Model 23
Dimensions
Product image
Product image

Model ED 23 | Drying and heating chambers with gravity convection

Benefits
  • Uniform drying conditions thanks to APT.line™ technology
  • Identical test conditions throughout the chamber interior independent of sample size and quantity
  • Outstanding thermal insulation saves operation costs

Configure your product with options and accessories and request it directly here.

Product description

The strengths of a BINDER ED series drying chamber include routine drying and sterilization tasks up to 300 °C. Thanks to the gravity convection, all thermal processes which use this drying chamber are highly efficient. The ED series ensures fast, uniform drying.

Important characteristics

  • Temperature range: room temperature plus 5 °C to 300 °C
  • APT.line™ preheating chamber technology
  • Gravity convection
  • Adjustable exhaust air flap
  • Controller with timer functions
  • 2 chrome-plated racks
  • Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm

Technical specifications

Data
DesignationED023-230VED023UL-120V
Article Number9010-01909010-0192
Option modelStandardStandard
Performance Data Temperature
Temperature range+5 °C above ambient temperature to 300 °C+5 °C above ambient temperature to 300 °C
Temperature uniformity at 150°C2.8 ± K2.8 ± K
Temperature fluctuation at 150°C0.5 ± K0.5 ± K
Heating up time to 150°C28 min28 min
Recovery time after door was opened for 30 s at 150°C5 min5 min
Air change data
Air change rate at 150°C13 x/h13 x/h
Electrical data
Rated Voltage230 V120 V
Power frequency50/60 Hz60 Hz
Nominal power0,8 kW0,8 kW
Unit fuse10,0 A12,5 A
Phase (Nominal voltage)1~1~
Dimensions and weights
Interior volume1 cu.ft.1 cu.ft.
Net weight of the unit (empty)60 lbs60 lbs
Load per rack26 lbs26 lbs
Permitted load55 lbs55 lbs
Wall clearance back3,94 in3,94 in
Wall clearance sidewise1,97 in1,97 in
Housing dimensions not incl. fittings and connections
Width net17.13 in17.13 in
Height net19.49 in19.49 in
Depth net20.47 in20.47 in
Internal Dimensions
Interior width8.74 in8.74 in
Interior height12.99 in12.99 in
Interior depth11.81 in11.81 in
Unit doors11
Environment-specific data
Energy consumption at 150°C148 Wh/h148 Wh/h
Fixtures
Number of shelves (std./max.)2/42/4

Downloads

Drying and heating chambers

ED | ED-S | FD | FD-S | FED | FP | M series

pdf, 2.21 MB

Drying and heating chambers (FP series) - Drying chambers, constant climate chambers, dynamic climate chambers and cooling incubators test the service life of MinebeaMitsumi spindle motors

MinebeaMitsumi Technology Center Europe GmbH

pdf, 479.77 KB

BD | ED |  FD series

pdf, 10.14 MB

Data Sheet Model ED 23

pdf, 189.59 KB

Options, Accessories, Services

Sorry, no result could be found. Check the spelling or use a new search term.

Rubber pads

Set anti-slip feet

8012-2030

Rack

Chrome plated

8012-2031

Shelves

Rack - chrome plated

Rack

Stainless steel

8012-2046

Shelves

Rack - stainless steel

pH-neutral detergent

Concentrated, for gentle remove of residual contaminants; 1 kg

8012-2250

1002-0016-Neutralreiniger-100x85_Web_670x515.jpg

pH-neutral detergent - concentrated, for gentle remove of residual contaminants; 1 kg

Qualification documents

IQ/OQ documents – supporting documents for validation performed by customers, consisting of: IQ/OQ checklists incl. calibration guide and comprehensive unit documentation; parameters: temperature, CO2, O2, pressure, depending on unit. Hard copy inside folder

7007-0001

Qualification documents

IQ/OQ documents – supporting documents for validation performed by customers, consisting of: IQ/OQ checklists incl. calibration guide and comprehensive unit documentation; parameters: temperature, CO2, O2, pressure, depending on unit. Digital in PDF format

7057-0001

Qualification documents

IQ/OQ/PQ documents – supporting documents for validation performed by customers, according to customer requirements, PQ section added to qualification folder IQ/OQ; parameters: temperature, CO2, O2 – or pressure, depending on unit. Hard copy inside folder

7007-0005

Qualification documents

IQ/OQ/PQ documents – supporting documents for validation performed by customers, according to customer requirements, PQ section added to qualification folder IQ/OQ; parameters: temperature, CO2, O2 – or pressure, depending on unit. Digital in PDF format

7057-0005

Sorry, no result could be found. Check the spelling or use a new search term.

