Model FDL 115 | Safety drying chambers for limited quantities of solvent
Benefits
- Safety concept meets the DIN EN 1539-2010-08 standard
- Short drying time with high temperature accuracy and defined ventilation
- Silicone free and dust-free inner chamber
- Replaceable fresh air cartridges are easy to change and ensure constant fresh air quality
Product description
The BINDER FDL 115 with silicone and dust-free inner chamber, as well as symmetric airflow dries specimens containing solvents and thus this safety drying oven meets all EN 1539 requirements.
Important characteristics
- Temperature range: +10°C above ambient temperature to +300 °C
- APT.line™ preheating chamber technology
- Controller with time-segment and real-time programming
- 2 chrome-plated racks
- Fresh-air monitoring with audible and visual alarm and automatic deactivation of heating
- Replaceable filter cartridge, Class M6
- Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
- Computer interface: RS 422
- Standard door gasket made from solvent-resistant FKM for use at up to 200 °C
- Optional door gasket made from silicone with limited solvent resistance for use at up to 300 °C
Technical specifications
| Data | |
| Designation | FDL115-230V |
| Article Number | 9010-0292 |
| Option model | Standard |
| Performance Data Temperature | |
| Temperature range | +10 °C above ambient temperature to 300 °C |
| Temperature uniformity at 150°C | 2.5 ± K |
| Temperature fluctuation at 150°C | 0.8 ± K |
| Heating up time to 150°C | 20 min |
| Recovery time after door was opened for 30 s at 150°C | 12 min |
| Air change data | |
| Air circulation (approx.) | 20 x/min |
| Volumetric flow rate of exhaust air acc. to EN 1539 at 50 °C | 400 ca. L/min |
| Solvent quantity | |
| Highest permitted solvent quantity (at T-180°C, M-100g/mol, U-40g/m3, K=0,5) | 4,67 g |
| Electrical data | |
| Rated Voltage | 230 V |
| Power frequency | 50/60 Hz |
| Nominal power | 2,9 kW |
| Phase (Nominal voltage) | 1~ |
| Dimensions and weights | |
| Interior volume | 115 L |
| Net weight of the unit (empty) | 90 kg |
| Load per rack | 30 kg |
| Permitted load | 60 kg |
| Wall clearance back | 160 mm |
| Wall clearance sidewise | 100 mm |
| Housing dimensions not incl. fittings and connections | |
| Width net | 830 mm |
| Height net | 805 mm |
| Depth net | 685 mm |
| Internal Dimensions | |
| Interior width | 600 mm |
| Interior height | 435 mm |
| Interior depth | 435 mm |
| Unit doors | 1 |
| Environment-specific data | |
| Sound-pressure level | 57 dB(A) |
| Energy consumption at 150°C | 1,200 Wh/h |
| Fixtures | |
| Number of shelves (std./max.) | 2/5 |
Downloads
Options, Accessories, Services
Sorry, no result could be found. Check the spelling or use a new search term.
stable cart, casters with locking brakes, dimensions: W 1,000 x D 800 x H 780 mm
Rack - stainless steel
Rack - chrome plated
Shelf, perforated - Stainless steel
APT-COM™ 4 BASIC-Edition - for simple logging and documentation requirements with up to 5 networked units.
APT-COM™ 4 PROFESSIONAL-Edition - convenient unit and user management built on the BASIC edition. Suitable for networking up to 100 units.
APT-COM™ 4 GLP-Edition - for working under GLP-compliant conditions. Measured values are documented in a tamper-proof way in line with the requirements of FDA Regulation 21 CFR 11.
pH-neutral detergent - concentrated, for gentle remove of residual contaminants; 1 kg
Sorry, no result could be found. Check the spelling or use a new search term.
Coil coating door flap - door flap for coil coating tests
Door gasket - made from silicone, resistant to high temperatures for operating temperatures above 200°C. Note: The gasket wears out more quickly in temperatures above 250°C
Door lock - lockable door handle
Access port with silicone plug - 10 mm, right
Access port with silicone plug - 10 mm, left
Access port with silicone plug - 10 mm, back
Access port with silicone plug - 30 mm, right
Access port with silicone plug - 30 mm, left
Sorry, no result could be found. Check the spelling or use a new search term.
Calibration certificate, temperature - Extension of calibration of one (1) additional test temperature specified by the user in the center of the usable space, including certificate
Temperature measurement, 9 measuring points - Temperature measurement with 9 measuring points with a set value specified by the user, including certificate
Temperature measurement, 18 measuring points - Temperature measurement with 18 measuring points with a set value specified by the user, including certificate
Temperature measurement, 27 measuring points - Temperature measurement with 27 measuring points with a set value specified by the user, including certificate
Execution of IQ/OQ/PQ - Execution of IQ/OQ/PQ in accordance with qualification folder
BRONZE 3-year maintenance contract - Maintenance service as contractually agreed, visual inspection of mechanical and electrical components, check of control response, 20% discount on spare parts
Unit commissioning - Connect the unit to the customer-side connections (electricity, water, wastewater, gas), basic functions check, brief operating instructions. (excl.: unpacking, setup, controller instructions, programming, installation work)
Unit instructions - Instruction regarding operating principle and basic functions of the unit, operation of the control electronics including programming
1-year warranty extension - The warranty is extended by 1 year from the delivery date, wear parts are excluded
Execution of IQ/OQ - Execution of IQ/OQ in accordance with qualification folder
SILVER 3-year maintenance contract - Maintenance service as contractually agreed, visual inspection of mechanical and electrical components, check of control response, 20% discount on spare parts, testing of all key functions, calibration of one test temperature specified by the user in the center of the usable space, without certificate
Maintenance - One-off maintenance service in accordance with maintenance schedule. Visual inspection of mechanical and electrical components, testing of all key functions. Calibration of a test temperature specified by the user in center of usable space without certificate
Calibration certificate, temperature - Calibration of one (1) test temperature specified by the user in center of chamber, including certificate
Measurement of air exchange rate - including certificate (in accordance with ASTM D5374)
GOLD 3-year maintenance contract - Maintenance service as contractually agreed, visual inspection of mechanical and electrical components, check of control response, 20% discount on spare parts, testing of all key functions, replacement of wear parts, calibration of one test temperature specified by the user in the center of the usable space, including certificate