The BINDER MDL series
Safety drying oven with expended temperature range
The BINDER MDL 115 with enhanced control function works at temperatures up to 350 °C (662 °F). This safety drying oven meets all EN 1539 requirements 1539 and is perfectly suited for high-temperature testing such as in the coil coating area.
Advantages:
- Defined solvent quantity according to EN 1539
- Wide temperature range up to 350 °C (662 °F)
Areas of application:
-
Chemicals -
Surface Technology
| Features | Customer benefits | Characteristics |
|---|---|---|
| APT.line™ |
|
APT.line™ preheating chamber for maximum precision
|
| Safety concept |
|
|
| Performance feature |
|
|
| Inner chamber concept |
|
|
| Controller |
|
|
| Quality |
|
|
| Unit options and service |
|
Comprehensive product portfolio:
|
- Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results
- Temperature range from 5 °C (9 °F) above ambient temperature to 350 °C (662 °F)
- All safety features met according to EN 1539
- Heating load 9.0 kW
- Door gasket made of of high temperature resistant silicone
- Rear exhaust connector Ø 100 mm (3.9 inch)
- MCS controller with 25 storable programs of 100 sections each for a maximum of 500 program segments
-
User-friendly LCD screen
- Easy-to-read menu guide
- Integrated electronic chart recorder
- Variety of options for the graphic display of process parameters
- Real-time clock
- Replacable fresh-air filter cartridge, class F6 (EU6-fine particle filter for particle sizes between 1 ... 10 µm)
- Independent adjustable temperature safety device class 2 (DIN 12880), with acoustic and visual temperature alarm
- Fresh-air monitoring with audible and visual alarm and automatic deactivation of heating
- RS 422 interface for communication software APT-COM™ DataControlSystem
- 2 chrome-plated racks included
- BINDER test confirmation
| MDL 115 | ||
|---|---|---|
|
|
Exterior dimensions | |
| Width (mm/inch) | 834 / 32.83 | |
| Height (incl. feet) (mm/inch) | 800 / 31.5 | |
| Depth (plus door handle 50 mm (2 inch)) (mm/inch) | 685 / 27.0 | |
| Wall clearance, rear (mm/inch) | 100 / 3.9 | |
| Wall clearance, side (mm/inch) | 100 / 3.9 | |
| Exhaust duct (outer Ø mm) | 100 / 3.9 | |
| Steam space volume (l/cu.ft.) | 156 / 5.5 | |
|
|
Interior dimensions | |
| Width (mm/inch) | 600 / 23.6 | |
| Height (mm/inch) | 435 / 17.1 | |
| Depth (mm/inch) | 435 / 17.1 | |
| Interior volume (l/cu.ft.) | 115 / 4.1 | |
| Racks (number standard/max.) | 2 / 5 | |
| Load per rack (kg/lbs.) | 20 / 44 | |
| Permitted total load (kg/lbs.) | 50 / 110 | |
| Weight (empty) (kg/lbs.) | 90 / 199 | |
|
|
Temperature range | |
| Temperature range approx. 5 °C (9 °F) above ambient temperature to (°C/°F) | 350 / 662 | |
| Temperature variation | ||
| at 70 °C (158 °F) (± K) | 2 | |
| at 150 °C (302 °F) (± K) | 3.4 | |
| at 300 °C (572 °F) (± K) | 7 | |
| Temperature variation, extension in the door | ||
| at 70 °C (158 °F) (± K) | 2 | |
| at 150 °C (302 °F) (± K) | 3 | |
| at 300 °C (572 °F) (± K) | 8 | |
| Temperature fluctuation (± K) | 0.5 | |
| Warm-up time from 2) | ||
| to 70 °C (158 °F) (min.) | 3.5 | |
| to 150 °C (302 °F) (min.) | 6 | |
| to 300 °C (572 °F) (min.) | 10 | |
| Recovery time after doors were open for 30 sec. 2) | ||
| at 70 °C (158 °F) (min.) | 0.5 | |
| at 150 °C (302 °F) (min.) | 2 | |
| at 300 °C (572 °F) (min.) | 4 | |
| Recovery time after door was open for 30 sec. 2), extension in the door | ||
| at 70 °C (158 °F) (min.) | 0.5 | |
| at 150 °C (302 °F) (min.) | 1 | |
| at 300 °C (572 °F) (min.) | 2 | |
| Ventilation (approx. x/min.) | 3 | |
| Air circulation (approx. x/min.) | 40 | |
| Exhaust air volume flow (approx. l/min. m³/h) | 400 (24.0) | |
| Air flow velocity (m/sec) | 0.8 - 1.2 | |
| Maximum amount of solvent (g)(lbs.) (at T-180 °C, M-100 g/mol, U-40 g/m3, K = 0,5) | 6.65/0.014 | |
|
|
Electrical data | |
| IP protection class acc. to EN 60529 | IP 33 | |
| Voltage (± 10%) 50 / 60 Hz (V) | 400 | |
| Nominal power (kW) | 9.0 | |
| Energy consumption at 150 °C (302 °F)(W) | 1130 |
1)
T = drying temperature
M = molecular mass
U = lower explosion limit
K = solvent vapor concentration as percentage at lower explosion limit
2) To 98% of the set value
All technical data are specified for units with standard equipment at an ambient temperature of 25 °C (77 °F) and a line voltage fluctuation of ±10%. The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. All figures are typical average values for series devices. We reserve the right to alter technical specifications at any time.
Door lock
Prevents unauthorized access to the process sequences in the chamber.
Extension for coil coating applications
Extension in the door for quick loading for coil coating / hot air short cycle applications.
Calibration certificate & validation
BINDER can significantly reduce the workload in qualifying and validating devices. Nobody knows our devices as well and has as much experience in certifications as we do.
| MDL 115 | |
|---|---|
| Rack,chrome-plated or stainless steel | O |
| Shelf, perforated, stainless steel | O |
| Locking door handle with key | O |
| Keyboard lock | O |
| Base on castors | O |
| Door flap for coil-coating tests | O |
| Replacement air filter (class F6/EU6 – fine particle filter for particle sizes between 1 µm and 10 µm), 100 × 520 × 22 mm (3.94 x 20.47 x 0.87 inch), with aluminum frame | O |
| Additional measuring channel for digital display of specimen temperature, with clip temperature sensor. Measured data recorded via RS 422 interface | O |
| Temperature measurement acc. to DIN 12880 (27 measuring points) at 150 °C or at specified temperature with measuring protocol and certificate | O |
| Factory calibration certificate. Measurement in center of chamber at 150 °C (302 °F) or at specified testing temperature | O |
| Extension to factory calibration certificate. Each additional measurement at an additional measuring point or temperature | O |