Vacuum drying ovens from the APT.line™
VD 23
Vacuum drying oven for non-flammable solvents
The VD series offers safe drying with homogenous temperature distribution, thanks to its APT.line™ air jacket system. Optimum heat transfer through patented expansion rack technology; the racks can be positioned as needed and are easy to clean. In addition to these many features, the units have an individually programmable controller.
- Electronically controlled APT.line™ preheating chamber technology with 2 patented expansion racks assures temperature accuracy and reproducible results
- Temperature range 15 °C (27 °F) above ambient temperature up to 200 °C (392 °F), optional up to 250 °C (482 °F) - not available for 115 V units
- MP controller with 2 programs with 10 sections each, or alternatively 1 program with 20 sections
- Integrated week program timer with real time function
- Digital temperature setting with an accuracy of one degree
- Elapsed time indicator
- Precision-adjustable ventilation valve (At VD 23 the standard inert gas connection is also used as ventilation valve.)
- Precision-adjustable inert gas valve with Cross-Flow-Technology
- Spring-mounted shatter proof safety glass panel
- All electrical componets are decoupled from the interior chamber
- Spring mounted safety glass panel with shatter protection
- Independent adjustable temperature safety device class 2 (DIN 12880), with visual temperature alarm
- Measuring port DN 16
- Analog pressure gauge (display pressure difference between the inner chamber and the ambient pressure)
- Electro polished inner chamber, suction and ventilation tubes, pressure container, expansion racks and ball valve are made of stainless steel
- Door gasket made of tempered silicone
- 2 x 24 V DC (max 0.4 A) switching outputs, switched via 2 control contact of the program editor
- RS 422 interface for use with optional GMP/GLP and FDA guideline 21 CFR Part 11 compliant APT-COM™ DataControlSystem software
- 2 patented, flexible aluminum expansion racks
- Available as complete system, with module and vacuum pump.
- Features:
- Reduced sound level
- Practical working height
- Well balanced system
- BINDER test certificate
|
1) value with aluminium racks
2) up to 98 % of the set value
All technical data are specified for units with standard equipment at an ambient temperature of 25 °C and a voltage fluctuation of ±10 %. The temperature data are determinated in accordance to factory standard following DIN 12880 respecting the recommended wall clearances of 10 % of the height, width and depth of the inner chamber. All indications are average values, typical for units produced in series. We reserve the right to alter technical specifications at all times.
-
Measuring port
Vacuum-tight bushing for instrument connection (9-pin) into the unit.
-
Specimen temperature display
Using a PT 100 sensor with digital temperature display.
-
BINDER Service
This BINDER service is available to maintain your vacuum drying ovens in top condition. In addition to calibrating your equipment, we will also perform a comprehensive safety inspection in accordance with DIN.
|
||
| VD 23 | ||
| Connection kit. Consists of: clamping ring, aluminum, centering ring, small flange with hose nozzle, 2.5 m (8.2 ft.) rubber hose and 2 hose clamps | O | |
| Measurement port, for air - tight lead - through of measurement lines into the unit (9 - pin) | O | |
| Specimen temperature measurement via flexible PT 100 sensor with digital temperature display | O | |
| Factory calibration certificate. Measurement in center of chamber at 100 °C (212 °F) or at specified testing temperature | O | |
| Extension to factory calibration certificate. Each additional measurement at an additional measuring point or temperature | O | |
| Factory calibration certificate for digital pressure display. Measurement at 100 mbar or at specified pressure (range: 20 - 900 mbar) | O | |
| Factory calibration certificate for digital pressure display, extension. Each additional measurement at an additional pressure (range: 20 - 900 mbar) | O | |
| Factory calibration certificate for digital specimen temperature display. Measurement at 100 °C (212 °F) or at specified testing temperature | O | |
| Expansion racks, aluminum or stainless steel | O | |
| Exchange and calibration to stainless steel expansion racks, standard equipment aluminum expansion racks will be replaced | O | |
| Door gasket, FKM | O | |
| Increased temperature range up to max. 250 °C (482 °F) | O | |
| Vacuum module for installation of vacuum pumps, with switchable power socket (ON / OFF) (230 V / 16 A) | O | |
| Vacuum module with chemical membrane pump VP 1 (nominal air flow 1.7 m³/hour), final pressure 9 mbar, with separator and emission condenser, including all necessary vacuum connection parts. Pump is supplied ready for connection in separate packaging | O | |
| Vacuum module with chemical membrane pump VP 2 (nominal air flow 3.0 m³ / hour), final pressure 2 mbar, with separator and emission condenser, including all necessary vacuum connection parts. Pump is supplied ready for connection in separate packaging | O | |
| Vacuum module with speed-controlled chemical membrane pump VP 3 (nominal air flow 3.5 m³ / hour), final pressure 2 mbar, with separator and emission condenser, including all necessary vacuum connection parts, as well as digital vacuum controller (measuring precision 1 mbar) | O |
Vacuum drying oven plus vacuum module with vacuum pump
-
BINDER’s one-stop solution consists of:
- A matching single-source solution consisting of the vacuum drying oven with a vacuum pump
- Space-saving, low-noise modular vacuum pump unit
- Easy and user-friendly condenser access
- A technically proven and approved solution, easy to start up and operate
- Time-saving, programmable intermediate air release and 2-stage drying
![]() |
![]() |
![]() |
|
| Vacuum module with integrated chemical membrane pump |
Vacuum module with chemical membrane pump |
Vacuum module with chemical membrane pump |
Vacuum module with speed-controlled chemical membrane pump |
| Type | VP 1 | VP 2 | VP 3 |
| Nominal airflow | 1.7 m³/h (60 cu.ft./h) | 3.0 m³/h (106 cu.ft./h) | 3.8 m³/h (134 cu.ft./h) |
| Final pressure (mbar / torr) | 9 / 7 | 2 / 1.5 | 2 / 1.5 |
| VD 23 | O | O | O |
| VD 53 | O | O | O |
| VD 115 | O | O | O |
| Application details | Powerful pump for each application | Increased volumetric displacement for samples with higher solvent content and for achieving a lower ultimate vacuum | Variably adjustable suction capacity and automatic vacuum adjustment during the process achieves up to 30 % shorter processing times |
O option - not available
Technical specifications subject to change