Model FED 720 | Drying and heating chambers with forced convection and enhanced timer functions
Benefits
- Best energy efficiency on the market
- Excellent temporal and spatial temperature accuracy
- USB connection for recording data
- Temperature range up to 300 °C
Product description
BINDER heating chambers of the FED Avantgarde.Line series provide almost unlimited capacity and are particularly adaptable. With advanced timing functions and a controllable fan, temperature and convection conditions are easily controlled.
Important characteristics
- Temperature range: +10°C above ambient temperature to +300 °C
- Up to 30% lower energy consumption compared to conventional units on the market
- High temperature accuracy thanks to APT.line™ technology
- Adjustable fan speed
- Controller with LCD display and enhanced time functions
- Electromechanical control of the exhaust air flap
- 2 chrome-plated racks
- Units up to 115 liters are stackable
- Class 2 integrated independent adjustable temperature safety device (DIN 12880) with visual alarm
- Ergonomic handle design
- Ethernet interface
- USB port for recording data
Technical specifications
| Data | ||
| Designation | FED720-400V | FED720UL-208V |
| Article Number | 9010-0301 | 9010-0302 |
| Option model | Standard | Standard |
| Performance Data Temperature | ||
| Temperature range | +12 °C above ambient temperature to 300 °C | +12 °C above ambient temperature to 300 °C |
| Temperature uniformity at 150°C | 2.0 ± K | 2.0 ± K |
| Temperature fluctuation at 150°C | 0.6 ± K | 0.6 ± K |
| Heating up time to 150°C | 25 min | 25 min |
| Recovery time after door was opened for 30 s at 150°C | 6 min | 6 min |
| Air change data | ||
| Air change rate at 100°C | 5 x/h | 5 x/h |
| Electrical data | ||
| Rated Voltage | 400 V | 208 V |
| Power frequency | 50/60 Hz | 60 Hz |
| Nominal power | 4,5 kW | 4,5 kW |
| Unit fuse | 16,0 A | 16,0 A |
| Phase (Nominal voltage) | 3~/N/PE | 3~ |
| Dimensions and weights | ||
| Interior volume | 26 cu.ft. | 26 cu.ft. |
| Net weight of the unit (empty) | 366 lbs | 366 lbs |
| Load per rack | 99 lbs | 99 lbs |
| Permitted load | 694 lbs | 694 lbs |
| Wall clearance back | 6,3 in | 6,3 in |
| Wall clearance sidewise | 3,94 in | 3,94 in |
| Housing dimensions not incl. fittings and connections | ||
| Width net | 45.87 in | 45.87 in |
| Height net | 62.6 in | 62.6 in |
| Depth net | 34.25 in | 34.25 in |
| Internal Dimensions | ||
| Interior width | 39.37 in | 39.37 in |
| Interior height | 51.18 in | 51.18 in |
| Interior depth | 22.44 in | 22.44 in |
| Unit doors | 2 | 2 |
| Environment-specific data | ||
| Sound-pressure level | 43 dB(A) | 43 dB(A) |
| Energy consumption at 100°C | 480 Wh/h | 480 Wh/h |
| Energy consumption at 150°C | 800 Wh/h | 800 Wh/h |
| Fixtures | ||
| Number of shelves (std./max.) | 2/16 | 2/16 |
Downloads
Options, Accessories, Services
Sorry, no result could be found. Check the spelling or use a new search term.
Rack - chrome plated
Rack - stainless steel
Shelf, perforated - Stainless steel
Rack, heavy load - Stainless steel
APT-COM™ 4 BASIC-Edition - for simple logging and documentation requirements with up to 5 networked units.
APT-COM™ 4 PROFESSIONAL-Edition - convenient unit and user management built on the BASIC edition. Suitable for networking up to 100 units.
APT-COM™ 4 GLP-Edition - for working under GLP-compliant conditions. Measured values are documented in a tamper-proof way in line with the requirements of FDA Regulation 21 CFR 11.
pH-neutral detergent - concentrated, for gentle remove of residual contaminants; 1 kg
Sorry, no result could be found. Check the spelling or use a new search term.
