The BINDER CB series
CO2 incubators with additional process controls
The BINDER CB series is the premium class for all sensitive incubation applications. Even during complex cultivation experiments or individual environments under hypoxic conditions, it is easy to grow thanks to a comprehensive assortment of options and accessories.
Advantages:
- Hot air sterilization at 180 °C
- Seamless, deep-drawn inner chamber made of stainless steel
- Unique BINDER technology (patented air jacket system, controlled condensation, etc.)
Areas of application:
-
Bio Tissue Engineering -
In vitro Fertilization (IVF) -
Clinics / University Hospitals
| Features | Customer benefits | Characteristics |
|---|---|---|
| Sterilization |
|
180 °C hot air sterilization
|
| Permadry™ humidity system |
|
Controlled humidification system
|
| APT.line™ heating technology |
|
Preheating chamber with VENTAIR™ air jacket
|
| Gas distribution |
|
Gas mixing head with venturi effect
|
| Inner chamber concept |
|
Seamless, deep-drawn inner chamber
|
| CO2 measurement |
|
Single-beam infrared sensor
|
| Operating costs |
|
|
- Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results
- Temperature range from 5 °C above ambient temperature to 60 °C
- MCS controller for temperature and CO2 concentration
-
User-friendly LCD screen
- Easy-to-read menu guide
- Integrated electronic chart recorder
- Variety of options for the graphic display of process parameters
- Real-time clock
- Standard-compliant hot air sterilization at 180 °C (DIN 58947)
- VENTAIR™ Jacket System
- Drift-free infrared CO2 measurement system
- Gas mixing head
- Permadry System - condensation-free double-pan humidification system, maintains dry interior walls
- Seamless deep-drawn inner chamber made of stainless steel with integrated shelf support system (flanges)
- Electronic self-diagnostic system for errors with visual and audible alarms, as well as relay potential changeover contact for central monitoring
- Independent temperature safety device class 3.1 (DIN 12880) with optical and audible temperature alarm
- Tightly-fitted inner glass door
- RS 422 interface for communication software APT-COM™ DataControlSystem
- 2 perforated stainless steel shelves with standard equipement and O2 control
- Units are stackable with stacking adapter
- Door lock
- BINDER test certificate
| CB 53 | ||
|---|---|---|
|
|
Exterior dimensions | |
| Width (mm) | 580 | |
| Height (incl. feet) (mm) | 720 | |
| Depth (plus 55 mm for door handle and connection) (mm) | 545 | |
| Wall clearance side / rear (mm) | 50 / 100 | |
| Number of doors (ea.) | 1 | |
| Inner glass door(s) | 1 | |
|
|
Interior dimensions | |
| Width (mm) | 400 | |
| Height (mm) | 400 | |
| Depth (mm) | 332 | |
| Interior volume (l) | 53 | |
| Number of shelves (standard/max.) | 2 / 3 | |
| Dimensions shelves width x depth (mm) | 396 x 289 | |
| Load per shelf (kg) | 10 | |
| Permitted total load (kg) | 30 | |
| Weight (empty) (kg) | 60 | |
|
|
Temperature data | |
| Temperature range from 5 °C above ambient temperature to (°C) | 60 | |
| Temperature variation at 37 °C (± K) | 0,3 | |
| Temperature fluctuation at 37 °C (± K) | 0,1 | |
| Recovery time after door was opened for 30 sec. at 37 °C (min.) | 3 | |
|
|
Humidity data | |
| Humidity (avg. value) (% RH) | ≥ 95 | |
|
|
CO2 data | |
| CO2 range (vol. % CO2) | 0 - 20 | |
| Setting accuracy (Vol.-% CO2) | 0,1 | |
| Recovery time after door was opened for 30 sec | ||
| at 5.0 vol. % CO2 (min.) | 6 | |
| CO2 measurement | IR | |
| Connection hose nozzle for CO2 for hose with internal diameter (mm) | 6 | |
|
|
O2 data | |
| O2 range (vol. % O2) | 0,2 - 95 | |
| Setting accuracy (Vol.-% O2) | 0,1 | |
| Recovery time after door was opened for 30 sec | ||
| at 1.0 vol. % O2 (min.) | 25 | |
| at 5.0 vol. % O2 (min.) | 9 | |
| O2 measurement | ZrO2 | |
| Connection hose nozzle for O2 / N2 for hose with internal diameter (mm) | 6 | |
|
|
Electrical data | |
| Voltage (± 10%) 50/60 Hz (V) | 230 | |
| Nominal power (kW) | 1,0 | |
| Energy consumption at 37 °C (W) 1) | 70 |
1) These values can be used for dimensioning air condition systems.
