CB056-CO2-Incubator-GF_Web_670x515.jpg
CB056-CO2-Incubator-OF_Web_670x515.jpg
Abmessungen-CB-056-9040-0168_Web_670x515.jpg
CO2-incubator-Stapel-100-85_Web_670x515.jpg
Product image
Product image
Product image
Product image

Model CB-56 | CO₂ incubator with hot air sterilization and heat sterilizable CO₂ sensor

  • Safe – cell cultures receive maximum protection against contamination
  • Reliable – constant well-being conditions for the cells
  • Smart – simple routine cleaning and convenient operation
  • Economical – efficient operation without consumables

Configure your product with options and accessories and request it directly here.

  • Optimized humidification system
    with condensation protection
  • Tried-and-tested anti-contamination concept
    with 180 °C hot air sterilization
  • Hot-air sterilizable CO₂ sensor
    with single-beam infrared technology
  • Seamless inner chamber made of stainless steel
    with flanges as shelf support system
  • Fanless interior
    with Venturi CO₂ gas mixing nozzle
  • Intuitive touchscreen controller
  • Internal data recording
    can be read out in open format via USB interface
  • Fault diagnosis system
    with visual and acoustic alarm
  • Interfaces:
    Ethernet, USB, zero-voltage alarm contact
  • Fail-safe CO₂ system
    for protecting the pH of cell cultures

Technical specifications

DesignationCB056-230V¹CB056-230V-O¹CB056UL-120V¹CB056UL-120V-O¹
Option modelStandardwith O₂ control 0.2–20%Standardwith O₂ control 0.2–20%
Order number9640-00059640-00069640-00079640-0008
Data
SerieCBCBCBCB
Performance Data Temperature
Recovery time after 30 seconds door open at 37 °C [min]5555
Temperature fluctuation at 37 °C [± K]0.10.10.10.1
Temperature range +4 °C above ambient temperature to [°C]60-60-
Temperature range +6 °C above ambient temperature to-60 °C-60 °C
Temperature variation at 37 °C [± K]0.30.30.30.3
Performance Data Climate
Humidity range90…9590…9590…9590…95
CO₂-Data
CO₂ measurementIRIRIRIR
CO₂ range [Vol.-% CO₂]0…200…200…200…20
CO₂ Recovery time after 30 seconds door open at 5 Vol.-% CO₂ [min]5555
O₂-Data
O₂ control ranges with option: O₂ range [Vol.-% O₂]-10…95-10…95
O₂ Recovery time after 30 seconds door open at 10 Vol.-% O₂ [min]-5-5
O₂ Recovery time after 30 seconds door open at 5 Vol.-% O₂ [min]-5-5
Standard O₂ control ranges: O₂ range [Vol.-% O₂]-0.2…20-0.2…20
Electrical data
Nominal power [kW]0.80.80.80.8
Phase (Nominal voltage)1~1~1~1~
Power frequency [Hz]50/6050/6050/6050/60
Rated Voltage [V]200…230200…230100…120100…120
Unit fuse [A]10101616
Doors
Inner doors1111
Unit doors1111
Housing dimensions not incl. fittings and connections
Depth net [mm]545545545545
Height net [mm]674674674674
Width net [mm]528528528528
Internal Dimensions
Depth [mm]330330330330
Height [mm]400400400400
Width [mm]400400400400
Measures
Interior volume [L]53535353
Load per rack [kg]10101010
Net weight of the unit (empty) [kg]53535353
Permitted load [kg]30303030
Wall clearance back [mm]100100100100
Wall clearance sidewise [mm]50505050
Environment-specific data
Energy consumption at 37 °C [Wh/h]45454545
Sound-pressure level [dB(A)]41414141
Fixtures
Number of shelves (std./max.)2/32/32/32/3
Performance Data Temperature
Recovery time after 30 seconds door open at 37 °C [min]5555
Temperature fluctuation at 37 °C [± K]0.10.10.10.1
Temperature range +4 °C above ambient temperature to [°C]60-60-
Temperature range +6 °C above ambient temperature to-60 °C-60 °C
Temperature variation at 37 °C [± K]0.30.30.30.3
Performance Data Climate
Humidity range90…9590…9590…9590…95
CO₂-Data
CO₂ measurementIRIRIRIR
CO₂ range [Vol.-% CO₂]0…200…200…200…20
CO₂ Recovery time after 30 seconds door open at 5 Vol.-% CO₂ [min]5555
O₂-Data
O₂ control ranges with option: O₂ range [Vol.-% O₂]-10…95-10…95
O₂ Recovery time after 30 seconds door open at 10 Vol.-% O₂ [min]-5-5
O₂ Recovery time after 30 seconds door open at 5 Vol.-% O₂ [min]-5-5
Standard O₂ control ranges: O₂ range [Vol.-% O₂]-0.2…20-0.2…20
Electrical data
Nominal power [kW]0.80.80.80.8
Phase (Nominal voltage)1~1~1~1~
Power frequency [Hz]50/6050/6050/6050/60
Rated Voltage [V]200…230200…230100…120100…120
Unit fuse [A]10101616
Doors
Inner doors1111
Unit doors1111
Housing dimensions not incl. fittings and connections
Depth net [in]21.4721.4721.4721.47
Height net [in]26.5626.5626.5626.56
Width net [in]20.820.820.820.8
Internal Dimensions
Depth [in]13131313
Height [in]15.7615.7615.7615.76
Width [in]15.7615.7615.7615.76
Measures
Interior volume [cu.ft.]1.91.91.91.9
Load per rack [lbs]22222222
Net weight of the unit (empty) [lbs]117117117117
Permitted load [lbs]66666666
Wall clearance back [in]3.943.943.943.94
Wall clearance sidewise [in]1.971.971.971.97
Environment-specific data
Energy consumption at 37 °C [Wh/h]45454545
Sound-pressure level [dB(A)]41414141
Fixtures
Number of shelves (std./max.)2/32/32/32/3

Downloads

Data Sheets CB 56

pdf

Operating ManualsTubular-Bag Sets | Humidification Water Set | Water Container Set for CB series with active humidification

pdf

Technical DrawingsVersion CB056-120V O₂-R. 0-20% 9040-0170

pdf

Technical DrawingsVersion Modellvariante CB056-120V 9040-0169

pdf

Technical DrawingsVersion CB056-230V O₂-R. 0-20% 9040-0168

pdf

Technical DrawingsVersion CB056-230V 9040-0167

pdf

White paper"Good laboratory practice" What is behind all this?

pdf

White paperGood Laboratory Practice Your checklist for GLP-compliant research and development

pdf

White paperCell cultivation: for and against particular technical solutions in CO₂ incubators The parameters to consider

pdf

White paperCell cultivation without contamination Hot air sterilization and other means of contamination control in CO₂ incubators

pdf

White paperValidation and Qualification In the regulated environment

pdf

BrochureBINDER INDIVIDUAL Series production devices for special requirements

pdf

Brochure CO₂ incubators CB-S | CBF | CB series

pdf