Vacuum drying ovens from the APT.line™
VDL EX-Zone 2 series:
Explosion-proof safety vacuum drying oven for EX-Zone 2
Same quality as the VDL series, but with some added safety features: The VDL series EX-Zone 2 is designed for operation in explosion hazard areas EX-Zone 2, in full compliance with ATEX requirements. Our special safety package eliminates any risk.
- Electronically controlled APT.line™ preheating chamber technology with 2 aluminum expansion racks
- Temperature range of 15 °C (27 °F) above ambient temperature up to 200 °C ( 392 °F)
- MP controller with 2 programs with 10 sections each, or alternatively 1 program with 20 sections
- The time interval of single program sections can be adjusted up to a maximum of 99:59 hours or 999:59 hours. This adjustment applies to all program sections.
- Digital temperature setting with an accuracy of one or tenth of a degree
- Adjustable ramp function via program editor
- Integrated week program timer with real time function
- Elapsed time indicator
- Pressure monitor for controlled heating release at < 125 mbar (94 torr)
- Pressure encapsulated instrument compartment with contolled overpressure by means of compressed air
- Flame protection gasket
- Fine dosing ventilation valve
- Fine dosing inert gas valve with cross-flow technology
- Spring-mounted safety glass pane with splinter protection
- Analog pressure gauge (display pressure difference between the inner chamber and the ambient pressure)
- Electro polished inner chamber
- All suction and ventilation tubes, as well as the internally-welded pressure container, the removable rack support and the vacuum valve are made of stainless steel Mat. No. 1.4571 (V4A) / AISI 316 Ti
- Door gasket made of tempered silicone
- Independent adjustable temperature safety device, Class 2 (DIN 12880), with visual alarm
- DN 16 measuring port in rear wall
- RS 422 interface for communication software APT-COM™ DataControlSystem, or switch over to printer output with RS 232 / RS 422 interface converter
- BINDER test certificate
| VDL EX 23 | VDL EX 53 | VDL EX 115 | |
| Exterior dimensions | |||
| Width (mm/inch) | 515 / 20.3 | 634 / 25.0 | 740 / 29.1 |
| Height (inclusive feet/castors) (mm/inch) | 655 / 25.8 | 775 / 30.5 | 900 / 35.4 |
| Height option vacuum module (mm/inch) | 624 / 24.6 | 624 / 24.6 | 622 / 24.5 |
| Total height VDL and option vacuum module (mm/inch) | 1279 / 50.4 | 1400 / 55.1 | 1522 / 59.9 |
| Depth (mm/inch) | 500 / 19.7 | 550 / 21.7 | 670 / 26.4 |
| Plus door handle, connection (mm/inch) | 100 / 3.9 | 100 / 3.9 | 100 / 3.9 |
| Wall clearance rear (mm/inch) | 100 / 3.9 | 100 / 3.9 | 100 / 3.9 |
| Wall clearance side (mm/inch) | 135 / 5.3 | 135 / 5.3 | 135 / 5.3 |
| Interior dimensions | |||
| Width (mm/inch) | 285 / 11.2 | 400 / 15.8 | 506 / 19.9 |
| Height (mm/inch) | 285 / 11.2 | 400 / 15.8 | 506 / 19.9 |
| Depth (mm/inch) | 285 / 11.2 | 330 / 13.0 | 450 / 17.7 |
| Interior volume (l/cu.ft.) | 23 / 0.8 | 53 / 1.9 | 115 / 4.1 |
| Expansion shelves (Aluminium) (number standard/max.) | 2 / 4 | 2 / 5 | 2 / 6 |
| Distance between the racks (width x depth) (mm/inch) | 53 / 2.1 | 62 / 2.4 | 68 / 2.7 |
| Usable space per rack (width x depth) (mm/inch) | 234 x 280 / 9.2 x 11 | 349 x 320 / 13.7 x 12.6 | 455 x 440 / 17.9 x 17.3 |
| Load per rack (kg/lbs.) | 20 / 44 | 20 / 44 | 20 / 44 |
| Permitted total load (kg/lbs.) | 35 / 77 | 45 / 99 | 65 / 143 |
| Weight of the unit (empty) (kg/lbs.) | 76 / 167.6 | 94 / 207.2 | 165 / 363.8 |
