The BINDER VDL series
Safety vacuum drying oven for flammable solvents
A BINDER safety vacuum drying oven of the VDL series ensures maximum safety when drying organic solvents standard with TÜV/GS. The inner chamber of the VDL safety vacuum drying oven is designed according to ATEX guidelines for Zone II 3 G.
Advantages:
- Maximum occupational safety with one-of-a-kind safety concept
- Fast, condensation-free drying processes
- Homogeneous specimen drying
Areas of application:
-
Chemicals -
Surface Technology -
Pharmaceuticals Industry
| Features | Customer benefits | Characteristics |
|---|---|---|
| APT.line™ |
|
APT.line™ preheating chamber technology
|
| Drying |
|
BINDER Cross-Flow Principle
|
| Safety concept |
|
Tested security
|
| Cleaning |
|
Smooth inner chamber with rounded corners
|
| Complete system |
|
Coordinated, modular system
|
| Accessories and Services |
|
Comprehensive product portfolio
|
- Inner chamber is designed according to ATEX guidelines for Zone II 3G
- Electronically controlled APT.line™ preheating chamber with 2 expansion racks assuring temperature accuracy and reproducible results
- Temperature range from 15 °C (27 °F) above ambient temperature to 200 °C (392 °F)
- MP controller with 2 programs with 10 sections each or switchable to 1 program with 20 sections
- Integrated weekly program timer with real-time function
- Digital temperature setting with an accuracy of one degree
- Elapsed time indicator
- Spring-mounted safety glass panel with shatter protection
- Precision-adjustable ventilation valve
- Precision-adjustable inert gas valve with Cross-Flow-Technology
- Safety features:
- Pressure control device for heating activated < 125 mbar (94 torr)
- Over pressure capsuled instrument panel with compressed air connection and maintenance unit
- Flame protection gasket
- Analog pressure gauge (displays pressure difference between the inner chamber and ambient pressure)
- Electro polished inner chamber, suction and ventilation tubes, pressure container, expansion racks, and ball valve are made of stainless steel
- Door gasket made of tempered silicone
- Independent adjustable temperature safety device class 2 (DIN 12880), with visual temperature alarm
- Measuring port DN 16 in rear panel
- RS 422 interface for use with optional GMP/GLP and FDA guideline 21 CFR Part 11 compliant APT-COM™ DataControlSystem software
- 2 patented, flexible aluminum expansion racks
- All electrical components are decoupled from the inner chamber
- Available as a complete system, with module and vacuum pump. Features:
- Reduced sound level
- Practical working height
- Well balanced system
- BINDER test confirmation
| VDL 23 | ||
|---|---|---|
|
|
Exterior dimensions | |
| Width (mm/inch) | 515 / 20.3 | |
| Height (incl. feet) (mm/inch) | 655 / 25.8 | |
| Height with optional vacuum module (mm/inch) | 625 / 24.6 | |
| Total Height with optional vacuum module (mm/inch) | 1280 / 50.4 | |
| Depth (mm/inch) | 500 / 19.7 | |
| Plus door handle, connection (mm/inch) | 100 / 4.0 | |
| Wall clearance, rear (mm/inch) | 100 / 4.0 | |
| Wall clearance, side (mm/inch) | 135 / 5.3 | |
|
|
Interior dimensions | |
| Width (mm/inch) | 285 / 11.2 | |
| Height (mm/inch) | 285 / 11.2 | |
| Depth (mm/inch) | 285 / 11.2 | |
| Interior volume (l/cu.ft.) | 23 / 0.8 | |
| Expansion racks (aluminum) (number standard/max.) | 2 / 4 | |
| Distance between the racks (mm/inch) | 53 / 2.1 | |
| Usable space per rack (width x depth) (mm/inch) | 234 x 280 / 9.2 x 11 | |
| Load per shelf (kg/lbs.) | 20 / 44 | |
| Permitted total load (kg/lbs.) | 35 / 77 | |
| Weight (empty) (kg/lbs.) | 63 / 139 | |
|
|
Temperature data | |
| Temperature range approx. 15 °C (27 °F) above ambient temperature to (°C/°F) | 200 / 392 | |
| Temperature variation 1) | ||
| at 100 °C (212 °F) (± K) | 1.5 | |
| at 200 °C (392 °F) (± K) | 3 | |
| Temperature fluctuation (± K) 1) | 0.1 | |
| Heating up time 1), 2) | ||
| to 100 °C (212 °F) (min.) | 65 | |
| to 200 °C (392 °F) (min.) | 100 | |
|
|
Vacuum data | |
| Vacuum connection with small flange (DN mm/inch) | 16 / 0.63 | |
| Measuring access port with small flange (DN mm/inch) | 16 / 0.63 | |
| Inert gas connection with flow limiter thread (RP) | 3 / 8 | |
| Air pressure connection for pressurized apparatus (tube clip Ø / mm/inch) | 8 / 0.3 | |
| Permitted end vacuum (mbar / inchHg) | 0.01 / 0.000295 | |
| Leak rate max. (bar/h / inchHg/h) | 0.01 / 2.95 | |
|
|
Electrical data | |
| IP protection class acc. to EN 60529 | IP 54 | |
| Voltage (± 10%) 50 / 60 Hz | 230 | |
| Nominal power (kW) | 0.8 | |
| Energy consumption | ||
| at 100 °C (212 °F) (W) | 105 | |
| at 200 °C (392 °F) (W) | 280 |
1) Value with aluminum racks
2) To 98% of the set value
All technical data are specified for units with standard equipment at an ambient temperature of 25 °C (77 °F) and a line voltage fluctuation of ±10%. These average values have been determined according to the BINDER factory standard, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. Differing ambient temperatures and production-related device-specific variances can lead to varying technical data.
Therefore, we recommend individual customer-specific calibration or validation for applications on the limit of the permitted ambient temperature range.
Measuring port
Vacuum-tight access port into the device for measuring lines (9-pin).
Specimen temperature display
Uses PT 100 sensor and digital temperature display.
Calibration certificate
Measurement in center at predefined test value. Expandable with additional measuring points or test values based on your requirements.
| VDL 23 | |
|---|---|
| ATEX connection kit VP 4, consists of: clamping rings, aluminum universal centering ring, transition piece adapter, anti-static vacuum hose, union nut, elbow | O |
| ATEX connection kit VP 5, consists of: clamping rings, aluminum, universal centering ring, outer centering ring, transition piece adapter, union nut, anti-static vacuum hose, elbow | O |
| Measuring port. For vacuum-tight access port of measuring lines into the device (9-pin) | O |
| Temperature measurement of the specimen via flexible PT 100 sensor (EX-proof via measurement current access port) and digital temperature display | O |
| Factory calibration certificate, measurement in center of chamber at 100 °C (212 °F) or at specified testing temperature | O |
| Extension to factory calibration certificate. Each additional measurement at additional measuring point or testing temperature | O |
| VP 4 chemical membrane pump (nominal air flow 1.9 m³/hour, final pressure 12 mbar), with separator and emission condenser 230 V 1N ~ 50 Hz | O |
| Expansion racks, aluminum | O |
| Expansion racks, stainless steel | O |
| Exchange and calibration to stainless steel expansion racks, standard equipment aluminum expansion racks are replaced | O |
| Door gasket, FKM (Viton) | O |
| Vacuum module (empty) for installation of vacuum pumps | O |
| Vacuum module with membrane pump VP 4, mechanical and electrical membrane pump components EX-proof, ATEX approved (nominal air flow 1.9 m³ / hour, final pressure 12 mbar), with separator and emission condenser, including all necessary vacuum connection parts | O |