The BINDER KBF P series
Constant climate chamber with ICH-compliant illumination
The BINDER KBF P constant climate chamber, equipped with ICH-compliant illumination, is the expert for photostability testing and provides clear test results according to ICH guideline Q1B, Option 2. At the same time, the variable position illumination cassettes with BINDER Q1B Synergy™ Light ensure homogeneous lighting conditions.
Advantages:
- Long-term stable test conditions
- Independent of the water supply
- one-stop solution: Temperature/humidity/light simulation in one chamber
Areas of application:
-
Food / Beverage -
Pharmaceuticals Industry
| Features | Customer benefits | Characteristics |
|---|---|---|
| APT.line™ climate technology |
|
APT.line™
|
| Photostability tests |
|
|
| Water supply |
|
|
| Humidification system |
|
|
| Cooling system |
|
|
| Standard equipment |
|
Comprehensive standards
|
| Controller |
|
MCS screen controller
|
| Qualification |
|
More than 10 years of experience with:
|
| Accessories and Services |
|
|
- Electronically controlled APT.line™ preheating chamber and refrigerating system assuring temperature accuracy and reproducible results
- 3 variable positioned illumination cassettes with BINDER Q1B Synergy Light™ for unique homogenous light distribution
- Temperature range 0 °C to 70 °C (32 °F to 158 °F) (without illumination and humidity)
- Temperature range 20 °C to 60 °C (68 °F to 140 °F) (with illumination and humidity)
- Humidity range 10 % to 80 % RH (without illumination)
- Humidity range 10 % to 75 % RH (with illumination)
-
MCS controller with 25 storable programs of 100 sections each for a maximum of 500 program segments
- User friendly LCD screen
- Easy-to-read menu guide
- Integrated electronic chart recorder
- Variety of options for the graphic display of process parameters
- Real time clock
- Electronically controlled humidification and dehumidification system with capacitive humidity sensor suitable for stability tests according to ICH guideline Q1A (R2)
- Inner glass door with sealing
- Independent adjustable temperature safety device class 3.1 (DIN 12880), with visual and audible temperature alarm
- Access port with silicone plug Ø 30 mm (1.2 inch), left side
- Safety connection kit for water supply and drainage, including water hose, total length 6 m (19.7 ft.)
- Ethernet interface for APT-COM™ DataControlSystem software
- 3 stainless steel racks
- BINDER technical specification conformance certificate
| KBF P 720 (E5.2) | ||
|---|---|---|
|
|
Exterior dimensions | |
| Width (mm/inch) | 1255 / 49.4 | |
| Height (incl. casters) (mm/inch) | 1925 / 75.8 | |
| Depth (incl. door handle, I-triangle, connection 80 mm (3.15 inch)) (mm/inch) | 970 / 38.2 | |
| Wall clearance, rear (minimum) (mm/inch) | 100 / 3.9 | |
| Wall clearance, side (minimum) (mm/inch) | 200 / 7.9 | |
| Steam space volume (l/cu.ft.) | 918 / 32.4 | |
| Number of doors (ea.) | 2 | |
| Number of inner glass doors | 2 | |
|
|
Interior dimensions | |
| Width (mm/inch) | 973 / 38.3 | |
| Height (mm/inch) | 1250 / 49.2 | |
| Depth (mm/inch) | 576 / 22.7 | |
| Interior volume (l/cu.ft) | 700 / 24.7 | |
| Racks (number standard/max.) | 3 / 12 | |
| Load per rack (kg/lbs.) | 45 / 99 | |
| Permitted total load (kg/lbs.) | 150 / 331 | |
| Weight (empty) (kg/lbs.) | 374 / 826 | |
|
|
Temperature data (without humidity) | |
| without illumination cassettes (°C/°F) | 0 - 70 / 32 - 158 | |
| with illumination cassettes, with illumination (°C/°F) | 20 - 60 / 68 - 140 | |
| Max. heat compensation up to 40 °C (104 °F) with illumination (W) | 1000 | |
|
|
Climate data (with humidity) | |
| Temperature range without illumination cassettes (°C/°F) | 0 - 70 / 32 - 158 | |
| Temperature range with illumination cassettes (°C/°F) | 20 - 60 / 68 - 140 | |
| Temperature variation with illumination | ||
| at 25 °C (77 °F) and 60 % RH (± K) | 1.2 | |
| at 40 °C (104 °F) and 75 % RH (± K) | 1.2 | |
| Temperature fluctuation with illumination | ||
| at 25 °C (77 °F) and 60 % RH (± K) | 0.2 | |
| at 40 °C (104 °F) and 75 % RH (± K) | 0.2 | |
| Humidity range without illumination cassettes (% RH) | 10 - 80 | |
| Humidity range with illumination cassettes, with illumination (% RH) | 10 - 75 | |
| Humidity fluctuation with illumination | ||
| at 25 °C (77 °F) and 60 % RH (± % RH) | 2 | |
| at 40 °C (104 °F) and 75 % RH (± % RH) | 2 | |
| Recovery time after doors were open for 30 sec. | ||
| at 25 °C (77 °F) and 60 % RH with illumination (min.) | 1 | |
| at 40 °C (104 °F) and 75 % RH with illumination (min.) | 5 | |
| Illumination data per light cassette | ||
| ICH-compliant illumination device for photostability testing (Lux) / (UVA W/m²) | 8750 / 2.0 | |
|
|
Electrical data | |
| IP protection class acc. to EN 60529 | IP 20 | |
| Nominal voltage (± 10 %) 50 Hz (V) | 200 - 240 | |
| Nominal voltage (± 10 %) 60 Hz (V) | 208 - 240 | |
| Nominal power at 240 V (kW) | 3.5 | |
| Energy consumption at 40 °C (104 °F) / 75 % RH 1) (W) | 2350 | |
| Noise level approx. (dB (A)) | 53 |
Variable position illumination cassette
Temperature-humidity chart without illumination
Temperature-humidity chart with illumination
Heat compensation without illumination
Heat compensation with illumination
Light spectrum
1) These values can be used for dimensioning air condition systems.
