Model ED 260 | Drying and heating chambers with natural convection

Benefits
  • Best energy efficiency on the market
  • Excellent temporal and spatial temperature accuracy
  • USB connection for recording data
  • Temperature range up to +300 °C

Configure your product with options and accessories and request it directly here.

Product description

The strengths of a BINDER ED series drying chamber include routine drying and sterilization tasks up to +300 °C. Thanks to the natural convection, all thermal processes which use this drying chamber are highly efficient. The ED series ensures fast, uniform drying.

Important characteristics

  • Temperature range: +5°C above ambient temperature to +300 °C
  • Up to 30% lower energy consumption compared to conventional units on the market
  • High temperature accuracy thanks to APT.line™ technology
  • Natural convection
  • Controller with LCD display
  • Electromechanical control of the exhaust air flap
  • 2 chrome-plated racks
  • Class 2 integrated independent adjustable temperature safety device (DIN 12880) with visual alarm
  • Ergonomic handle design
  • USB port for recording data

Technical specifications

Data
Designation

ED260-230V

ED260UL-240V

Article Number9010-03399010-0340
Option model

Standard

Standard

Performance Data Temperature
Temperature range+5 °C above ambient temperature to 300 °C+5 °C above ambient temperature to 300 °C
Temperature uniformity at 150°C1.8 ± K1.8 ± K
Temperature fluctuation at 150°C0.8 ± K0.8 ± K
Heating up time to 150°C50 min50 min
Recovery time after door was opened for 30 s at 150°C20 min20 min
Air change data
Air change rate at 100°C3 x/h3 x/h
Electrical data
Rated Voltage230 V240 V
Power frequency50/60 Hz60 Hz
Nominal power2,25 kW2,45 kW
Unit fuse12,5 A16,0 A
Phase (Nominal voltage)1~1~
Dimensions and weights
Interior volume255 L255 L
Net weight of the unit (empty)85 kg85 kg
Load per rack40 kg40 kg
Permitted load270 kg270 kg
Wall clearance back160 mm160 mm
Wall clearance sidewise100 mm100 mm
Housing dimensions not incl. fittings and connections
Width net810 mm810 mm
Height net940 mm940 mm
Depth net760 mm760 mm
Internal Dimensions
Interior width610 mm610 mm
Interior height760 mm760 mm
Interior depth550 mm550 mm
Unit doors11
Environment-specific data
Energy consumption at 100°C195 Wh/h195 Wh/h
Energy consumption at 150°C355 Wh/h355 Wh/h
Fixtures
Number of shelves (std./max.)2/82/8

Downloads

Drying and heating chambers

ED | ED-S | FD | FD-S | FED | FP | M series

pdf, 2.21 MB

Drying and heating chambers (ED | FED series) - Carbon frames for racing bikes

SWT Swiss Manufacturing Technology AG, BMC Swiss Cycling Technologie

pdf, 1.9 MB

Drying and heating chambers (ED | FD series) - Testing the functionality and service life of cables

U. I. Lapp GmbH

pdf, 189.08 KB

BD | BF | ED | FD | FED series

pdf, 18.66 MB

Quick Start Guide BD | BF | ED | FD | FED series

pdf, 1.65 MB

Data Sheet Model ED 260

pdf, 407.66 KB

Options, Accessories, Services

Sorry, no result could be found. Check the spelling or use a new search term.

Rubber pads

Set anti-slip feet

8012-2030

Rack

Max. load per rack 40 kg. chrome plated

8012-2043

Shelves

Rack - chrome plated

Rack

Max. load per rack 40 kg. stainless steel

8012-2173

Shelves

Rack - stainless steel

Shelf, perforated

Max. load per rack 40 kg. Stainless steel

8012-2181

Shelf

Shelf, perforated - Stainless steel

Rack, heavy load

Max. load per rack 70 kg. Stainless steel

8012-2187

Reinforced shelf racks

Rack, heavy load - Stainless steel

Table on castors

Stable cart, casters with locking brakes, dimensions: W 1,300 x D 800 x H 780 mm

9051-0019

Ethernet interface

For Multi Management Software APT-COM™

8012-1645

APT-COM™ 4 BASIC-Edition

For simple logging and documentation requirements with up to 5 networked units.. version 4, BASIC edition 

9053-0039

Multi Management Software APT-COM™

APT-COM™ 4 BASIC-Edition - for simple logging and documentation requirements with up to 5 networked units.

APT-COM™ 4 PROFESSIONAL-Edition

Convenient unit and user management built on the BASIC edition. Suitable for networking up to 100 units.. version 4, PROFESSIONAL edition

9053-0040

Multi Management Software APT-COM™

APT-COM™ 4 PROFESSIONAL-Edition - convenient unit and user management built on the BASIC edition. Suitable for networking up to 100 units.

