Model 115
Model 115
Dimensions
Solvent curve
Product image
Product image
Product image
Product image

Model FDL 115 | Safety drying chambers for limited quantities of solvent

Benefits
  • Safety concept meets the DIN EN 1539-2010-08 standard
  • Short drying time with high temperature accuracy and defined ventilation
  • Silicone free and dust-free inner chamber
  • Replaceable fresh air cartridges are easy to change and ensure constant fresh air quality

Configure your product with options and accessories and request it directly here.

Product description

The BINDER FDL 115 with silicone and dust-free inner chamber, as well as symmetric airflow dries specimens containing solvents and thus this safety drying oven meets all EN 1539 requirements.

Important characteristics

  • Temperature range: +10°C above ambient temperature to +300 °C
  • APT.line™ preheating chamber technology
  • Controller with time-segment and real-time programming
  • 2 chrome-plated racks
  • Fresh-air monitoring with audible and visual alarm and automatic deactivation of heating
  • Replaceable filter cartridge, Class M6
  • Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
  • Computer interface: RS 422
  • Standard door gasket made from solvent-resistant FKM for use at up to 200 °C
  • Optional door gasket made from silicone with limited solvent resistance for use at up to 300 °C

Technical specifications

Data
DesignationFDL115-230V
Article Number9010-0292
Option model

Standard

Performance Data Temperature
Temperature range+10 °C above ambient temperature to 300 °C
Temperature uniformity at 150°C2.5 ± K
Temperature fluctuation at 150°C0.8 ± K
Heating up time to 150°C20 min
Recovery time after door was opened for 30 s at 150°C12 min
Air change data
Air circulation (approx.)20 x/min
Volumetric flow rate of exhaust air acc. to EN 1539 at 50 °C400 ca. L/min
Solvent quantity
Highest permitted solvent quantity (at T-180°C, M-100g/mol, U-40g/m3, K=0,5)4,67 g
Electrical data
Rated Voltage230 V
Power frequency50/60 Hz
Nominal power2,9 kW
Phase (Nominal voltage)1~
Dimensions and weights
Interior volume115 L
Net weight of the unit (empty)90 kg
Load per rack30 kg
Permitted load60 kg
Wall clearance back160 mm
Wall clearance sidewise100 mm
Housing dimensions not incl. fittings and connections
Width net830 mm
Height net805 mm
Depth net685 mm
Internal Dimensions
Interior width600 mm
Interior height435 mm
Interior depth435 mm
Unit doors1
Environment-specific data
Sound-pressure level57 dB(A)
Energy consumption at 150°C1,200 Wh/h
Fixtures
Number of shelves (std./max.)2/5

Downloads

Safety drying chambers

FDL series

pdf, 1.35 MB

Safety drying chambers (FDL series) - BINDER chambers ensure quality

Car manufacturer in Lower Bavaria

pdf, 134.33 KB

Safety drying chambers (FDL series) - Perfect conditions for tests involving solvent-based coatings

FREILACKE - Emil Frei GmbH & Co. KG

pdf, 359.42 KB

FDL series

pdf, 6.76 MB

Data Sheet Model FDL 115

pdf, 435.12 KB

Options, Accessories, Services

Sorry, no result could be found. Check the spelling or use a new search term.

Table on castors

Stable cart, casters with locking brakes, dimensions: W 1,000 x D 800 x H 780 mm

9051-0018

Tischwagen

stable cart, casters with locking brakes, dimensions: W 1,000 x D 800 x H 780 mm

Rack

Stainless steel

8012-2048

Shelves

Rack - stainless steel

Rack

Chrome plated

8012-2034

Shelves

Rack - chrome plated

Shelf, perforated

Stainless steel

8012-2161

Shelf

Shelf, perforated - Stainless steel

APT-COM™ 4 BASIC-Edition

For simple logging and documentation requirements with up to 5 networked units.. version 4, BASIC edition 

9053-0039

Multi Management Software APT-COM™

APT-COM™ 4 BASIC-Edition - for simple logging and documentation requirements with up to 5 networked units.

APT-COM™ 4 PROFESSIONAL-Edition

Convenient unit and user management built on the BASIC edition. Suitable for networking up to 100 units.. version 4, PROFESSIONAL edition

9053-0040

Multi Management Software APT-COM™

APT-COM™ 4 PROFESSIONAL-Edition - convenient unit and user management built on the BASIC edition. Suitable for networking up to 100 units.