Access port with silicone plug - 10 mm, right
Access port with silicone plug - 10 mm, left
Access port with silicone plug - 30 mm, top
Access port with silicone plug - 50 mm, top
Pt 100 temperature sensor - additional flexible Pt 100, interior, for displaying the temperature on the unit display
Access port with silicone plug - 30 mm, right
Access port with silicone plug - 30 mm, left
Access port with silicone plug - 50 mm, right
Access port with silicone plug - 50 mm, left
Access port with silicone plug - 100 mm, top
Access port with silicone plug - 10 mm, top
Access port with silicone plug - 100 mm, left
Access port with silicone plug - 100 mm, right
Door lock - lockable door handle
Viewing window and interior lighting - 208/400 V option model, viewing window (triple insulating glass) in door, 300 x 480 mm, and 30 W interior lighting
HEPA air filter - On fresh air supply for unit; filter classification H14 (acc. to. EN 1822-1:2009, > 99.995% at 0.3 µm); not in conjunction with reinforced heating output, reinforced fan
Sorry, no result could be found. Check the spelling or use a new search term.
Temperature calibration - Extension of calibration of one (1) additional test temperature specified by the user in the center of the usable space, including certificate Travel to be quoted separate.
Temperature measurement, 9 measuring points - Temperature measurement with 9 measuring points with a set value specified by the user, including certificate. Travel to be quoted separate. A total of 3 shelvesare required.
Temperature measurement, 18 measuring points - Temperature measurement with 18 measuring points with a set value specified by the user, including certificate. Travel to be quoted separate. A total of 2 shelvesare required.
Temperature measurement, 27 measuring points - Temperature measurement with 27 measuring points with a set value specified by the user, including certificate. Travel to be quoted separate. A total of 3 shelvesare required.
Execution of IQ/OQ/PQ - Execution of IQ/OQ/PQ in accordance with qualification folder. Travel to be quoted separate.
BRONZE 3-year maintenance contract - Annual Device Inspection Includes Preventive Maintenance with Calibration Certificate. 10% Discount on Spare Parts and Labor for any additional Services needed Wear + tear parts to be purchased or billed to customer. Travel to be quoted as separate.
Unit installation - Connect the unit to the customer-side connections (electricity, water, wastewater, gas), basic functions check, brief operating instructions. (excl.: unpacking, setup, controller instructions, programming, installation work) Travel to be quoted as separate.
Unit instructions - Instruction regarding operating principle and basic functions of the unit, operation of the control electronics including programming. Travel to be quoted as separate Limited to 3 hours.
1-year warranty extension - The warranty is extended by 1 year from the delivery date, wear parts are excluded
Execution of IQ/OQ - Execution of IQ/OQ in accordance with qualification folder. Travel to be quoted separate.
SILVER 3-year maintenance contract - Annual Device Inspection Includes Preventive Maintenance with Calibration Certificate 15% Discount on Spare Parts and Labor for any additional Services needed One additional courtesy visit per year can be used for service visit or operator training. Wear + tear parts to be purchased or billed to customer. Travel to be quoted separate.
GOLD 3-year maintenance contract - Annual Device Inspection Includes Preventive Maintenance with Calibration Certificate of Temperature. 25% Discount on Spare Parts and Labor for any additional Services needed. One additional courtesy visit per year can be used for service visit or operator. Wear + tear parts to be purchased or billed to customer. Travel to be quoted separate.
Preventive maintenance - One-off maintenance service in accordance with maintenance schedule. Visual inspection of mechanical and electrical components, testing of all key functions. Checking a test temperature specified by the user in center of usable space without certificate. Wear + tear parts need to be purchased or billed to customer. Travel to be quoted as separate.
Temperature calibration - Calibration of one (1) test temperature specified by the user in center of chamber, including certificate Travel to be quoted separate.
Measurement of air exchange rate - including certificate (in accordance with ASTM D5374)