The BINDER VD series
Vacuum drying oven for non-flammable solvents
A BINDER vacuum drying oven of the VD series is impressive while gently drying with its homogeneous temperature distribution. The patented expansion shelf technology from BINDER ensures optimal heat transfer. The shelves are easy to position where required and the inner chamber of the vacuum drying oven is easy to clean.
Advantages:
- Maximum occupational safety with one-of-a-kind safety concept
- Fast, condensation-free drying processes
- Homogeneous specimen drying
Areas of application:
-
Electronics / Semiconductor Industry -
Plastics Industry -
Pharmaceuticals Industry
| Features | Customer benefits | Characteristics |
|---|---|---|
| APT.line™ |
|
APT.line™ preheating chamber technology
|
| Drying |
|
BINDER Cross-Flow Principle
|
| Safety concept |
|
Tested security
|
| Cleaning |
|
Smooth inner chamber with rounded corners
|
| Complete system |
|
Coordinated, modular system
|
| Accessories and Services |
|
Comprehensive product portfolio
|
- Electronically controlled APT.line™ preheating chamber with 2 expansion racks assuring temperature accuracy and reproducible results
- Temperature range from 15 °C (27 °F) above ambient temperature to 200 °C (392 °F)
- MP controller with 2 programs with 10 sections each or switchable to 1 program with 20 sections
- Integrated weekly program timer with real-time function
- Digital temperature setting with an accuracy of one degree
- Elapsed time indicator
- Precision-adjustable ventilation valve
- Precision-adjustable inert gas valve with Cross-Flow-Technology
- All electrical components are decoupled from the inner chamber
- Spring-mounted safety glass panel with shatter protection
- Independent adjustable temperature safety device class 2 (DIN 12880), with visual temperature alarm
- Measuring port DN 16 in rear panel
- Analog pressure gauge (displays pressure difference between the inner chamber and ambient pressure)
- Electro polished inner chamber, suction and ventilation tubes, pressure container, expansion racks, and ball valve are made of stainless steel
- Door gasket made of tempered silicone
- 2 x 24 V DC (max 0.4 A) switching outputs, switched via 2 control contacts in the program editor
- RS 422 interface for communication software APT-COM™ DataControlSystem
- 2 patented, flexible aluminum expansion racks
- Also available as complete system with module and vacuum pump. Features:
- Reduced noise level
- Practical working height
- Coordinated system
- BINDER test confirmation
| VD 53 | ||
|---|---|---|
|
|
Exterior dimensions | |
| Width (mm/inch) | 635 / 25.0 | |
| Height (incl. feet) (mm/inch) | 775 / 30.5 | |
| Height with optional vacuum module (mm/inch) | 625 / 24.6 | |
| Height with optional vacuum module (mm/inch) | 1400 / 55.1 | |
| Depth (mm/inch) | 550 / 21.7 | |
| Plus door handle, connection (mm/inch) | 100 / 3.9 | |
| Wall clearance, rear (mm/inch) | 100 / 3.9 | |
| Wall clearance, side (mm/inch) | 135 / 5.3 | |
|
|
Interior dimensions | |
| Width (mm/inch) | 400 / 15.8 | |
| Height (mm/inch) | 400 / 15.8 | |
| Depth (mm/inch) | 340 / 13.4 | |
| Interior volume (l/cu.ft.) | 53 / 1.9 | |
| Expansion racks (aluminum) (number standard/max.) | 2 / 5 | |
| Distance between the racks (mm/inch) | 62 / 2.4 | |
| Usable space per rack (width x depth) (mm/inch) | 349 x 320 / 13.7 x 12.6 | |
| Load per shelf (kg/lbs.) | 20 / 44 | |
| Permitted total load (kg/lbs.) | 45 / 99 | |
| Weight (empty) (kg/lbs.) | 95 / 209 | |
|
|
Temperature range | |
| Temperature range approx. 15 °C (27 °F) above ambient temperature to (°C/°F) | 200 / 392 | |
| Temperature variation 1) | ||
| 100 °C (212 °F) (± K) | 2 | |
| 200 °C (392 °F) (± K) | 4.5 | |
| Temperature fluctuation (± K) 1) | 0.1 | |
| Warm-up time 1), 2) | ||
| to 100 °C (212 °F) (min.) | 80 | |
| to 200 °C (392 °F) (min.) | 115 | |
| Vacuum connection with small flange (DN mm/inch) | 16 / 0.63 | |
| Measuring access port with small flange (DN mm/inch) | 16 / 0.63 | |
| Inert gas connection with flow limiter thread (RP) | 3 / 8 | |
| Permitted end vacuum (mbar / inchHg) | 0.01 / 0.000295 | |
| Leak rate max. (bar/h / inchHg/h) | 0.01 / 0.295 | |
|
|
Electrical data | |
| IP protection class acc. to EN 60529 | IP 20 | |
| Voltage (± 10%) 50 / 60 Hz | 230 | |
| Voltage (± 10%) 60 Hz | 115 | |
| Nominal power (kW) | 1.2 | |
| Energy consumption | ||
| at 100 °C (212 °F) (W) | 150 | |
| at 200 °C (392 °F) (W) | 445 |
1) Value with aluminum racks
2) To 98% of the set value
All technical data are specified for units with standard equipment at an ambient temperature of 25 °C (77 °F) and a line voltage fluctuation of ±10%. These average values have been determined according to the BINDER factory standard, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. Differing ambient temperatures and production-related device-specific variances can lead to varying technical data.
