Drying ovens from the APT.line™

ED series:
Drying ovens with natural convection

Routine drying and sterilization applications up to 300° C and storage at precisely controlled elevated temperatures are the strengths of ED drying ovens. Because of the natural air circulation with a high rate of air exchange, thermal processes run with significantly increased efficiency.

BINDER Wärmetrockenschränke

  • Electronically-controlled APT.line™ preheating chamber technology with natural convection
  • Temperature range of 5 °C (9 °F) above ambient temperature up to 300 °C (572 °F)
  • DS controller with integrated timer 0 to 99 hours
  • Digital temperature setting with one degree accuracy
  • One ramp function
  • Independent adjustable temperature safety device, Class 2 (DIN 12880), with visual temperature alarm
  • Adjustable ventilation by means of rear exhaust duct Ø 50 mm (2 inch) with ventilation flap and front ventilation slide
  • 2 chrome-plated racks
  • Units up to 115 liters (4.1 cu.ft.) are stackable
  • Optional RS 422 interface for communication software APT-COM™ DataControlSystem
  • Only at BINDER: Precision regardless of size – the 20-liter (0.7 cu.ft.) bench-top units.
  • BINDER test certificate
ED 23ED 53ED 115ED 240ED 400ED 720
Exterior dimensions      
Width (mm/inch) 433 / 17.1 634 / 25.0 834 / 32.8 1034 / 40.7 1234 / 48.6 1234 / 48.6
Height (inclusive feet/castors) (mm/inch) 492 / 19.4 617 / 24.3 702 / 27.6 822 / 32.4 1022 / 40.2 1528 / 60.2
Depth (mm/inch) 516 / 20.3 575 / 22.6 645 / 25.4 745 / 29.3 765 / 30.1 865 / 34.1
plus door handle, I-panel and exhaust duct (mm/inch) 85 / 3.4 85 / 3.4 85 / 3.4 85 / 3.4 85 / 3.4 85 / 3.4
Wall clearance rear (mm/inch) 100 / 3.9 100 / 3.9 100 / 3.9 100 / 3.9 100 / 3.9 100 / 3.9
Wall clearance side (mm/inch) 100 / 3.9 160 / 6.3 160 / 6.3 160 / 6.3 160 / 6.3 160 / 6.3
Exhaust duct outer-Ø (mm/inch) 52 / 2.1 52 / 2.1 52 / 2.1 52 / 2.1 52 / 2.1 52 / 2.1
Steam space volume (l/cu.ft.) 36 / 1.3 70 / 2.5 142 / 5.0 283 / 10.0 457 / 16.2 808 / 28.6
Number of doors 1 1 1 2 2 2
       
Interior dimensions      
Width (mm/inch) 222 / 8.7 400 / 15.8 600 / 23.6 800 / 31.5 1000 / 39.4 1000 / 39.4
Height (mm/inch) 330 / 13.0 400 / 15.8 480 / 18.9 600 / 23.6 800 / 31.5 1200 / 47.2
Depth (mm/inch) 277 / 10.9 330 / 13.0 400 / 15.8 500 / 19.7 500 / 19.7 600 / 23.6
Interior volume (l/cu.ft.) 20 / 0.7 53 / 1.9 115 / 4.1 240 / 8.6 400 / 14.3 720 / 25.7
Racks, chrome-plated (number standard/max.) 2/3 2/5 2/6 2/7 2/10 2/16
Load per rack (kg/lbs.) 12 / 26 15 / 33 20 / 44 30 / 66 35 / 77 45 / 99
Permitted total load (kg/lbs.) 25 / 55 40 / 88 50 / 110 70 / 155 90 / 199 120 / 265
Weight of the unit (empty) (kg/lbs.) 25 / 55 42 / 93 57 / 126 86 / 190 125/276 174 / 384
       
Temperature data      
Temperature range, 5°C (9°F) above ambient up to (°C/°F) 300 / 572 300 / 572 300 / 572 300 / 572 300 / 572 300 / 572
Temperature variation 1)      
at 70 °C (158 °F) (± °C) 1,5 2 1,5 1,5 1,7 1,5
at 150 °C (302 °F) (± °C) 2,5 3,2 2,5 2,5 3 2,8
at 300 °C (572 °F) (± °C) 3,8 4,5 4,5 5,0 5,0 5,0
Temperature fluctuation (± °C) 0.2 0,3 0,3 0,3 0,3 0,3
Heating up time 2)      
to 70 °C (158 °F) (Min.) 13 14 15 40 49 56
to 150 °C (302 °F) (Min.) 24 27 29 48 62 69
to 250 °C (482 °F) (Min.) 35 61 66 61 74 80
Recov. time after door was opened for 30 sec. 2)      
at 70 °C (158 °F) (Min.) 2,5 2 2 5 4 4
at 150 °C (302 °F) (Min.) 5 6 9 13 20 14
at 300 °C (572 °F) (Min.) 8 11 14 18 24 18
Air change 3)      
at 150 °C (302 °F) (x/h) 19 12 10 10 10 9
       
Electrical data      
Housing protection acc. to EN 60529 IP 20 IP 20 IP 20 IP 20 IP 20 IP 20
Nominal voltage (±10 %) 50/60 Hz (V) 230 / 115 230 / 115 230 / 115 230 / 208 (3N) 400 / 208 (3N) 400 / 208 (3N)
Nominal power (W) 800 1200 1600 2700 3400 5000
Energy consumption      
at 70 °C (158 °F) (W) 56 56 90 143 201 220
at 150 °C (302 °F) (W) 188 188 300 447 672 750
at 300 °C (572 °F) (W) 320 320 360 700 1000 1200
       

1) value without window
2) up to 98 % of the set value
3) The air change depends on the inner chamber- and ambient temperature. Further measurement of air change rate in acc. to ASTM D5374 with definition and protocol are optional.

All specified technical data apply for standard equipment operating at an ambient temperature of + 25 °C (77 °F), with line voltage fluctuations of ± 10%. These typical average values for series equipment were determined in accordance with the BINDER factory standard. They are based upon the recommended distances from the inner chamber walls, namely 10% of height, width and depth. Differing ambient temperatures or variances in the design of individual equipment may produce different performance data.

We therefore recommend that equipment be calibrated and/or validated on a case-by-case basis when working at the extremes of the permissible ambient temperature range.