APT-COM™ 4 GLP-Edition

For working under GLP-compliant conditions. Measured values are documented in a tamper-proof way in line with the requirements of FDA Regulation 21 CFR 11.. version 4, GLP edition

9053-0042

Multi Management Software APT-COM™

APT-COM™ 4 GLP-Edition - for working under GLP-compliant conditions. Measured values are documented in a tamper-proof way in line with the requirements of FDA Regulation 21 CFR 11.

pH-neutral detergent

Concentrated, for gentle remove of residual contaminants; 1 kg

8012-2250

1002-0016-Neutralreiniger-100x85_Web_670x515.jpg

pH-neutral detergent - concentrated, for gentle remove of residual contaminants; 1 kg

Qualification documents

IQ/OQ documents – supporting documents for validation performed by customers, consisting of: IQ/OQ checklists incl. calibration guide and comprehensive unit documentation; parameters: temperature, CO2, O2, pressure, depending on unit. Hard copy inside folder

7007-0001

Qualification documents

IQ/OQ documents – supporting documents for validation performed by customers, consisting of: IQ/OQ checklists incl. calibration guide and comprehensive unit documentation; parameters: temperature, CO2, O2, pressure, depending on unit. Digital in PDF format

7057-0001

Qualification documents

IQ/OQ/PQ documents – supporting documents for validation performed by customers, according to customer requirements, PQ section added to qualification folder IQ/OQ; parameters: temperature, CO2, O2 – or pressure, depending on unit. Hard copy inside folder

7007-0005

Qualification documents

IQ/OQ/PQ documents – supporting documents for validation performed by customers, according to customer requirements, PQ section added to qualification folder IQ/OQ; parameters: temperature, CO2, O2 – or pressure, depending on unit. Digital in PDF format

7057-0005

Sorry, no result could be found. Check the spelling or use a new search term.

Coil coating door flap

Door flap for coil coating tests

8012-0179

Coil coating door flap

Coil coating door flap - door flap for coil coating tests

Door gasket

Made from silicone, resistant to high temperatures for operating temperatures above 200°C. Note: The gasket wears out more quickly in temperatures above 250°C

8012-0499

Door gasket

Door gasket - made from silicone, resistant to high temperatures for operating temperatures above 200°C. Note: The gasket wears out more quickly in temperatures above 250°C  

Calibration certificate, expanded

For temperature; for extending the measurement in center of chamber to include another test temperature

8012-1117

Calibration certificate, temperature

For temperature, measurement in center of chamber at specified temperature

8012-1136

Measurement of air exchange rate

In accordance with ASTM D5374, definition and protocol according to ambient temperature

8012-1196

Door lock

Lockable door handle

8012-1248

Door lock

Door lock - lockable door handle

Access port with silicone plug

Right. 10 mm

8012-1270

Access port with silicone plug

Access port with silicone plug - 10 mm, right

Access port with silicone plug

Left. 10 mm

8012-1277

Access port with silicone plug

Access port with silicone plug - 10 mm, left

Access port with silicone plug

Back. 10 mm

8012-1284

Access port with silicone plug

Access port with silicone plug - 10 mm, back

Access port with silicone plug

Right. 30 mm

8012-1355

Access port with silicone plug

Access port with silicone plug - 30 mm, right

Access port with silicone plug

Left. 30 mm

8012-1360

Access port with silicone plug

Access port with silicone plug - 30 mm, left

Calibration certificate, temperature

Temperature measurement incl. certificate, 9 measuring points at specified temperature

8012-1553

Calibration certificate, temperature

Temperature measurement incl. certificate, 15- 18 measuring points at specified temperature

8012-1574

Calibration certificate, temperature

Temperature measurement incl. certificate and 27 measuring points at specified temperature

8012-1595

Sorry, no result could be found. Check the spelling or use a new search term.

