC 60068-3-5 | 5.0 K/min | 5.0 K/min |
| Electrical data | ||
| Rated Voltage | 400 V | 480 V |
| Power frequency | 50 Hz | 60 Hz |
| Nominal power | 3,5 kW | 3,5 kW |
| Unit fuse | 16,0 A | 16,0 A |
| Phase (Nominal voltage) | 3~/N/PE | 3~ |
| Dimensions and weights | ||
| Interior volume | 115 L | 115 L |
| Net weight of the unit (empty) | 255 kg | 255 kg |
| Load per rack | 30 kg | 30 kg |
| Permitted load | 60 kg | 60 kg |
| Wall clearance back | 300 mm | 300 mm |
| Wall clearance sidewise | 200 mm | 200 mm |
| Housing dimensions not incl. fittings and connections | ||
| Width net | 980 mm | 980 mm |
| Height net | 1,725 mm | 1,725 mm |
| Depth net | 865 mm | 865 mm |
| Internal Dimensions | ||
| Interior width | 600 mm | 600 mm |
| Interior height | 480 mm | 480 mm |
| Interior depth | 400 mm | 400 mm |
| Unit doors | 1 | 1 |
| Environment-specific data | ||
| Sound-pressure level | 62 dB(A) | 67 dB(A) |
| Energy consumption at 20°C | 650 Wh/h | 650 Wh/h |
| Fixtures | ||
| Number of shelves (std./max.) | 1/4 | 1/4 |
Downloads
Options, Accessories, Services
Sorry, no result could be found. Check the spelling or use a new search term.
Rack - stainless steel
Shelf, perforated - Stainless steel
Rack, reinforced - stainless steel, with fasteners (1 set of 4)
APT-COM™ 4 PROFESSIONAL-Edition - convenient unit and user management built on the BASIC edition. Suitable for networking up to 100 units.
APT-COM™ 4 GLP-Edition - for working under GLP-compliant conditions. Measured values are documented in a tamper-proof way in line with the requirements of FDA Regulation 21 CFR 11.
pH-neutral detergent - concentrated, for gentle remove of residual contaminants; 1 kg
Sorry, no result could be found. Check the spelling or use a new search term.
Access port with silicone plug - 30 mm, right
Access port with silicone plug - 30 mm, left
Access port with silicone plug - 50 mm, right
Access port with silicone plug - 50 mm, left
Access port with silicone plug - 80 mm, right
Access port with silicone plug - 80 mm, left
Access port with silicone plug - 80 mm, top
Pt 100 temperature sensor - additional flexible Pt 100, interior, for displaying the temperature on the unit display
Access port with silicone plug - 125 mm, right
Access port with silicone plug - 125 mm, left
Access port with silicone plug - 100 mm, top
Access port with silicone plug - 125 mm, top
Access port with silicone plug - 100 mm, right
Access port with silicone plug - 100 mm, left
Door lock - lockable door handle
Dry-air purge - uncontrolled, incl. connection; for preventing condensation
Notch-type access port - notch-type access port in door, 100 x 35 mm
Sorry, no result could be found. Check the spelling or use a new search term.
BRONZE 3-year maintenance contract - Maintenance service as contractually agreed, visual inspection of mechanical and electrical components, check of control response, 20% discount on spare parts
Calibration certificate, temperature - Extension of calibration of one (1) additional test temperature specified by the user in the center of the usable space, including certificate
Temperature measurement, 9 measuring points - Temperature measurement with 9 measuring points with a set value specified by the user, including certificate
Temperature measurement, 18 measuring points - Temperature measurement with 18 measuring points with a set value specified by the user, including certificate
Temperature measurement, 27 measuring points - Temperature measurement with 27 measuring points with a set value specified by the user, including certificate
Execution of IQ/OQ - Execution of IQ/OQ in accordance with qualification folder
Execution of IQ/OQ/PQ - Execution of IQ/OQ/PQ in accordance with qualification folder
Unit commissioning - Connect the unit to the customer-side connections (electricity, water, wastewater, gas), basic functions check, brief operating instructions. (excl.: unpacking, setup, controller instructions, programming, installation work)
Unit instructions - Instruction regarding operating principle and basic functions of the unit, operation of the control electronics including programming
Leakage test - Testing and logging of the leak-tightness of systems containing hydrofluorocarbons in accordance with Regulation EC842/2006, incl. test book
Maintenance - One-off maintenance service in accordance with maintenance schedule. Visual inspection of mechanical and electrical components, testing of all key functions. Calibration of a test temperature specified by the user in center of usable space without certificate
SILVER 3-year maintenance contract - Maintenance service as contractually agreed, visual inspection of mechanical and electrical components, check of control response, 20% discount on spare parts, testing of all key functions, calibration of one test temperature specified by the user in the center of the usable space, without certificate
GOLD 3-year maintenance contract - Maintenance service as contractually agreed, visual inspection of mechanical and electrical components, check of control response, 20% discount on spare parts, testing of all key functions, replacement of wear parts, calibration of one temperature/humidity/pressure value, including certificate
Calibration certificate, temperature - Calibration of one (1) test temperature specified by the user in center of chamber, including certificate
1-year warranty extension - The warranty is extended by 1 year from the delivery date, wear parts are excluded