All technical data are specified for units with standard equipment at an ambient temperature of 25 °C and a line voltage fluctuation of ±10%. The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. All figures are typical average values for series devices. We reserve the right to alter technical specifications at any time.
Silicone access port
For introducing external measuring instruments into the chamber. The access port has a diameter of 30 mm and can be sealed on both sides using a silicone plug. It can only be positioned on the back for CB 53.
APT-COM™ DataControlSystem GLP Edition
Software for GLP-compliant control, programming and documentation. Allows networking of up to 30 devices or controllers. Meets FDA 21 CFR Part 11 requirements.
Calibration certificate & validation
BINDER can significantly reduce the workload in qualifying and validating devices. Nobody knows our devices as well and has as much experience in certifications as we do.
Control for variable O2 values
For hyper- or hypoxic culture conditions. The O2 or N2 gas supply can be controlled using an additional control loop as needed. A zirconium oxide sensor (ZrO2) is used for the measurement.
Stacking adapter
Used for stacking of two CB incubators which are thermally isolated from each other. The incubators can thus be sterilized with hot air independent of each other. Available as stacking adapters, stacking frames or base frames.
| CB 53 | |
|---|---|
| Silicone access port with two silicone plugs, 30 mm, back | O |
| Lockable controller keyboard | O |
| Gas tank connection set for CO2 consisting of a gas tank pressure regulator (max. pressure 10 bar) with connection parts and 5 m hose | O |
| Gas tank connection set for O2 consisting of a gas tank pressure regulator (max. pressure 10 bar) with connection parts and 5 m hose | O |
| Gas tank connection set for N2 consisting of a gas tank pressure regulator max. pressure 10 bar) with connection parts and 5 m hose | O |
| BINDER Gas Supply Service external gas tank replacement for connecting 2 gas tanks, either CO2 or N2, with audible and visual alarms, as well as potential-free alarm output | O |
| Unit conversion from RS 422 to Ethernet interface | O |
| 4 - 20 mA analog outputs for temperature and CO2 measurements (e.g. chart recorder connection) with 6-pin DIN socket (output not adjustable) | O |
| Switched LEMO inner chamber socket (coverable) and LEMO connector (max. current carrying capacity 230 V AC - 3 A) (protection class IP 65) | O |
| Independent Intelligent Fail Safe monitoring function. Unique safety function used to prevent unnoticed deviations of the CO2 concentration from the set value. Note: Not possible in conjunction with access ports or divided inner glass door | O |
| Calibration certificate for temperature and CO2 (temperature measurement in center / CO2 measurement performed using analyzed test gas at 37 °C and 5% CO2) | O |
| Calibration certificate for optional O2 control, O2 measurement with analyzed test gas 1% O2 | O |
| Temperature measurement acc. to DIN 12880 (27 measuring points) at 37 °C or at specified temperature with measuring protocol and certificate | O |
| Manual for Primary Human Cell Culture, in English | O |
| Perforated shelf, stainless steel | O |
| Base on casters | O |
| Stacking adapter for direct thermal decoupled stacking of two C 53 CO2 incubators | O |