Measurement of air exchange rate

In accordance with ASTM D5374, definition and protocol according to ambient temperature

8012-0195

Access port with silicone plug

Top. 30 mm

8012-0539

Access port with silicone plug

Access port with silicone plug - 30 mm, top

Access port with silicone plug

Back. 30 mm

8012-0540

Access port with silicone plug

Access port with silicone plug - 30 mm, back

Access port with silicone plug

Top. 50 mm

8012-0545

Access port with silicone plug

Access port with silicone plug - 50 mm, top

Access port with silicone plug

Back. 50 mm

8012-0546

Access port with silicone plug

Access port with silicone plug - 50 mm, back

Calibration certificate, expanded

For temperature; for extending the measurement in center of chamber to include another test temperature

8012-1115

Calibration certificate, temperature

For temperature, measurement in center of chamber at specified temperature

8012-1134

Access port with silicone plug

Top. 10 mm

8012-1262

Access port with silicone plug

Access port with silicone plug - 10 mm, top

Access port with silicone plug

Right. 10 mm

8012-1268

Access port with silicone plug

Access port with silicone plug - 10 mm, right

Access port with silicone plug

Left. 10 mm

8012-1275

Access port with silicone plug

Access port with silicone plug - 10 mm, left

Access port with silicone plug

Back. 10 mm

8012-1282

Access port with silicone plug

Access port with silicone plug - 10 mm, back

Calibration certificate, temperature

Temperature measurement incl. certificate, 9 measuring points at specified temperature

8012-1551

Calibration certificate, temperature

Temperature measurement incl. certificate, 15- 18 measuring points at specified temperature

8012-1572

Calibration certificate, temperature

Temperature measurement incl. certificate and 27 measuring points at specified temperature

8012-1593

Door gasket

Made of FKM, silicone-free

8012-0493

Door lock

Lockable door handle

8012-1246

Door lock

Door lock - lockable door handle

Analog output 4-20 mA

For temperature values (output not adjustable)

8012-0430

Access port with silicone plug

Left. 30 mm

8012-0537

Access port with silicone plug

Access port with silicone plug - 30 mm, left

Access port with silicone plug

Right. 50 mm

8012-0541

Access port with silicone plug

Access port with silicone plug - 50 mm, right

Access port with silicone plug

Left. 50 mm

8012-0542

Access port with silicone plug

Access port with silicone plug - 50 mm, left

Pt 100 temperature sensor

Additional flexible Pt 100, interior, with external connection incl. LEMO connector (3-pin)

8012-1173

Pt 100 temperature sensor

Pt 100 temperature sensor - additional flexible Pt 100, interior, with external connection incl. LEMO connector (3-pin)

Access port with silicone plug

Right. 30 mm

8012-0536

Access port with silicone plug

Access port with silicone plug - 30 mm, right

Viewing window and interior lighting

Viewing window in door, 180 x 180 mm, and 15 W interior lighting. 230 V option model

8012-0378

Viewing window in the door

Viewing window and interior lighting - 230 V option model, viewing window in door, 180 x 180 mm, and 15 W interior lighting

Class 3.1 independent temperature safety device

With visual alarm (DIN 12880)

8012-0032

Alarm function for overheating

Switchable acoustic alarm, with adjustable limit value on the independent temperature safety device. 230/400 V option model

8012-1299

Sorry, no result could be found. Check the spelling or use a new search term.

Temperature calibration

Extension of calibration of one (1) additional test temperature specified by the user in the center of the usable space, including certificate Travel to be quoted separate

DL30-0102

Services

Temperature calibration - Extension of calibration of one (1) additional test temperature specified by the user in the center of the usable space, including certificate Travel to be quoted separate.

Temperature measurement, 9 measuring points

Temperature measurement with 9 measuring points with a set value specified by the user, including certificate. Travel to be quoted separate.  A total of 3 shelvesare required

DL30-0109

Services

Temperature measurement, 9 measuring points - Temperature measurement with 9 measuring points with a set value specified by the user, including certificate. Travel to be quoted separate.  A total of 3 shelvesare required.

Temperature measurement, 18 measuring points

Temperature measurement with 18 measuring points with a set value specified by the user, including certificate. Travel to be quoted separate.  A total of 2 shelvesare required

DL30-0118

Services

Temperature measurement, 18 measuring points - Temperature measurement with 18 measuring points with a set value specified by the user, including certificate. Travel to be quoted separate.  A total of 2 shelvesare required.

Temperature measurement, 27 measuring points

Temperature measurement with 27 measuring points with a set value specified by the user, including certificate. Travel to be quoted separate.  A total of 3 shelvesare required

DL30-0127

Services

Temperature measurement, 27 measuring points - Temperature measurement with 27 measuring points with a set value specified by the user, including certificate. Travel to be quoted separate.  A total of 3 shelvesare required.