Therefore, we recommend individual customer-specific calibration or validation for applications on the limit of the permitted ambient temperature range.
Measuring port
Vacuum-tight access port into the device for measuring lines (9-pin).
Specimen temperature display
Uses PT 100 sensor and digital temperature display.
Calibration certificate
Measurement in center at predefined test value. Expandable with additional measuring points or test values based on your requirements.
| VD 53 | |
|---|---|
| Connection kit consists of: clamping ring, aluminum centering ring, small flange with hose nozzle, (2.5 m/8.2 ft.) rubber hose, and 2 hose clamps | O |
| Measuring port for vacuum-tight access port of measuring lines into the device (9-pin) | O |
| Factory calibration certificate, measurement in center of chamber at 100 °C (212 °F) or at specified testing temperature | O |
| Extension to factory calibration certificate. Each additional measurement at additional measuring point or testing temperature | O |
| Factory calibration certificate for digital pressure gauge. Measurement at 100 mbar or at specified test pressure (range 20-900 mbar) | O |
| Extension to factory calibration certificate for digital pressure gauge. Each additional measurement with additional test pressure (range 20-900 mbar) | O |
| Temperature measurement of the specimen with flexible PT 100 sensor and digital specimen temperature display | O |
| Factory calibration certificate for digital specimen temperature display. Measurement at 100 °C or at specified testing temperature | O |
| VP 1.1 chemical membrane pump (nominal air flow 2.0 m³/hour, final pressure 7 mbar), with separator and emission condenser (230 V 1N ~ 50 / 60 Hz) | O |
| VP 2.1 chemical membrane pump (nominal air flow 3.4 m³/hour, final pressure 7 mbar), with separator and emission condenser (230 V 1N ~ 50 / 60 Hz) | O |
| VP 3.1 speed-controlled chemical membrane pump (nominal air flow 4.6 m³/hour, final pressure 1.5 mbar), with separator and emission condenser (230 V 1N ~ 50 / 60 Hz) | O |
| Expansion racks, aluminum | O |
| Expansion racks, stainless steel | O |
| Exchange and calibration to stainless steel expansion racks, standard equipment aluminum expansion racks are replaced | O |
| Door gasket, FKM (Viton) | O |
| Digital pressure display, measuring range from 1 mbar to atm. pressure, display accuracy 1 mbar | O |
| Vacuum module for installation of vacuum pumps, with switchable power socket (ON/OFF) (230 V / 16 A) | O |
| Vacuum module with chemical membrane pump VP 1.1 (nominal air flow 2.0 m³/hour, final pressure 7 mbar), with separator and emission condenser, including all necessary vacuum connection parts 230 V 1N ~ 50 / 60 Hz | O |
| Vacuum module with chemical membrane pump VP 2.1 (nominal air flow 3.4 m³/hour, final pressure 7 mbar), with separator and emission condenser, including all necessary vacuum connection parts 230 V 1N ~ 50 / 60 Hz | O |
| Vacuum module with speed-controlled chemical membrane pump VP 3.1 (nominal air flow 4.6 m³/hour, final pressure 1.5 mbar), with separator and emission condenser, including all necessary vacuum connection parts, as well as digital vacuum controller (measuring precision ≤1 mbar) 230 V 1N ~ 50 / 60 Hz | O |