Hybridization ovens from the APT.line™

BFD series:
Hybridization ovens

The BFD series by far exceeds the requirements expected from standard hybridization ovens, with its unmistakably precise temperature accuracy, above-average thermal insulation, and perfect cleaning options. The wide temperature range is particularly pleasing from economic aspects. The BFD series can also be used as an incubator, if needed.

BINDER Hybridization ovens

  • Electronically controlled APT.line™ preheating chamber technology
  • Temperature range of 5 °C (9 °F) above ambient temperature up to 100 °C (212 °F)
  • Digital temperature setting with tenth of a degree accuracy
  • DS controller
  • Timer 0 to 24 hours
  • Removable rotation unit for 12 hybridization bottles
  • Low profile collection pan
  • Adjustable ventilation by means of rear exhaust duct, 50 mm (2 inch) diameter, with ventilation flap and front ventilation slide
  • Inner glass door
  • Independent adjustable temperature safety device, Class 3.1 (DIN 12880), with visual temperature alarm
  • Units are stackable
  • 1 chrome-plated rack
  • BINDER test certificate
BFD 53
Exterior dimensions 
Width (mm/inch) 659 / 25.9
Height (inclusive feet/castors) (mm/inch) 617 / 24.3
Depth (mm/inch) 575 / 22.6
Plus door handle, I-panel and exhaust duct (mm/inch) 85 / 3.4
Wall clearance rear (mm/inch) 100 / 3.9
Wall clearance side (mm/inch) 160 / 6.3
Steam space volume (l/cu.ft.) 77 / 2.7
Number of doors 1
  
Interior dimensions 
Width (mm/inch) 400 / 15.8
Height (mm/inch) 400 / 15.8
Depth (mm/inch) 330 / 13.0
Interior volume (l/cu.ft.) 53 / 1.9
Racks, chrome-plated (number standard/max.) 1/4
Load per rack (kg/lbs.) 15 / 33
Permitted total load (kg/lbs.) 40 / 88
Weight of the unit (empty) (kg/lbs.) 45 / 99
  
Temperature data 
Temperature range, 5 °C (9 °F) above ambient up to (°C/°F) 100 / 212
Temperature variation 
at 37 °C (98.6 °F) (± °C) 0,5
at 50 °C (122 °F) (± °C) 1,5
Temperature fluctuation 
at 37 °C (98.6 °F) (± °C) 0,1
at 50 °C (122 °F) (± °C) 0,2
Heating up time 1) 
to 37 °C (98.6 °F) (Min.) 12
to 50 °C (122 °F) (Min.) 22
Recov. time after door was opened for 30 sec. 1) 
at 37 °C (98.6 °F) (Min.) 1
at 50 °C (122 °F) (Min.) 1,5
  
Electrical data 
Housing protection acc. to EN 60529 IP 20
Nominal voltage (±10 %) 50 Hz (V) 230 / 115
Nominal power (W) 400
Energy consumption 
at 37 °C (98.6 °F) (W) 20
at 50 °C (122 °F) (W) 130
  

1) up to 98 % of the set value

All specified technical data apply for standard equipment operating at an ambient temperature of + 25 °C (77 °F), with line voltage fluctuations of ± 10%. These typical average values for series equipment were determined in accordance with the BINDER factory standard. They are based upon the recommended distances from the inner chamber walls, namely 10% of height, width and depth. Differing ambient temperatures or variances in the design of individual equipment may produce different performance data.