Execution of IQ/OQ/PQ

Execution of IQ/OQ/PQ in accordance with qualification folder. Travel to be quoted separate

DL44-0500

Services

Execution of IQ/OQ/PQ - Execution of IQ/OQ/PQ in accordance with qualification folder. Travel to be quoted separate.

BRONZE 3-year maintenance contract

Annual Device Inspection Includes Preventive Maintenance with Calibration Certificate. 10% Discount on Spare Parts and Labor for any additional Services needed Wear + tear parts to be purchased or billed to customer. Travel to be quoted as separate

DL20-0720

Services

BRONZE 3-year maintenance contract - Annual Device Inspection Includes Preventive Maintenance with Calibration Certificate. 10% Discount on Spare Parts and Labor for any additional Services needed Wear + tear parts to be purchased or billed to customer. Travel to be quoted as separate.

SILVER 3-year maintenance contract

Annual Device Inspection Includes Preventive Maintenance with Calibration Certificate 15% Discount on Spare Parts and Labor for any additional Services needed One additional courtesy visit per year can be used for service visit or operator training. Wear + tear parts to be purchased or billed to customer. Travel to be quoted separate

DL20-0815

Services

SILVER 3-year maintenance contract - Annual Device Inspection Includes Preventive Maintenance with Calibration Certificate 15% Discount on Spare Parts and Labor for any additional Services needed One additional courtesy visit per year can be used for service visit or operator training. Wear + tear parts to be purchased or billed to customer. Travel to be quoted separate.

GOLD 3-year maintenance contract

Annual Device Inspection Includes Preventive Maintenance with Calibration Certificate of Temperature. 25% Discount on Spare Parts and Labor for any additional Services needed. One additional courtesy visit per year can be used for service visit or operator. Wear + tear parts to be purchased or billed to customer. Travel to be quoted separate

DL20-0915

Services

GOLD 3-year maintenance contract - Annual Device Inspection Includes Preventive Maintenance with Calibration Certificate of Temperature. 25% Discount on Spare Parts and Labor for any additional Services needed. One additional courtesy visit per year can be used for service visit or operator. Wear + tear parts to be purchased or billed to customer. Travel to be quoted separate.

Unit installation

Connect the unit to the customer-side connections (electricity, water, wastewater, gas), basic functions check, brief operating instructions. (excl.: unpacking, setup, controller instructions, programming, installation work) Travel to be quoted as separate

DL10-0100

Services

Unit installation - Connect the unit to the customer-side connections (electricity, water, wastewater, gas), basic functions check, brief operating instructions. (excl.: unpacking, setup, controller instructions, programming, installation work) Travel to be quoted as separate.

Unit instructions

Instruction regarding operating principle and basic functions of the unit, operation of the control electronics including programming. Travel to be quoted as separate Limited to 3 hours

DL10-0500

Services

Unit instructions - Instruction regarding operating principle and basic functions of the unit, operation of the control electronics including programming. Travel to be quoted as separate Limited to 3 hours.

Preventive maintenance 

One-off maintenance service in accordance with maintenance schedule. Visual inspection of mechanical and electrical components, testing of all key functions. Checking a test temperature specified by the user in center of usable space without certificate. Wear + tear parts need to be purchased or billed to customer. Travel to be quoted as separate

DL20-0601

Services

Preventive maintenance  - One-off maintenance service in accordance with maintenance schedule. Visual inspection of mechanical and electrical components, testing of all key functions. Checking a test temperature specified by the user in center of usable space without certificate. Wear + tear parts need to be purchased or billed to customer. Travel to be quoted as separate.

Temperature calibration

Calibration of one (1) test temperature specified by the user in center of chamber, including certificate Travel to be quoted separate

DL30-0110

Services

Temperature calibration - Calibration of one (1) test temperature specified by the user in center of chamber, including certificate Travel to be quoted separate.

1-year warranty extension

The warranty is extended by 1 year from the delivery date, wear parts are excluded

DL50-0010

Services

1-year warranty extension - The warranty is extended by 1 year from the delivery date, wear parts are excluded

Execution of IQ/OQ

Execution of IQ/OQ in accordance with qualification folder. Travel to be quoted separate

DL40-0100

Services

Execution of IQ/OQ - Execution of IQ/OQ in accordance with qualification folder. Travel to be quoted separate.

Measurement of air exchange rate

Including certificate (in accordance with ASTM D5374)

DL33-0000

Services

Measurement of air exchange rate - including certificate (in accordance with ASTM D5374)

Suggested products

Compare products:

Or add another product.