Model BF 115 | Standard-Incubators with forced convection
- Excellent temporal and spatial temperature accuracy
- High process reliability
- 100 °C disinfection routine
- Temperature range:
+8 °C above ambient temperature to +100 °C - High temperature accuracy
thanks to APT.line™ technology - Forced convection
- Controller
with LCD display - Electromechanical control
of the exhaust air flap - Inner door
made of tempered safety glass - 2 chrome-plated racks
- Stackable
- Class 3.1 integrated independent temperature safety device
(DIN 12880) with visual alarm - Ergonomic handle design
- USB port
for recording data
Technical specifications
Designation | BF115-230V¹ | BF115UL-120V¹ |
---|---|---|
Option model | Standard | Standard |
Order number | 9010-0315 | 9010-0316 |
Data | ||
Serie | BF | BF |
Performance Data Temperature | ||
Heating-up time to 37 °C [min] | 8 | 8 |
Recovery time after 30 seconds door open at 37 °C [min] | 3 | 3 |
Temperature fluctuation at 37 °C [± K] | 0.1 | 0.1 |
Temperature range +8 °C above ambient temperature to [°C] | 100 | 100 |
Temperature variation at 37 °C [± K] | 0.3 | 0.3 |
Electrical data | ||
Nominal power [kW] | 0.4 | 0.4 |
Phase (Nominal voltage) | 1~ | 1~ |
Power frequency [Hz] | 50/60 | 60 |
Rated Voltage [V] | 230 | 120 |
Unit fuse [A] | 6.3 | 12.5 |
Doors | ||
Unit doors | 1 | 1 |
Housing dimensions not incl. fittings and connections | ||
Depth net [mm] | 605 | 605 |
Height net [mm] | 725 | 725 |
Width net [mm] | 710 | 710 |
Internal Dimensions | ||
Depth [mm] | 380 | 380 |
Height [mm] | 550 | 550 |
Width [mm] | 550 | 550 |
Measures | ||
Interior volume [L] | 114 | 114 |
Load per rack [kg] | 30 | 30 |
Net weight of the unit (empty) [kg] | 56 | 56 |
Permitted load [kg] | 150 | 150 |
Wall clearance back [mm] | 160 | 160 |
Wall clearance sidewise [mm] | 100 | 100 |
Environment-specific data | ||
Energy consumption at 37 °C [Wh/h] | 60 | 60 |
Sound-pressure level [dB(A)] | 43 | 43 |
Fixtures | ||
Number of shelves (std./max.) | 2/5 | 2/5 |
Performance Data Temperature | ||
Heating-up time to 37 °C [min] | 8 | 8 |
Recovery time after 30 seconds door open at 37 °C [min] | 3 | 3 |
Temperature fluctuation at 37 °C [± K] | 0.1 | 0.1 |
Temperature range +8 °C above ambient temperature to [°C] | 100 | 100 |
Temperature variation at 37 °C [± K] | 0.3 | 0.3 |
Electrical data | ||
Nominal power [kW] | 0.4 | 0.4 |
Phase (Nominal voltage) | 1~ | 1~ |
Power frequency [Hz] | 50/60 | 60 |
Rated Voltage [V] | 230 | 120 |
Unit fuse [A] | 6.3 | 12.5 |
Doors | ||
Unit doors | 1 | 1 |
Housing dimensions not incl. fittings and connections | ||
Depth net [in] | 23.84 | 23.84 |
Height net [in] | 28.57 | 28.57 |
Width net [in] | 27.97 | 27.97 |
Internal Dimensions | ||
Depth [in] | 14.97 | 14.97 |
Height [in] | 21.67 | 21.67 |
Width [in] | 21.67 | 21.67 |
Measures | ||
Interior volume [cu.ft.] | 4 | 4 |
Load per rack [lbs] | 66 | 66 |
Net weight of the unit (empty) [lbs] | 123 | 123 |
Permitted load [lbs] | 331 | 331 |
Wall clearance back [in] | 6.3 | 6.3 |
Wall clearance sidewise [in] | 3.94 | 3.94 |
Environment-specific data | ||
Energy consumption at 37 °C [Wh/h] | 60 | 60 |
Sound-pressure level [dB(A)] | 43 | 43 |
Fixtures | ||
Number of shelves (std./max.) | 2/5 | 2/5 |