Size ranges of E series

Drying ovens with mechanical control

With the BINDER E series drying oven, you obtain a reliable and powerful unit in a solid configuration at a very attractive price.

Advantages:

  • Reliable powerful drying oven
  • Very attractive price
  • "Made in Germany" quality

Areas of application:

  • Basic Research / Research Institutes
    Basic Research / Research Institutes
  • Human / Veterinary Medicine
    Human / Veterinary Medicine
  • Clinics / University Hospitals
    Clinics / University Hospitals
Features Customer benefits Characteristics
Reliable powerful drying oven
  • Very attractive price
  • "Made in Germany" quality
  • Long service life
  • Temperature range 60 °C to 230 °C
  • Mechanical control and adjustable ventilation flap slide
  • Hydro-mechanical thermostat
  • Timer 0 – 120 min
  • Optional independent temperature safety device class 1 (DIN 12880)
  • 2 chrome-plated racks included
  • Units are stackable
  • BINDER test certificate
E 28
Exterior dimensions
Width (mm) 580
Height (incl. feet) (mm) 402
Depth (mm) 425
Plus door handle, I-panel, connection (mm) 50
Wall clearance, rear (mm) 100
Wall clearance, side (mm) 160
Number of doors (ea.) 1
     
Interior dimensions
Width (mm) 400
Height (mm) 280
Depth (mm) 250
Interior volume (l) 28
Racks (number standard/max.) 2 / 4
Load per rack (kg) 10
Permitted total load (kg) 25
Weight (empty) (kg) 22
     
Temperature data
Temperature range approx. 5 °C above ambient temperature to (°C) 60 - 230
Temperature variation at 70 °C (± K) 3
Temperature fluctuation at 70 °C (± K) 1,5
Warm-up time 1)
at 70 °C (min.) 28
at 150 °C (min.) 36
Recovery time after doors were open for 30 sec. 1)
at 70 °C (min.) 11
at 150 °C (min.) 19
Ventilation data
Ventilation at 70 °C (x/h) 30
     
Electrical data
IP protection class acc. to EN 60529 IP 20
Voltage (± 10%) 50 / 60 Hz (V) 230
Nominal power (kW) 0,8

1) To 98% of the set value

All technical data are specified for units with standard equipment at an ambient temperature of 25 °C and a line voltage fluctuation of ±10%. The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. All figures are typical average values for series devices.
We reserve the right to alter technical specifications at any time.

E 28
Rack, chrome-plated O
Perforated shelf, stainless steel O