Hybridization ovens from the APT.line™
BFED series:
Hybridization ovens with shaker platform
BFED is just as ideally suited for reproducible incubation results as the closely related BFD unit, although it offers a little extra when it comes to versatility. The shaking platform is particularly suited for hybridization in bags, or for in situ hybridization. The shaking frequency is digitally adjustable.
- Electronically controlled APT.line™ preheating chamber technology
- Temperature range of 5 °C (9 °F) above ambient temperature up to 100 °C (212 °F)
- Digital temperature setting with tenth of a degree accuracy
- MS controller with several timer functions
- Time functions: delayed ON, delayed OFF, and temperature-dependent delayed OFF
- Shaking unit, digitally adjustable (20–100 %)
- Adjustable ramp function
- Adjustable heat output (0 to 100 %)
- Removable rotation unit for 12 hybridization bottles
- Low profile collection pan
- Adjustable ventilation by means of rear exhaust duct, 50 mm (2 inch) diameter, with ventilation flap and front ventilation slide
- Inner glass door
- Independent adjustable temperature safety device, Class 3.1 (DIN 12880), with visual temperature alarm
- RS 232 printer and communication interface for communication software APT-COM™ DataControlSystem
- Adjustable intervals for printer
- Units are stackable
- 1 chrome-plated rack
- BINDER test certificate
| BFED 53 | |
| Exterior dimensions | |
| Width (mm/inch) | 659 / 25.9 |
| Height (inclusive feet/castors) (mm/inch) | 617 / 24.3 |
| Depth (mm/inch) | 575 / 22.6 |
| Plus door handle, I-panel and exhaust duct (mm/inch) | 85 / 3.4 |
| Wall clearance rear (mm/inch) | 100 / 3.9 |
| Wall clearance side (mm/inch) | 160 / 6.3 |
| Steam space volume (l/cu.ft.) | 77 / 2.7 |
| Number of doors | 1 |
| Interior dimensions | |
| Width (mm/inch) | 400 / 15.8 |
| Height (mm/inch) | 400 / 15.8 |
| Depth (mm/inch) | 330 / 13.0 |
| Interior volume (l/cu.ft.) | 53 / 1.9 |
| Racks, chrome-plated (number standard/max.) | 1/4 |
| Load per rack (kg/lbs.) | 15 / 33 |
| Permitted total load (kg/lbs.) | 40 / 88 |
| Weight of the unit (empty) (kg/lbs.) | 45 / 99 |
| Temperature data | |
| Temperature range, 5 °C (9 °F) above ambient up to (°C/°F) | 100 / 212 |
| Temperature variation | |
| at 37 °C (98.6 °F) (± °C) | 0,5 |
| at 50 °C (122 °F) (± °C) | 1,5 |
| Temperature fluctuation | |
| at 37 °C (98.6 °F) (± °C) | 0,1 |
| at 50 °C (122 °F) (± °C) | 0,2 |
| Heating up time 1) | |
| to 37 °C (98.6 °F) (Min.) | 12 |
| to 50 °C (122 °F) (Min.) | 22 |
| Recov. time after door was opened for 30 sec. 1) | |
| at 37 °C (98.6 °F) (Min.) | 1 |
| at 50 °C (122 °F) (Min.) | 1,5 |
| Electrical data | |
| Housing protection acc. to EN 60529 | IP 20 |
| Nominal voltage (±10 %) 50 Hz (V) | 230 / 115 |
| Nominal power (W) | 400 |
| Energy consumption | |
| at 37 °C (98.6 °F) (W) | 20 |
| at 50 °C (122 °F) (W) | 130 |
1) up to 98 % of the set value
All specified technical data apply for standard equipment operating at an ambient temperature of + 25 °C (77 °F), with line voltage fluctuations of ± 10%. These typical average values for series equipment were determined in accordance with the BINDER factory standard. They are based upon the recommended distances from the inner chamber walls, namely 10% of height, width and depth. Differing ambient temperatures or variances in the design of individual equipment may produce different performance data.
We therefore recommend that equipment be calibrated and/or validated on a case-by-case basis when working at the extremes of the permissible ambient temperature range.