Model 56
Model 56
Dimensions
Fail-safe CO₂ automatic diagnosis system for protecting the pH of cell cultures
DuoDoor™ system: Unit door and glass door can be opened with one handle
Variant: divided inner door
Product image
Product image
Product image
Product image
Product image
Product image

Model CB 56 | CO2 incubators with hot air sterilization and heat sterilizable CO2 sensor

Benefits
  • Safe – cell cultures receive maximum protection against contamination
  • Reliable – constant well-being conditions for the cells
  • Smart – simple routine cleaning and convenient operation
  • Economical – efficient operation without consumables

Configure your product with options and accessories and request it directly here.

Product description

The powerful and versatile CB CO₂ incubator is suitable for cell-based assays in drug and medical research. This BINDER incubator is even suitable for complex cultivation experiments or individual environments with hypoxic conditions thanks to its comprehensive assortment of options and accessories.

Important characteristics

  • Optimized humidification system with condensation protection
  • Tried-and-tested anti-contamination concept with 180°C hot air sterilization
  • Hot-air sterilizable CO₂ sensor with single-beam infrared technology
  • Seamless inner chamber made of stainless steel with flanges as shelf support system
  • Fanless interior with Venturi CO₂ gas mixing nozzle
  • Intuitive touchscreen controller
  • Internal data recording, can be read out in open format via USB interface
  • Fault diagnosis system with visual and acoustic alarm
  • Interfaces: Ethernet, USB, zero-voltage alarm contact
  • Fail-safe CO₂ system for protecting the pH of cell cultures

Technical specifications

Data
Designation

CB056-230V

CB056UL-120V

Article Number9640-00059640-0007
Option modelStandardStandard
Performance Data Temperature
Temperature range+4 °C above ambient temperature to 60 °C+4 °C above ambient temperature to 60 °C
Temperature uniformity at 37°C0.3 ± K0.3 ± K
Temperature fluctuation at 37°C0.1 ± K0.1 ± K
Recovery time after door was opened for 30 s at 37°C5 min5 min
Performance Data Climate
Humidity range90 ...95 % RH90 ...95 % RH
Gas Data
CO₂ range0 ...20 Vol.-% CO20 ...20 Vol.-% CO2
CO₂ measuring technologyIRIR
CO₂ recovery time after door was opened for 30 s at 5 vol. % CO₂5 min5 min
Electrical data
Rated Voltage200...230 V100...120 V
Power frequency50/60 Hz50/60 Hz
Nominal power0,8 kW0,8 kW
Unit fuse10,0 A16,0 A
Phase (Nominal voltage)1~1~
Measures
Interior volume53 L53 L
Net weight of the unit (empty)53 kg53 kg
Load per rack10 kg10 kg
Permitted load30 kg30 kg
Wall clearance back100 mm100 mm
Wall clearance sidewise50 mm50 mm
Housing dimensions not incl. fittings and connections
Width net528 mm528 mm
Height net674 mm674 mm
Depth net545 mm545 mm
Internal Dimensions
Interior width400 mm400 mm
Interior height400 mm400 mm
Interior depth330 mm330 mm
Inner doors11
Unit doors11
Environment-specific data
Sound-pressure level41 dB(A)41 dB(A)
Energy consumption at 37°C45 Wh/h45 Wh/h
Fixtures
Number of shelves (std./max.)2/32/3

Downloads

Stacking options for BINDER CO2 incubators

C | CB series

pdf, 161.65 KB

BINDER CO2-shaker

Series CB 170 | CB 260 | CB-S 170 | CB-S 260

pdf, 188.87 KB

CO₂ incubator (CB series) - Stem cell research into healing musculoskeletal disorders

University of Rochester School of Medicine and Dentistry

pdf, 224.21 KB

CO2 incubators (CB series) - Stem cell research

The Huck Institutes of the Life Sciences Pennsylvania State University

pdf, 1.23 MB

CO₂ incubators (CB series) – Successful reproduction of human skin cells

Sysmex Suisse AG

pdf, 198.36 KB

CO2 incubator (CB series) - Herd breeding

BETAGRAN Lipetsk LLC, Lipetsk

pdf, 306.7 KB

CO2 incubators (CB series) - BINDER chambers are used in cancer research

Klinikum rechts der Isar (Rechts der Isar Hospital) Technical University of Munich

pdf, 264.13 KB

CO₂ incubators (CB series) -  Lucerne University of Applied Sciences and Arts delivers new findings in the field of space biology

Lucerne University of Applied Sciences and Arts

pdf, 205.99 KB

Tubular-Bag Sets | Humidification Water Set | Water Container Set

for CB series with active humidification

pdf, 929.89 KB

CB | CBF series

pdf, 13.77 MB

Data Sheet Model CB 56

pdf, 450.75 KB

Compare products: