Heating chambers from the APT.line™
FED series:
Heating ovens with forced convection
The FED series is a true all-rounder. It has a virtually unlimited capacity and is at the same time particularly adaptable to the specific requirements of a large variety of testing applications. The enhanced time functions and the digitally controlled air turbine can be used to adjust ideal temperature parameters and recirculation air conditions.
- Electronically controlled APT.line™ preheating chamber technology with forced convection
- Temperature range of 5 °C (9 °F) above ambient temperature up to 300 °C (572 °F)
- MS controller with several timer functions
- Controller timer functions: delayed ON, delayed OFF, temperature dependent delayed OFF
- Digital temperature setting with one degree accuracy
- One ramp function
- Adjustable fan speed (0 to 100 %)
- Independent adjustable temperature safety device, Class 2 (DIN 12880), with visual temperature alarm
- Adjustable ventilation by means of rear exhaust duct Ø 50 mm (2 inch) with ventilation flap and front ventilation slide
- RS 422 interface for communication software APT-COM™ DataControlSystem, or switch over to printer output with RS 232/RS 422 interface converter
- Adjustable intervals for printer
- Units up to 115 liters (4.1 cu.ft.) are stackable
- 2 chrome-plated racks
- BINDER test certificate
| FED 53 | FED 115 | FED 240 | FED 400 | FED 720 | |
| Exterior dimensions | |||||
| Width (mm/inch) | 634 / 25.0 | 834 / 32.8 | 1034 / 40.7 | 1234 / 48.6 | 1234 / 48.6 |
| Height (inclusive feet/castors) (mm/inch) | 617 / 24.3 | 702 / 27.6 | 822 / 32.4 | 1022 / 40.2 | 1528 / 60.2 |
| Depth (mm/inch) | 575 / 22.6 | 645 / 25.4 | 745 / 29.3 | 765 / 30.1 | 865 / 34.1 |
| Plus door handle, I-panel and exhaust duct (mm/inch) | 105 / 4.1 | 105 / 4.1 | 105 / 4.1 | 105 / 4.1 | 105 / 4.1 |
| Wall clearance rear (mm/inch) | 100 / 3.9 | 100 / 3.9 | 100 / 3.9 | 100 / 3.9 | 100 / 3.9 |
| Wall clearance side (mm/inch) | 160 / 6.3 | 160 / 6.3 | 160 / 6.3 | 160 / 6.3 | 160 / 6.3 |
| Exhaust duct outer- Ø (mm/inch) | 52 / 2.1 | 52 / 2.1 | 52 / 2.1 | 52 / 2.1 | 52 / 2.1 |
| Steam space volume (l/cu.ft.) | 77 / 2.7 | 158 / 5.6 | 308 / 10.9 | 498 / 17.6 | 869 / 30.7 |
| Number of doors | 1 | 1 | 2 | 2 | 2 |
| Interior dimensions | |||||
| Width (mm/inch) | 400 / 15.8 | 600 / 23.6 | 800 / 31.5 | 1000 / 39.4 | 1000 / 39.4 |
| Height (mm/inch) | 400 / 15.8 | 480 / 18.9 | 600 / 23.6 | 800 / 31.5 | 1200 / 47.2 |
| Depth (mm/inch) | 330 / 13.0 | 400 / 15.8 | 500 / 19.7 | 500 / 19.7 | 600 / 23.6 |
| Interior volume (l/cu.ft.) | 53 / 1.9 | 115 / 4.1 | 240 / 8.6 | 400 / 14.3 | 720 / 25.7 |
| Racks, chrome-plated (number standard/max.) | 2/5 | 2/6 | 2/7 | 2/10 | 2/16 |
| Load per rack (kg/lbs.) | 15 / 33 | 20 / 44 | 30 / 66 | 35 / 77 | 45 / 99 |
| Permitted total load (kg/lbs.) | 40 / 88 | 50 / 110 | 70 / 155 | 90 / 199 | 120 / 265 |
| Weight of the unit (empty) (kg/lbs.) | 44 / 97 | 62 / 137 | 96 / 212 | 145 / 320 | 195 / 430 |
| Temperature data | |||||
| Temperature range, 5°C (9°F) above ambient up to (°C/°F) | 300 / 572 | 300 / 572 | 300 / 572 | 300 / 572 | 300 / 572 |
| Temperature variation 1) | |||||
| at 70 °C (158 °F) (± °C) | 0,8 | 0,7 | 0,8 | 1 | 1 |
| at 150 °C (302 °F) (± °C) | 2 | 1,8 | 2 | 2,5 | 2 |
| at 300 °C (572 °F) (± °C) | 3,7 | 3,9 | 4,3 | 4,8 | 5,5 |
| Temperature fluctuation (± °C) | 0,3 | 0,3 | 0,3 | 0,3 | 0,3 |
| Heating up time 2) | |||||
| to 70 °C (158 °F) (Min.) | 6 | 7 | 12 | 18 | 25 |
| to 150 °C (302 °F) (Min.) | 24 | 30 | 27 | 35 | 39 |
| to 250 °C (482 °F) (Min.) | 45 | 49 | 50 | 60 | 65 |
| Recov. time after door was opened for 30 sec. 2) | |||||
| at 70 °C (158 °F) (Min.) | 2 | 2 | 2 | 2 | 2 |
| at 150 °C (302 °F) (Min.) | 5 | 8 | 10 | 17 | 20 |
| at 300 °C (572 °F) (Min.) | 10 | 15 | 16 | 21 | 24 |
| Air change 3) | |||||
| at 150 °C (302 °F) (x/h) | 64 | 32 | 20 | 18 | 12 |
| Electrical data | |||||
| Housing protection acc. to EN 60529 | IP 20 | IP 20 | IP 20 | IP 20 | IP 20 |
| Nominal voltage (±10 %) 50/60 Hz (V) | 230 / 115 | 230 / 115 | 230 / 115 | 400 3/N | 400 3/N |
| Nominal power (W) | 1200 | 1600 | 2700 | 3400 | 5000 |
| Energy consumption | |||||
| at 70 °C (158 °F) (W) | 145 | 230 | 370 | 520 | 570 |
| at 150 °C (302 °F) (W) | 300 | 544 | 850 | 1200 | 1320 |
| at 300 °C (572 °F) (W) | 720 | 1100 | 1400 | 2340 | 2600 |
1) value without window
2) up to 98 % of the set value
3) The air change depends on the inner chamber- and ambient temperature. Further measurement of air change rate in acc. to ASTM D5374 with definition and protocol are optional.
All specified technical data apply for standard equipment operating at an ambient temperature of + 25 °C (77 °F), with line voltage fluctuations of ± 10%. These typical average values for series equipment were determined in accordance with the BINDER factory standard. They are based upon the recommended distances from the inner chamber walls, namely 10% of height, width and depth. Differing ambient temperatures or variances in the design of individual equipment may produce different performance data.
We therefore recommend that equipment be calibrated and/or validated on a case-by-case basis when working at the extremes of the permissible ambient temperature range.