Model BD 720 | Standard-Incubators with natural convection
- Excellent temporal and spatial temperature accuracy
- High process reliability
- 100 °C disinfection routine
- Temperature range:
+5 °C above ambient temperature to +100 °C - High temperature accuracy
thanks to APT.line™ technology - Natural convection
- Controller
with LCD display - Electromechanical control
of the exhaust air flap - Inner door
made of tempered safety glass - 2 chrome-plated racks
- Class 3.1 integrated
independent temperature safety device (DIN 12880) with visual alarm - Ergonomic handle design
- USB port
for recording data
Technical specifications
Designation | BD720-230V¹ | BD720UL-240V¹ |
---|---|---|
Option model | Standard | Standard |
Order number | 9010-0331 | 9010-0332 |
Data | ||
Serie | BD | BD |
Performance Data Temperature | ||
Heating-up time to 37 °C [min] | 70 | 70 |
Recovery time after 30 seconds door open at 37 °C [min] | 23 | 23 |
Temperature fluctuation at 37 °C [± K] | 0.1 | 0.1 |
Temperature range +5 °C above ambient temperature to [°C] | 100 | 100 |
Temperature variation at 37 °C [± K] | 0.7 | 0.7 |
Electrical data | ||
Nominal power [kW] | 1.65 | 1.75 |
Phase (Nominal voltage) | 1~ | 1~ |
Power frequency [Hz] | 50/60 | 50/60 |
Rated Voltage [V] | 230 | 240 |
Unit fuse [A] | 16 | 16 |
Doors | ||
Inner doors | 2 | 2 |
Unit doors | 2 | 2 |
Housing dimensions not incl. fittings and connections | ||
Depth net [mm] | 870 | 870 |
Height net [mm] | 1590 | 1590 |
Width net [mm] | 1165 | 1165 |
Internal Dimensions | ||
Depth [mm] | 600 | 600 |
Height [mm] | 1280 | 1280 |
Width [mm] | 960 | 960 |
Measures | ||
Interior volume [L] | 737 | 737 |
Load per rack [kg] | 45 | 45 |
Net weight of the unit (empty) [kg] | 170 | 170 |
Permitted load [kg] | 315 | 315 |
Wall clearance back [mm] | 160 | 160 |
Wall clearance sidewise [mm] | 100 | 100 |
Environment-specific data | ||
Energy consumption at 37 °C [Wh/h] | 78 | 78 |
Fixtures | ||
Number of shelves (std./max.) | 2/16 | 2/16 |
Performance Data Temperature | ||
Heating-up time to 37 °C [min] | 70 | 70 |
Recovery time after 30 seconds door open at 37 °C [min] | 23 | 23 |
Temperature fluctuation at 37 °C [± K] | 0.1 | 0.1 |
Temperature range +5 °C above ambient temperature to [°C] | 100 | 100 |
Temperature variation at 37 °C [± K] | 0.7 | 0.7 |
Electrical data | ||
Nominal power [kW] | 1.65 | 1.75 |
Phase (Nominal voltage) | 1~ | 1~ |
Power frequency [Hz] | 50/60 | 50/60 |
Rated Voltage [V] | 230 | 240 |
Unit fuse [A] | 16 | 16 |
Doors | ||
Inner doors | 2 | 2 |
Unit doors | 2 | 2 |
Housing dimensions not incl. fittings and connections | ||
Depth net [in] | 34.28 | 34.28 |
Height net [in] | 62.6 | 62.6 |
Width net [in] | 45.9 | 45.9 |
Internal Dimensions | ||
Depth [in] | 23.64 | 23.64 |
Height [in] | 50.4 | 50.4 |
Width [in] | 37.82 | 37.82 |
Measures | ||
Interior volume [cu.ft.] | 26 | 26 |
Load per rack [lbs] | 99 | 99 |
Net weight of the unit (empty) [lbs] | 375 | 375 |
Permitted load [lbs] | 694 | 694 |
Wall clearance back [in] | 6.3 | 6.3 |
Wall clearance sidewise [in] | 3.94 | 3.94 |
Environment-specific data | ||
Energy consumption at 37 °C [Wh/h] | 78 | 78 |
Fixtures | ||
Number of shelves (std./max.) | 2/16 | 2/16 |