All technical data are specified for units with standard equipment at an ambient temperature of 25 °C (77 °F) and a line voltage fluctuation of ±10%. The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. All figures are typical average values for series devices. We reserve the right to alter technical specifications at any time.
BINDER Pure Aqua Service
This convenient and flexible water treatment system extends maintenance intervals and is easily implemented independent of water quality.
The specific advantage: Point-of-use system with water quality display and replaceable filter cartridge
BINDER Data Logger Kits
The new BINDER Data Logger Kits – Makes independent recording of temperature and humidity data in the BINDER device possible. The tailored product solution contains helpful accessories: from mounting the logger to the BINDER device to cable access assistance to the sensor mount
External water supply set
External water supply set from fresh and waste water container, wiring and pump
Access port
With silicone plugs for introducing external measuring instruments into the chamber, access ports with 30, 50 or 100 mm (1.18, 1.97, 3.94 inch) diameters
APT-COM™ DataControlSystem GLP Edition
Software for GLP-compliant control, programming and documentation. Allows networking of up to 30 devices or controllers. Meets FDA 21 CFR Part 11 requirements
Calibration certificate & validation
BINDER can significantly reduce the workload in qualifying and validating devices. Nobody knows our devices as well and has as much experience in certifications as we do
Temperature measurement of the specimen
Additional PT 100 temperature sensor for accurate temperature measurement of the specimen and digital display of measured values. Recording of measured data via Ethernet or RS 422 interface optional
| KBF P 720 (E5.2) | |
|---|---|
| Access ports with silicone plug 30, 50, 100 mm (1.18, 1.97, 3,94 inch) | O |
| Securing elements for additional fastening of racks (1 set of 4 pieces) | O |
| Keyboard lock | O |
| RS 422 interface for communication software APTCOM™ DataControlSystem | O |
| Additional PT 100 temperature sensor, flexibly installed, with external connection, including DIN connector (6-pin) | O |
| Factory calibration certificate for temperature and humidity. Measurement in center of chamber at 25 °C (77 °F) / 60 % RH or at specified values | O |
| Extension to factory calibration certificate for temperature and humidity. Each additional measurement at an additional measuring point or set of values | O |
| Temperature precision measurement according to DIN 12880 and 9-point humidity measurement / factory standard with measurement log and certificate, measured at (25 °C /77 °F) / 60 % RH or at specified values | O |
| External fresh water supply set consits of fresh and waste water tank with 20 l (0.7 cu.ft.) each, cabling and pump | O |
| BINDER PURE AQUA SERVICE consisting of disposible cartridge, hose set and measuring unit | O |
| Disposable cartridge for BINDER PURE AQUA SERVICE | O |
| External chart recorder for temperature and humidity | O |
| Data Logger Kit TH 70: For the continuous temperature and humidity recording of -40 °C (-40 °F) to 70 °C (158 °F) / 0 % to 100 % RH. Kit includes 1 data logger, 1 attachable combined humidity/temperature sensor with 2 m extension cable and 1 fixture for the connection at the BINDER unit | O |
| Data Logger Kit TH 70/70: With two attachable combined sensors. One for the continuous temperature and humidity recording of -40 °C (-40 °F) to 70 °C (158 °F) / 0 % to 100 % RH. Second one at the data logger for recording the environmental conditions. Kit includes 1 data logger, 2 attachable combined humidity/temperature sensors with 2 m extension cable and 1 fixture for the connection at the BINDER unit | O |
| Data Logger Kit T 220: For the continuous temperature recording of -90 °C (-130 °F) to 220 °C (428 °F). Kit includes 1 data logger, PT 100 sensor with 2 m Teflon extension cable and 1 fixture for the connection at the BINDER unit | O |
| Data Logger Software: Configuration und evaluation software for all BINDER Data Logger Kits, incl. data cable (RS 232) | O |
| Rack, stainless steel | O |
| Reinforced rack, stainless steel, with 1 set of securing elements (1 set of 4 pieces), max. load 70 kg (154 lbs.) | O |
| Shelf, perforated, stainless steel | O |
| Temperature safety device, class 3.3 (DIN 12880) with visual alarm | O |
| 4-20 mA analog output for temperature and humidity measurements (e.g. chart recorder connection), with 6-pin DIN socket. Outputs are adjusted automatically as the controller is adjusted | O |
| Zero-voltage relay alarm outputs for temperature (±2 °C) and humidity (±5 % RH), accessible via 6-pin DIN socket, with audible signal that can be switched off (maximum power rating 24 V AC/DC, 2.5 A) | O |
| Locking door handle with key | O |
| Replacement set illumination tubes for one illumination cassette, including 3 cool white fluorescent tubes and 2 BINDER Q1B Synergy Light™ fluorescent tubes | O |