Calibration certificate, temperature

Extension of calibration of one (1) additional test temperature specified by the user in the center of the usable space, including certificate

DL30-0102

Services

Calibration certificate, temperature - Extension of calibration of one (1) additional test temperature specified by the user in the center of the usable space, including certificate

Temperature measurement, 9 measuring points

Temperature measurement with 9 measuring points with a set value specified by the user, including certificate

DL30-0109

Services

Temperature measurement, 9 measuring points - Temperature measurement with 9 measuring points with a set value specified by the user, including certificate

Temperature measurement, 18 measuring points

Temperature measurement with 18 measuring points with a set value specified by the user, including certificate

DL30-0118

Services

Temperature measurement, 18 measuring points - Temperature measurement with 18 measuring points with a set value specified by the user, including certificate

Temperature measurement, 27 measuring points

Temperature measurement with 27 measuring points with a set value specified by the user, including certificate

DL30-0127

Services

Temperature measurement, 27 measuring points - Temperature measurement with 27 measuring points with a set value specified by the user, including certificate

Execution of IQ/OQ/PQ

Execution of IQ/OQ/PQ in accordance with qualification folder

DL44-0500

Services

Execution of IQ/OQ/PQ - Execution of IQ/OQ/PQ in accordance with qualification folder

BRONZE 3-year maintenance contract

Maintenance service as contractually agreed, visual inspection of mechanical and electrical components, check of control response, 20% discount on spare parts

DL20-0720

Services

BRONZE 3-year maintenance contract - Maintenance service as contractually agreed, visual inspection of mechanical and electrical components, check of control response, 20% discount on spare parts

Unit commissioning

Connect the unit to the customer-side connections (electricity, water, wastewater, gas), basic functions check, brief operating instructions. (excl.: unpacking, setup, controller instructions, programming, installation work) 

DL10-0100

Services

Unit commissioning - Connect the unit to the customer-side connections (electricity, water, wastewater, gas), basic functions check, brief operating instructions. (excl.: unpacking, setup, controller instructions, programming, installation work) 

Unit instructions

Instruction regarding operating principle and basic functions of the unit, operation of the control electronics including programming

DL10-0500

Services

Unit instructions - Instruction regarding operating principle and basic functions of the unit, operation of the control electronics including programming

1-year warranty extension

The warranty is extended by 1 year from the delivery date, wear parts are excluded

DL50-0010

Services

1-year warranty extension - The warranty is extended by 1 year from the delivery date, wear parts are excluded

Execution of IQ/OQ

Execution of IQ/OQ in accordance with qualification folder

DL40-0100

Services

Execution of IQ/OQ - Execution of IQ/OQ in accordance with qualification folder

SILVER 3-year maintenance contract

Maintenance service as contractually agreed, visual inspection of mechanical and electrical components, check of control response, 20% discount on spare parts, testing of all key functions, calibration of one test temperature specified by the user in the center of the usable space, without certificate 

DL20-0810

Services

SILVER 3-year maintenance contract - Maintenance service as contractually agreed, visual inspection of mechanical and electrical components, check of control response, 20% discount on spare parts, testing of all key functions, calibration of one test temperature specified by the user in the center of the usable space, without certificate 

Maintenance

One-off maintenance service in accordance with maintenance schedule. Visual inspection of mechanical and electrical components, testing of all key functions. Calibration of a test temperature specified by the user in center of usable space without certificate

DL20-0602

Services

Maintenance - One-off maintenance service in accordance with maintenance schedule. Visual inspection of mechanical and electrical components, testing of all key functions. Calibration of a test temperature specified by the user in center of usable space without certificate

Calibration certificate, temperature

Calibration of one (1) test temperature specified by the user in center of chamber, including certificate

DL30-0120

Services

Calibration certificate, temperature - Calibration of one (1) test temperature specified by the user in center of chamber, including certificate

Measurement of air exchange rate

Including certificate (in accordance with ASTM D5374)

DL33-0000

Services

Measurement of air exchange rate - including certificate (in accordance with ASTM D5374)

GOLD 3-year maintenance contract

Maintenance service as contractually agreed, visual inspection of mechanical and electrical components, check of control response, 20% discount on spare parts, testing of all key functions, replacement of wear parts, calibration of one test temperature specified by the user in the center of the usable space, including certificate

DL20-0985

Services

GOLD 3-year maintenance contract - Maintenance service as contractually agreed, visual inspection of mechanical and electrical components, check of control response, 20% discount on spare parts, testing of all key functions, replacement of wear parts, calibration of one test temperature specified by the user in the center of the usable space, including certificate

Compare products:

Or add another product.