The BINDER KBWF series
Climate chambers with illumination and humidity
The BINDER KBWF series shines with its homogeneous light distribution with natural illumination from above. It creates consistent climate conditions – even with illumination.Advantages:
- Homogeneous light distribution
- Natural growth conditions
- Temperature, humidity & light in one unit
Areas of application:
-
Cosmetics Industry -
Plant / Insect Growth -
Packaging Industry
| Features | Customer benefits | Characteristics |
|---|---|---|
| APT.line™ climate technology |
|
APT.line™
|
| Light |
|
|
| Cooling system |
|
|
| Water supply |
|
Suitable for any water quality
|
| Humidification system |
|
|
| Standard equipment |
|
Comprehensive standards
|
| Unit design |
|
|
| Accessories and Services |
|
Comprehensive product portfolio
|
- Electronically controlled APT.line™ preheating chamber with cooling system assuring temperature accuracy and reproducible results
- Temperature range 0 °C to 70 °C (without humidity and illumination)
- Temperature range 20 °C to 60 °C (with humidity and illumination)
- Humidity range 10% to 80% RH (without illumination)
- Humidity range 10% to 75% RH (with illumination)
- 3 variable position illumination cassettes each with 5 daylight fluorescent illumination tubes
- 3 racks, stainless steel
- Patented light system for unique homogeneous light distribution over the entire usable area
- MCS controller for temperature, humidity, and lighting control with 25 storable programs of 100 sections each for a maximum of 500 program segments, for programming of day/night cycles
- Integrated electronic chart recorder
- Variety of options for the graphic display of process parameters
- Real-time clock
- Controlled humidification and dehumidification system with capacitive humidity sensor
- Independent temperature safety device class 3.1 (DIN 12880) with optical and audible temperature alarm
- Access port with silicone plug Ø 30 mm, left side
- Inner glass door
- Complete safety connection kit for water supply and drainage, including water hose, (total length 6 m)
- RS 422 interface for communication software APT-COM™ DataControlSystem
- BINDER test certificate
| KBWF 720 (E5.2) | ||
|---|---|---|
|
|
Exterior dimensions | |
| Width (mm) | 1249 | |
| Height (incl. casters) (mm) | 1924 | |
| Depth (plus 52 mm door handle, I-triangle, connection) (mm) | 920 | |
| Wall clearance, rear (mm) | 100 | |
| Wall clearance, side (mm) | 100 | |
| Steam space volume (l) | 918 | |
| Number of doors (ea.) | 2 | |
| Inner glass door(s) (ea.) | 2 | |
|
|
Interior dimensions | |
| Width (mm) | 970 | |
| Height (mm) | 1250 | |
| Depth (mm) | 576 | |
| Interior volume (l) | 698 | |
| Racks (number standard/max.) | 3 / 12 | |
| Load per rack (kg) | 45 | |
| Permitted total load (kg) | 150 | |
| Weight (empty) (kg) | 374 | |
| Variable position illumination cassettes | 3 | |
|
|
Temperature data (without humidity) | |
| Temperature range without illumination cassettes (°C) | 0 - 70 | |
| Temperature range with illumination cassettes with illumination (°C) | 10 - 60 | |
| Max. heat compensation up to 40 °C with illumination (W) | 1000 | |
|
|
Climate data (with humidity) | |
| Temperature range without illumination cassettes (°C) | 0 - 70 | |
| Temperature range with illumination cassettes with illumination (°C) | 20 - 60 | |
| Temperature fluctuation with illumination (± K) | 0,1 - 0,5 | |
| Temperature variation with illumination (± K) | 1,0 - 1,5 | |
| Humidity range without illumination cassettes (% RH) | 10 - 80 | |
| Humidity range with illumination cassettes, with illumination (% RH) | 10 - 75 | |
| Humidity fluctuation with illumination (± % RH) | ≤ 2,5 | |
|
|
Illumination data per light cassette | |
| Daylight tubes (Lux) | 14200 | |
| Daylight tubes (W/m²) | 38 | |
| Fluora® growth lamps (Lux) | 10500 | |
| Fluora® growth lamps (W/m²) | 31 | |
| Arabidopsis tubes (Lux) | 14400 | |
| Arabidopsis tubes (W/m²) | 38 | |
|
|
Electrical data | |
| IP protection class acc. to EN 60529 | IP 20 | |
| Voltage (± 10%) 50 / 60 Hz (V) | 200-240 1N~ | |
| Nominal power at 240 V (kW) | 3,5 | |
| Energy consumption at 37 °C / 75 % RH 1) (W) | 2350 | |
| Noise level approx. (dB (A)) | 53 |
Variable position illumination cassette
Temperature-humidity chart without illumination
Temperature-humidity chart with illumination
Heat compensation without illumination
Heat compensation with illumination
Light spectrum
1) These values can be used for dimensioning air condition systems.
