The BINDER E series
Size ranges of E series
Heating oven with mechanical control
With a BINDER E series heating oven, you obtain a reliable and powerful unit in a solid configuration at a very attractive price. This heating oven is preferred for tasks in basic research, as well as in human and veterinary medicine.Advantages:
- Reliable powerful heating oven
- Very attractive price
- Quality "Made in Germany"
Areas of application:
-
Basic Research / Research Institutes -
Human / Veterinary Medicine -
Clinics / University Hospitals
| Features | Customer benefits | Characteristics |
|---|---|---|
| Reliable powerful heating oven |
|
|
- Temperature range 60 °C (140 °F) up to 230 °C
- Robust, space-saving small unit with hydraulic-mechanical control and adjustable ventilation
- Hydro-mechanical thermostat
- Timer 0 to 120 min
- Optional independent temperature safety device class 1 (DIN 12880)
- 2 chrome-plated racks included
- Units are stackable
- BINDER technical specification conformance certificate
| E 28 | ||
|---|---|---|
|
|
Exterior dimensions | |
| Width (mm/inch) | 580 / 22.8 | |
| Height (incl. feet) (mm/inch) | 405 / 15.9 | |
| Depth (mm/inch) | 425 / 16.7 | |
| Plus door handle, I-panel, connection (mm/inch) | 50 / 2.0 | |
| Wall clearance, rear (mm/inch) | 100 / 3.9 | |
| Wall clearance, side (mm/inch) | 160 / 6.3 | |
| Number of doors (ea.) | 1 | |
|
|
Interior dimensions | |
| Width (mm/inch) | 400 / 15.8 | |
| Height (mm/inch) | 280 / 11.0 | |
| Depth (mm/inch) | 250 / 9.8 | |
| Interior volume (l/cu.ft.) | 28 / 1.0 | |
| Racks (number standard/max.) | 2 / 4 | |
| Load per rack (kg/lbs.) | 10 / 22 | |
| Permitted total load (kg/lbs.) | 25 / 55 | |
| Weight (empty) (kg/lbs.) | 22 / 49 | |
|
|
Temperature data | |
| Temperature range approx. 5 °C (41 °F) above ambient temperature to (°C/°F) | 60 - 230 / 140 - 446 | |
| Temperature variation at 70 °C (158 °F) (± K) | 3 | |
| Temperature fluctuation at 70 °C (158 °F) (± K) | 1.5 | |
| Warm-up time 1) | ||
| at 70 °C (158 °F) (min.) | 28 | |
| at 150 °C (302 °F) (min.) | 36 | |
| Recovery time after doors were open for 30 sec. 1) | ||
| at 70 °C (158 °F) (min.) | 11 | |
| at 150 °C (302 °F) (min.) | 19 | |
| Ventilation data | ||
| Ventilation at 70 °C (158 °F) (x/h) | 30 | |
|
|
Electrical data | |
| IP protection class acc. to EN 60529 | IP 20 | |
| Voltage (± 10%) 50 / 60 Hz (V) | 230 | |
| Voltage (±10 %) 60 Hz (V) | 115 | |
| Nominal power (kW) | 0.8 |
1) To 98% of the set value
All technical data are specified for units with standard equipment at an ambient temperature of 25 °C (77 °F) and a line voltage fluctuation of ±10%. The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. All figures are typical average values for series devices.
We reserve the right to alter technical specifications at any time.
| E 28 | |
|---|---|
| Rack, chrome-plated | O |
| Shelf, perforated, stainless steel | O |