APT-COM™ 4 GLP-Edition

For working under GLP-compliant conditions. Measured values are documented in a tamper-proof way in line with the requirements of FDA Regulation 21 CFR 11.. version 4, GLP edition

9053-0042

Multi Management Software APT-COM™

APT-COM™ 4 GLP-Edition - for working under GLP-compliant conditions. Measured values are documented in a tamper-proof way in line with the requirements of FDA Regulation 21 CFR 11.

pH-neutral detergent

Concentrated, for gentle remove of residual contaminants; 1 kg

8012-2250

1002-0016-Neutralreiniger-100x85_Web_670x515.jpg

pH-neutral detergent - concentrated, for gentle remove of residual contaminants; 1 kg

Qualification documents

IQ/OQ documents – supporting documents for validation performed by customers, consisting of: IQ/OQ checklists incl. calibration guide and comprehensive unit documentation; parameters: temperature, CO2, O2, pressure, depending on unit. Hard copy inside folder

7007-0001

Qualification documents

IQ/OQ documents – supporting documents for validation performed by customers, consisting of: IQ/OQ checklists incl. calibration guide and comprehensive unit documentation; parameters: temperature, CO2, O2, pressure, depending on unit. Digital in PDF format

7057-0001

Qualification documents

IQ/OQ/PQ documents – supporting documents for validation performed by customers, according to customer requirements, PQ section added to qualification folder IQ/OQ; parameters: temperature, CO2, O2 – or pressure, depending on unit. Hard copy inside folder

7007-0005

Qualification documents

IQ/OQ/PQ documents – supporting documents for validation performed by customers, according to customer requirements, PQ section added to qualification folder IQ/OQ; parameters: temperature, CO2, O2 – or pressure, depending on unit. Digital in PDF format

7057-0005

Sorry, no result could be found. Check the spelling or use a new search term.

Door lock

Lockable door handle

8012-1664

Door lock

Door lock - lockable door handle

Measurement of air exchange rate

In accordance with ASTM D5374, definition and protocol according to ambient temperature

8012-0195

Access port with silicone plug

Top. 30 mm

8012-0539

Access port with silicone plug

Access port with silicone plug - 30 mm, top

Access port with silicone plug

Back. 30 mm

8012-0540

Access port with silicone plug

Access port with silicone plug - 30 mm, back

Access port with silicone plug

Top. 50 mm

8012-0545

Access port with silicone plug

Access port with silicone plug - 50 mm, top

Access port with silicone plug

Back. 50 mm

8012-0546

Access port with silicone plug

Access port with silicone plug - 50 mm, back

Class 3.1 independent temperature safety device

With visual alarm (DIN 12880)

8012-0985

Access port with silicone plug

Right. 30 mm

8012-1017

Access port with silicone plug

Access port with silicone plug - 30 mm, right

Access port with silicone plug

Left. 30 mm

8012-1018

Access port with silicone plug

Access port with silicone plug - 30 mm, left

Access port with silicone plug

Right. 50 mm

8012-1021

Access port with silicone plug

Access port with silicone plug - 50 mm, right

Access port with silicone plug

Left. 50 mm

8012-1022

Access port with silicone plug

Access port with silicone plug - 50 mm, left

Access port with silicone plug

Top. 100 mm

8012-1029

Access port with silicone plug

Access port with silicone plug - 100 mm, top

Calibration certificate, expanded

For temperature; for extending the measurement in center of chamber to include another test temperature

8012-1115

Calibration certificate, temperature

For temperature, measurement in center of chamber at specified temperature

8012-1134

Access port with silicone plug

Top. 10 mm

8012-1262

Access port with silicone plug

Access port with silicone plug - 10 mm, top

Access port with silicone plug

Right. 10 mm

8012-1268

Access port with silicone plug

Access port with silicone plug - 10 mm, right

Access port with silicone plug

Left. 10 mm

8012-1275

Access port with silicone plug

Access port with silicone plug - 10 mm, left

Access port with silicone plug

Back. 10 mm

8012-1282

Access port with silicone plug

Access port with silicone plug - 10 mm, back

Calibration certificate, temperature

Temperature measurement incl. certificate, 9 measuring points at specified temperature

8012-1551

Calibration certificate, temperature

Temperature measurement incl. certificate, 15- 18 measuring points at specified temperature

8012-1572

Calibration certificate, temperature

Temperature measurement incl. certificate and 27 measuring points at specified temperature

8012-1593

Pt 100 temperature sensor

Additional flexible Pt 100, interior, for displaying the temperature on the unit display

8012-1619

Pt 100 temperature sensor

Pt 100 temperature sensor - additional flexible Pt 100, interior, for displaying the temperature on the unit display

Analog output 4-20 mA

For temperature values (output not adjustable)

8012-1623

Alarm function for overheating

Switchable acoustic alarm, with adjustable limit value on the independent temperature safety device. Not in combination with the following option: Door with window

8012-1635

Clock

Battery-backed

8012-1640

Access port with silicone plug

Left. 100 mm

8012-1030

Access port with silicone plug

Access port with silicone plug - 100 mm, left

Access port with silicone plug

Right. 100 mm

8012-1031

Access port with silicone plug

Access port with silicone plug - 100 mm, right

Door gasket

Made of FKM, silicone-free

8012-1011

Door gasket

Door gasket - made of FKM, silicone-free  

Interior lighting

With 27 W light bulb. 230/240 V option model

8012-1627

Viewing window and interior lighting

Viewing window (triple-glazed insulating glass) 370 x 370 mm in the door and interior lighting 15 W. Not in combination with the following option: Alarm function for overheating. 230 V option model

8012-1064

Viewing window in the door

Viewing window and interior lighting - 230 V option model, Viewing window (triple-glazed insulating glass) 370 x 370 mm in the door and interior lighting 15 W. Not in combination with the following option: Alarm function for overheating

Sorry, no result could be found. Check the spelling or use a new search term.