| Temperature data | |||
| Temperature range, 15 °C (27 °F) above ambient up to (°C/°F) | 200 / 392 | 200 / 392 | 200 / 392 |
| Temperature variation 3) | |||
| at 100 °C (212 °F) (± °C) | 2 | 2,5 | 4 |
| at 200 °C (392 °F) (± °C) | 4 | 4,5 | 7 |
| Temperature fluctuation 3) (± °C) | 0.4 | 0.4 | 0.4 |
| Heating up time 2) | |||
| to 100 °C (212 °F) (Min.) | 80 | 70 | 80 |
| to 200 °C (392 °F) (Min.) | 110 | 130 | 150 |
| Vacuum connection with small flange (DN mm/inch) | 16 / 0.63 | 16 / 0.63 | 16 / 0.63 |
| Measuring access port with small flange (DN mm/in) | 16 / 0.63 | 16 / 0.63 | 16 / 0.63 |
| Inert gas connection with flow limiter - thread (RP") | 3 / 8 | 3 / 8 | 3 / 8 |
| Inert gas connection with flow limiter Adapter with hose olive (DN mm/inch) | 8 / 0.31 | 8 / 0.31 | 8 / 0.31 |
| Permitted end vacuum (mbar / torr) | 0.01 / 0.008 | 0.01 / 0.008 | 0.01 / 0.008 |
| Leak rate (max. bar 1/h / torr 1/h) | 0.01 / 0.008 | 0.01 / 0.008 | 0.01 / 0.008 |
| Compressed air connect. f. pressure-encapsulation (Ø mm/in.) | 8 / 0.32 | 8 / 0.32 | 8 / 0.32 |
| Compressed air use (l/Min.) | 50 | 50 | 50 |
| Electrical data | |||
| Housing protection acc. to EN 60529 | IP 54 | IP 54 | IP 54 |
| Nominal voltage (±10 %) 50/60 Hz (V) | 230 | 230 | 230 |
| Nominal power (W) | 800 | 1200 | 1900 |
| Energy consumption | |||
| at 100 °C (212 °F) (W) | 105 | 150 | 250 |
| at 200 °C (392 °F) (W) | 280 | 445 | 785 |
1) value with aluminium racks
2) up to 98 % of the set value
All specified technical data apply for standard equipment operating at an ambient temperature of + 25 °C (77 °F), with line voltage fluctuations of ± 10%. These typical average values for series equipment were determined in accordance with the BINDER factory standard. They are based upon the recommended distances from the inner chamber walls, namely 10% of height, width and depth. Differing ambient temperatures or variances in the design of individual equipment may produce different performance data.
We therefore recommend that equipment be calibrated and/or validated on a case-by-case basis when working at the extremes of the permissible ambient temperature range.
Vacuum drying oven plus vacuum module with vacuum pump
BINDER’s one-stop solution consists of:
- A matching single-source solution consisting of the vacuum drying oven with a vacuum pump
- Space-saving, low-noise modular vacuum pump unit
- Easy and user-friendly condenser access
- A technically proven and approved solution, easy to start up and operate
- Time-saving, programmable intermediate air release and 2-stage drying
- Full compliance with ATEX Directives
![]() |
![]() |
|
| Vacuum module with integrated chemical membrane pump |
Vacuum module with chemical membrane pump, Ex-proof |
Vacuum module with chemical membrane pump, Ex-proof |
| Type | VP 4 | VP 5 |
| Nominal airflow | 1.9 m³/h (67 cu.ft./h) | 3.7 m³/h (131 cu.ft./h) |
| Final pressure (mbar / torr) | 12 / 9 | 3 / 2 |
| ATEX approved according to ATEX Directive 94/9/EC | yes | yes |
| VDL 23 | O | - |
| VDL 53 | O | - |
| VDL 115 | O | O |
| Application details | Approved ATEX-compliant diaphragm pump for VDL 23, VDL 53 and VDL 115 series | Approved ATEX-compliant diaphragm pump with a reduced final vacuum for the VDL 115 series |
O option - not available
Technical specifications subject to change
As of July 1, 2003, only ATEX-compliant equipment may be used in explosion hazard areas.
All BINDER VDL series vacuum pumps are ATEX certified.