Model KMF 115 | Constant climate chambers with expanded temperature / humidity range
- Safe thanks to climatic homogeneity that far exceeds the accuracy required by ICH guidelines as well as THB (85/85), even with a fully loaded unit.
- Reliable thanks to failsafe operation without compromise. The interior and Longlife evaporator plate are made entirely from stainless steel.
- Smart, as a wide range of accessories makes it highly compatible for adaptation to specific customer requirements.
- Economical thanks to maximum effective volume which allows for 30% larger load compared to the competition.
- Temperature range: -10 °C to +100 °C
- Humidity range: 10 % RH to 98 % RH
- APT.line™ preheating chamber technology
- Humidity regulation with capacitive humidity sensor and vapor humidification
- Inner chamber made of stainless steel
- BINDER Multi Management Software APT-COM™ Basic Edition
- Intuitive touchscreen controller with time-segment and real-time programming
- Internal data logger, measured values can be read out in open format via USB
- Unit self-test for comprehensive status analysis
- Tight-sealing inner door made of safety glass (ESG)
- Avoidance of glass corrosion by special TIMELESS coating
- 1 stainless steel rack
- Access port with silicone plug, 30 mm, left
- Class 3.1 independent temperature safety device (DIN 12880) with visual and audible temperature alarm
- Computer interface: Ethernet
- Door heating
Technical specifications
Designation | KMF115-230V¹ | KMF115-240V¹ |
---|---|---|
Option model | Standard | Standard |
Order number | 9020-0341 | 9020-0342 |
Data | ||
Serie | KMF | KMF |
Performance Data Temperature | ||
Average cooling down time according to IEC 60068-3-5 [K/min] | 0.5 | 0.5 |
Average heating-up rate according to IEC 60068-3-5 [K/min] | 1.3 | 1.3 |
Max. heat compensation at 25 °C [W] | 150 | 150 |
Temperature range | -10…100 | -10…100 |
Performance Data Climate | ||
Dew point temperature range [°C] | 5…90 | 5…90 |
Humidity fluctuation depending on setpoint | ≤2,5 ± % RH | ≤2,5 ± % RH |
Humidity range | 10…98 | 10…98 |
Temperature fluctuation depending on setpoint [± K] | 0.1…0.3 | 0.1…0.3 |
Temperature range [°C] | 10…90 | 10…90 |
Temperature variation depending on setpoint [± K] | 0.2…1 | 0.2…1 |
Electrical data | ||
Nominal power [kW] | 2 | 2 |
Phase (Nominal voltage) | 1~ | 1~ |
Power frequency [Hz] | 50/60 | 50/60 |
Rated Voltage [V] | 200…230 | 200…240 |
Unit fuse [A] | 16 | 16 |
Doors | ||
Inner doors | 1 | 1 |
Unit doors | 1 | 1 |
Housing dimensions not incl. fittings and connections | ||
Depth net [mm] | 650 | 650 |
Height net [mm] | 1050 | 1050 |
Width net [mm] | 880 | 880 |
Internal Dimensions | ||
Depth [mm] | 351 | 351 |
Height [mm] | 483 | 483 |
Width [mm] | 600 | 600 |
Measures | ||
Interior volume [L] | 102 | 102 |
Load per rack [kg] | 30 | 30 |
Net weight of the unit (empty) [kg] | 128 | 128 |
Permitted load [kg] | 100 | 100 |
Wall clearance back [mm] | 100 | 100 |
Wall clearance sidewise [mm] | 100 | 100 |
Environment-specific data | ||
Energy consumption at 85 °C and 85 % RH [Wh/h] | 570 | 570 |
Sound-pressure level [dB(A)] | 52 | 52 |
Fixtures | ||
Number of shelves (std./max.) | 1/5 | 1/5 |
Performance Data Temperature | ||
Average cooling down time according to IEC 60068-3-5 [K/min] | 0.5 | 0.5 |
Average heating-up rate according to IEC 60068-3-5 [K/min] | 1.3 | 1.3 |
Max. heat compensation at 25 °C [W] | 150 | 150 |
Temperature range | -10…100 | -10…100 |
Performance Data Climate | ||
Dew point temperature range [°C] | 5…90 | 5…90 |
Humidity fluctuation depending on setpoint | ≤2,5 ± % RH | ≤2,5 ± % RH |
Humidity range | 10…98 | 10…98 |
Temperature fluctuation depending on setpoint [± K] | 0.1…0.3 | 0.1…0.3 |
Temperature range [°C] | 10…90 | 10…90 |
Temperature variation depending on setpoint [± K] | 0.2…1 | 0.2…1 |
Electrical data | ||
Nominal power [kW] | 2 | 2 |
Phase (Nominal voltage) | 1~ | 1~ |
Power frequency [Hz] | 50/60 | 50/60 |
Rated Voltage [V] | 200…230 | 200…240 |
Unit fuse [A] | 16 | 16 |
Doors | ||
Inner doors | 1 | 1 |
Unit doors | 1 | 1 |
Housing dimensions not incl. fittings and connections | ||
Depth net [in] | 25.61 | 25.61 |
Height net [in] | 41.4 | 41.4 |
Width net [in] | 34.67 | 34.67 |
Internal Dimensions | ||
Depth [in] | 13.83 | 13.83 |
Height [in] | 19.03 | 19.03 |
Width [in] | 23.64 | 23.64 |
Measures | ||
Interior volume [cu.ft.] | 3.6 | 3.6 |
Load per rack [lbs] | 66 | 66 |
Net weight of the unit (empty) [lbs] | 282 | 282 |
Permitted load [lbs] | 220 | 220 |
Wall clearance back [in] | 3.94 | 3.94 |
Wall clearance sidewise [in] | 3.94 | 3.94 |
Environment-specific data | ||
Energy consumption at 85 °C and 85 % RH [Wh/h] | 570 | 570 |
Sound-pressure level [dB(A)] | 52 | 52 |
Fixtures | ||
Number of shelves (std./max.) | 1/5 | 1/5 |