Calibration certificate, temperature

Extension of calibration of one (1) additional test temperature specified by the user in the center of the usable space, including certificate

DL30-0102

Services

Calibration certificate, temperature - Extension of calibration of one (1) additional test temperature specified by the user in the center of the usable space, including certificate

Temperature measurement, 9 measuring points

Temperature measurement with 9 measuring points with a set value specified by the user, including certificate

DL30-0109

Services

Temperature measurement, 9 measuring points - Temperature measurement with 9 measuring points with a set value specified by the user, including certificate

Temperature measurement, 18 measuring points

Temperature measurement with 18 measuring points with a set value specified by the user, including certificate

DL30-0118

Services

Temperature measurement, 18 measuring points - Temperature measurement with 18 measuring points with a set value specified by the user, including certificate

Temperature measurement, 27 measuring points

Temperature measurement with 27 measuring points with a set value specified by the user, including certificate

DL30-0127

Services

Temperature measurement, 27 measuring points - Temperature measurement with 27 measuring points with a set value specified by the user, including certificate

Execution of IQ/OQ/PQ

Execution of IQ/OQ/PQ in accordance with qualification folder

DL44-0500

Services

Execution of IQ/OQ/PQ - Execution of IQ/OQ/PQ in accordance with qualification folder

BRONZE 3-year maintenance contract

Maintenance service as contractually agreed, visual inspection of mechanical and electrical components, check of control response, 20% discount on spare parts

DL20-0720

Services

BRONZE 3-year maintenance contract - Maintenance service as contractually agreed, visual inspection of mechanical and electrical components, check of control response, 20% discount on spare parts

SILVER 3-year maintenance contract

Maintenance service as contractually agreed, visual inspection of mechanical and electrical components, check of control response, 20% discount on spare parts, testing of all key functions, calibration of one test temperature specified by the user in the center of the usable space, without certificate 

DL20-0815

Services

SILVER 3-year maintenance contract - Maintenance service as contractually agreed, visual inspection of mechanical and electrical components, check of control response, 20% discount on spare parts, testing of all key functions, calibration of one test temperature specified by the user in the center of the usable space, without certificate 

GOLD 3-year maintenance contract

Maintenance service as contractually agreed, visual inspection of mechanical and electrical components, check of control response, 20% discount on spare parts, testing of all key functions, replacement of wear parts, calibration of one test temperature specified by the user in the center of the usable space, including certificate

DL20-0915

Services

GOLD 3-year maintenance contract - Maintenance service as contractually agreed, visual inspection of mechanical and electrical components, check of control response, 20% discount on spare parts, testing of all key functions, replacement of wear parts, calibration of one test temperature specified by the user in the center of the usable space, including certificate

Unit commissioning

Connect the unit to the customer-side connections (electricity, water, wastewater, gas), basic functions check, brief operating instructions. (excl.: unpacking, setup, controller instructions, programming, installation work) 

DL10-0100

Services

Unit commissioning - Connect the unit to the customer-side connections (electricity, water, wastewater, gas), basic functions check, brief operating instructions. (excl.: unpacking, setup, controller instructions, programming, installation work) 

Unit instructions

Instruction regarding operating principle and basic functions of the unit, operation of the control electronics including programming

DL10-0500

Services

Unit instructions - Instruction regarding operating principle and basic functions of the unit, operation of the control electronics including programming

Maintenance

One-off maintenance service in accordance with maintenance schedule. Visual inspection of mechanical and electrical components, testing of all key functions. Calibration of a test temperature specified by the user in center of usable space without certificate

DL20-0601

Services

Maintenance - One-off maintenance service in accordance with maintenance schedule. Visual inspection of mechanical and electrical components, testing of all key functions. Calibration of a test temperature specified by the user in center of usable space without certificate

Calibration certificate, temperature

Calibration of one (1) test temperature specified by the user in center of chamber, including certificate

DL30-0110

Services

Calibration certificate, temperature - Calibration of one (1) test temperature specified by the user in center of chamber, including certificate

1-year warranty extension

The warranty is extended by 1 year from the delivery date, wear parts are excluded

DL50-0010

Services

1-year warranty extension - The warranty is extended by 1 year from the delivery date, wear parts are excluded

Execution of IQ/OQ

Execution of IQ/OQ in accordance with qualification folder

DL40-0100

Services

Execution of IQ/OQ - Execution of IQ/OQ in accordance with qualification folder

Measurement of air exchange rate

Including certificate (in accordance with ASTM D5374)

DL33-0000

Services

Measurement of air exchange rate - including certificate (in accordance with ASTM D5374)

Suggested products

Compare products:

Or add another product.