The BINDER MK series
Environmental simulation chamber for cyclical temperature
The BINDER environmental simulation chamber of the MK series is suitable for heat or cold testing between -40 °C and 180 °C (-40 °F and 356 °F). The APT.line™ preheating chamber technology uniquely simulates a natural environment. For cyclical temperature testing, this environmental simulation chamber is a smart alternative to complex individual solutions.
Advantages:
- State-of-the-art reliability
- User-friendly chamber interior
- Comprehensive standard equipment
Areas of application:
-
Automotive -
Electronics / Semiconductor Industry -
Plastics Industry
| Features | Customer benefits | Characteristics |
|---|---|---|
| APT.line™ climate technology |
|
APT.line™
|
| Standard equipment |
|
Well equipped
|
| Unit design |
|
Good use of space
|
| Production |
|
|
| Accessories and Services |
|
Comprehensive product portfolio
|
- Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results
- Temperature range of -40 °C to 180 °C (-40 °F to 356 °F)
-
MCS controller with 25 storable programs of 100 sections each for a maximum of 500 program segments
- User friendly LCD screen
- Easy-to-read menu guide
- Integrated electronic chart recorder
- Variety of options for the graphic display of process parameters
- Real time clock
- Adjustable ramp function via program editor
- Access port Ø 80 mm (3.15 inch), top
- Heated viewing window with LED interior lighting
- Temperature safety device class 2 (DIN 12880) with visual and audible temperature alarm
- RS 422 interface for communication software APT-COM™ DataControlSystem
- 1 stainless steel rack
- BINDER technical specification conformance certificate
- BINDER Communication software APT-COM™ 3 Basic Edition
| MK 53 (E2.1) | ||
|---|---|---|
|
|
Exterior dimensions | |
| Width (incl. 80 mm (3.15 inch) access port with plug) (mm/inch) | 745 / 29.3 | |
| Height (incl. feet) (mm/inch) | 1245 / 49.0 | |
| Depth (incl. 55 mm (2.17 inch) I-triangle, door handle) (mm/inch) | 795 / 31.3 | |
| Wall clearance, side (mm/inch) | 160 / 6.3 | |
| Wall clearance, rear (mm/inch) | 100 / 3.9 | |
| Viewing window width (mm/inch) | 280 / 11.0 | |
| Viewing window height (mm/inch) | 280 / 11.0 | |
| Number of doors (ea.) | 1 | |
|
|
Interior dimensions | |
| Width (mm/inch) | 402 / 15.8 | |
| Height (mm/inch) | 402 / 15.8 | |
| Depth (mm/inch) | 330 / 13.0 | |
| Interior volume (l/cu.ft.) | 53 / 1.9 | |
| Shelves (number standard/max.) | 1 / 5 | |
| Load per shelf (kg/lbs.) | 15 / 33 | |
| Permitted total load (kg/lbs.) | 40 / 88 | |
| Weight (empty) (kg/lbs.) | 150 / 331 | |
|
|
Temperature data | |
| Temperature range (°C/°F) | -40 - 180 / -40 - 356 | |
| Temperature variation (±K) | 0.4 - 2.0 | |
| Temperature fluctuation (±K) | 0.1 - 0.5 | |
| Recovery time after door was open for 30 sec. | ||
| at -10 °C (14 °F) (min.) | 5 | |
| at 70 °C (158 °F) (min.) | 1 | |
| at 150 °C (302 °F) (min.) | 5 | |
| Mean warm-up rate acc. to factory standard (°C (°F) /min.) -40 °C to 180 °C | 4.6 | |
| Mean cooling rate acc. factory standard (°C (°F) /min.) 180 °C to -40 °C | 4.1 | |
| Heat compensation, max. (W) | 500 | |
|
|
Electrical data | |
| IP protection class acc. to EN 50529 | IP 20 | |
| Voltage (± 10 %) 50 Hz (V) | 230, 1 N ~ | |
| Nominal power (kW) | 2.6 | |
| Energy consumption at 20 °C (68 °F) (kW) 1) | 1.02 | |
| Noise level (approx. dB(A)) | 59 |
1) These values can be used for dimensioning air condition systems.
All technical data are specified for units with standard equipment at an ambient temperature of 25 °C (77 °F)and a line voltage fluctuation of ±10%. The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. All figures are typical average values for series devices. We reserve the right to alter technical specifications at any time.
BINDER Data Logger Kits
The new BINDER Data Logger Kits – Makes independent recording of temperature data in the BINDER device possible. The tailored product solution contains helpful accessories: from mounting the logger to the BINDER device to cable access assistance to the sensor mount
Access port
With silicone plugs for introducing external measuring instruments into the chamber, access ports with 30, 50, 80 mm (1.18, 1.97, 3.15 inch) diameters
APT-COM™ DataControlSystem GLP Edition
Software for GLP-compliant control, programming and documentation. Allows networking of up to 30 devices or controllers. Meets FDA 21 CFR Part 11 requirements
Analog outputs
For temperature or temperature and humidity 4 - 20 mA with 6-pin DIN socket (output cannot be adjusted)
Calibration certificate + validation
BINDER can significantly reduce the workload in qualifying devices. Nobody knows our devices as well and has as much experience in certifications as we do
| MK 53 (E2.1) | |
|---|---|
| Access port with silicone plug 30, 50, 80 mm (1.18, 1.97, 3.15 inch) | O |
| Securing elements for additional fastening of racks (1 set of 4 pieces) | O |
| Keyboard lock | O |
| Analog temperature output, 4-20 mA, with 6-pin DIN socket (output not adjustable) | O |
| Factory calibration certificate. Measurement in center of chamber at 150 °C (302 °F) or at specified testing temperature | O |
| Extension to factory calibration certificate. Each additional measurement at an additional measuring point or temperature | O |
| Additional measuring channel for digital display of specimen temperature, with flexible PT 100 temperature sensor. Measuring data recorded through device interface | O |
| Data Logger Kit T 220: For the continuous temperature recording of -90 °C to 220 °C (-130 °F - 428 °F). Kit includes 1 data logger, PT 100 sensor with 2 m Teflon extension cable and 1 fixture for the connection at the BINDER unit | O |
| Data Logger Software: Configuration und evaluation software for all BINDER Data Logger Kits, incl. data cable (RS 232) | O |
| Rack, stainless steel | O |
| Shelf, perforated, stainless steel | O |
| Locking door handle with key | O |