Model KBF-S ECO 400 | Humidity test chambers with environmentally friendly thermoelectric cooling
Benefits
- Safe thanks to standard-compliant testing according to the ICH Q1A guideline, even with full load.
- Reliable thanks to failsafe operation without compromise.
- Smart, as a wide range of accessories makes it highly compatible for adaptation to specific customer requirements.
- Economical, as energy consumption is minimal thanks to optimized thermoelectric cooling.
Product description
Thanks to thermoelectric cooling technology with patented heat dissipation, the new KBF-S ECO climatic chamber is one of the most energy-efficient constant climate chambers on the market. The thermoelectric cooling technology means the KBF-S ECO series is also very quiet during operation.
Important characteristics
- Temperature range: 0 °C to +70 °C (max. 26 °C under ambient temperature)
- Humidity range: 10% to 80% RH
- APT.line™ preheating chamber technology
- Independent water supply via tank
- LCD to display temperature and humidity along with additional information and alarms
- Internal data logger, measured values can be read out in open format via USB
- Unit self-test for comprehensive status analysis
- Inner chamber made completely of stainless steel
- 2 stainless steel racks, including shelf carrier
- Access port with silicone plug Ø 30 mm, left
- Computer interface: Ethernet
Technical specifications
Data | ||
Label | KBFSECO400-230V | KBFSECO400UL-120V |
Article Number | 9020-0463 | 9020-0464 |
Option model | Standard | Standard |
NEU Leistungsdaten Temperatur | ||
Temperature range | (max. 26 °C under ambient temperature) 0...70 °C | (max. 26 °C under ambient temperature) 0...70 °C |
Max. heat compensation at 40°C | 300 W | 300 W |
Performance Data Climate | ||
Temperature range with humidity | (max. 26 °C under ambient temperature) 5...70 °C | (max. 26 °C under ambient temperature) 5...70 °C |
Humidity range | 10...80 % r.F. | 10...80 % r.F. |
Temperature uniformity at 25°C and 60% RH | 0.1 ± K | 0.1 ± K |
Temperature uniformity at 40°C and 75% RH | 0.5 ± K | 0.5 ± K |
Temperature fluctuation at 25°C and 60% RH | 0.1 ± K | 0.1 ± K |
Temperature fluctuation at 40°C and 75% RH | 0.1 ± K | 0.1 ± K |
Recovery time after door was opened for 30 s at 25°C and 60% RH | 6 min | 6 min |
Recovery time after door was opened for 30 s at 40°C and 75% RH | 15 min | 15 min |
Electrical data | ||
Nominal Voltage | 200...230 V | 100...120 V |
Power frequency | 50/60 Hz | 50/60 Hz |
Nominal power | 1,1 kW | 1,1 kW |
Unit fuse | 16,0 A | 16,0 A |
Phase (Nominal voltage) | 1~ | 1~ |
Maße und Gewichte | ||
Interior volume | 400 L | 400 L |
Net weight of the unit (empty) | 192 kg | 192 kg |
Load per rack | 30 kg | 30 kg |
permitted load | 120 kg | 120 kg |
Wall clearance back | 100 mm | 100 mm |
Wall clearance sidewise | 180 mm | 180 mm |
Housing dimensions not incl. fittings and connections | ||
Width net | 925 mm | 925 mm |
Height net | 1,946 mm | 1,946 mm |
Depth net | 796 mm | 796 mm |
Internal Dimensions | ||
Interior width | 650 mm | 650 mm |
Interior height | 1,268 mm | 1,268 mm |
Interior depth | 485 mm | 485 mm |
Unit doors | 1 | 1 |
Environment-specific data | ||
Sound-pressure level | 45 dB(A) | 45 dB(A) |
Energy consumption at 25°C and 60% RH | 60 Wh/h | 60 Wh/h |
Energy consumption at 40°C and 75% RH | 70 Wh/h | 70 Wh/h |
Fixtures | ||
Number of shelves (std./max.) | 2/15 | 2/15 |