All technical data are specified for units with standard equipment at an ambient temperature of 25 °C and a line voltage fluctuation of ±10%. The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. All figures are typical average values for series devices. We reserve the right to alter technical specifications at any time.
BINDER Data Logger Kits
The new BINDER Data Logger Kits – Makes independent recording of temperature and humidity data in the BINDER device possible. The tailored product solution contains helpful accessories: from mounting the logger to the BINDER device to cable access assistance to the sensor mount.
BINDER Pure Aqua Service
This convenient and flexible water treatment system extends maintenance intervals and is easily implemented independent of water quality.
The specific advantage: Point-of-use system with water quality display and replaceable filter cartridge.
External water supply set
External water supply set from fresh and waste water container, wiring and pump
Temperature measurement of the specimen
Additional PT 100 temperature sensor for accurate temperature measurement of the specimen and digital display of measured values. Recording of measured data via Ethernet or RS 422 interface optional.
| KBWF 720 (E5.2) | |
|---|---|
| Access ports with silicone plug, 30, 50, 100 mm | O |
| Fasteners for additional security for racks (1 set of 4) | O |
| Ethernet interface for communication software APT-COM™ DataControlSystem | O |
| External water supply set including fresh and waste water containers (20 liters each), cabling and pump | O |
| BINDER PURE AQUA SERVICE containing single-use cartridge, cabling and measuring device | O |
| Single-use cartridge for BINDER PURE AQUA SERVICE | O |
| Temperature precision measurement according to DIN 12880 and 9-point humidity measurement / factory standard with measurement log and certificate, measured at 25 °C / 60% RH or at specified values | O |
| Calibration certificate for temperature and humidity for all models. Measurement in center of chamber at 25 °C / 60% RH or at specified values | O |
| Extension to calibration certificate for temperature and humidity. Each additional measurement at an additional measuring point or set of values | O |
| Data Logger Kit TH 70: For continuous temperature and humidity recording of -40 °C to 70 °C / 0% to 100% RH. Kit includes 1 data logger, 1 attachable combined humidity/temperature sensor, 2 m extension cable and 1 fixture for mounting to the BINDER unit | O |
| Data Logger Kit T 220: For continuous temperature recording of -90 °C to 220 °C. Kit includes 1 data logger, PT 100 sensor with 2 m Teflon extension cable and 1 fixture for mounting to the BINDER unit | O |
| Data Logger Software: Configuration and evaluation software for all BINDER Data Logger Kits, incl. data cable | O |
| Rack, stainless steel | O |
| Perforated shelf, stainless steel | O |
| Reinforced rack, stainless steel, with 1 set of fasteners (1 set of 4), max. load 70 kg | O |
| Stable shelf (positioned at bottom shelf level) with additional mounting for shaker operation. Other positions possible on request | O |
| 4 - 20 mA analog outputs for temperature and humidity measurements (e.g. chart recorder connection) with 6-pin DIN socket. Outputs are adjusted automatically when adjusting the controller | O |
| Zero-voltage relay alarm outputs for temperature (±2 °C) and humidity (±5% RH), accessible via 6-pin DIN socket, with audible signal that can be switched off (maximum power rating 24 V AC/DC, 2.5 A) | O |
| FLUORA® growth fluorescent tube set for replacing the daylight fluorescent tubes | O |
| Arabidopsis fluorescent tube set for replacing the daylight fluorescent tubes | O |
| Replacement fluorescent tube set for one illumination cassette | O |
| Replacement FLUORA® illumination tube set for one illumination cassette | O |
| Replacement Arabidopsis lamp set for one illumination cassette | O |
| Door lock | O |
| Independent temperature safety device class 3.3 (DIN 12880) with optical alarm | O |