Vacuum drying ovens from the APT.line™

VDL EX-Zone 2 series:
Explosion-proof safety vacuum drying oven for EX-Zone 2

Same quality as the VDL series, but with some added safety features: The VDL series EX-Zone 2 is designed for operation in explosion hazard areas EX-Zone 2, in full compliance with ATEX requirements. Our special safety package eliminates any risk.

BINDER Vakuumtrockenschränke

  • Electronically controlled APT.line™ preheating chamber technology with 2 aluminum expansion racks
  • Temperature range of 15 °C (27 °F) above ambient temperature up to 200 °C ( 392 °F)
  • MP controller with 2 programs with 10 sections each, or alternatively 1 program with 20 sections
  • The time interval of single program sections can be adjusted up to a maximum of 99:59 hours or 999:59 hours. This adjustment applies to all program sections.
  • Digital temperature setting with an accuracy of one or tenth of a degree
  • Adjustable ramp function via program editor
  • Integrated week program timer with real time function
  • Elapsed time indicator
  • Pressure monitor for controlled heating release at < 125 mbar (94 torr)
  • Pressure encapsulated instrument compartment with contolled overpressure by means of compressed air
  • Flame protection gasket
  • Fine dosing ventilation valve
  • Fine dosing inert gas valve with cross-flow technology
  • Spring-mounted safety glass pane with splinter protection
  • Analog pressure gauge (display pressure difference between the inner chamber and the ambient pressure)
  • Electro polished inner chamber
  • All suction and ventilation tubes, as well as the internally-welded pressure container, the removable rack support and the vacuum valve are made of stainless steel Mat. No. 1.4571 (V4A) / AISI 316 Ti
  • Door gasket made of tempered silicone
  • Independent adjustable temperature safety device, Class 2 (DIN 12880), with visual alarm
  • DN 16 measuring port in rear wall
  • RS 422 interface for communication software APT-COM™ DataControlSystem, or switch over to printer output with RS 232 / RS 422 interface converter
  • BINDER test certificate
VDL EX 23VDL EX 53VDL EX 115
Exterior dimensions   
Width (mm/inch) 515 / 20.3 634 / 25.0 740 / 29.1
Height (inclusive feet/castors) (mm/inch) 655 / 25.8 775 / 30.5 900 / 35.4
Height option vacuum module (mm/inch) 624 / 24.6 624 / 24.6 622 / 24.5
Total height VDL and option vacuum module (mm/inch) 1279 / 50.4 1400 / 55.1 1522 / 59.9
Depth (mm/inch) 500 / 19.7 550 / 21.7 670 / 26.4
Plus door handle, connection (mm/inch) 100 / 3.9 100 / 3.9 100 / 3.9
Wall clearance rear (mm/inch) 100 / 3.9 100 / 3.9 100 / 3.9
Wall clearance side (mm/inch) 135 / 5.3 135 / 5.3 135 / 5.3
    
Interior dimensions   
Width (mm/inch) 285 / 11.2 400 / 15.8 506 / 19.9
Height (mm/inch) 285 / 11.2 400 / 15.8 506 / 19.9
Depth (mm/inch) 285 / 11.2 330 / 13.0 450 / 17.7
Interior volume (l/cu.ft.) 23 / 0.8 53 / 1.9 115 / 4.1
Expansion shelves (Aluminium) (number standard/max.) 2 / 4 2 / 5 2 / 6
Distance between the racks (width x depth) (mm/inch) 53 / 2.1 62 / 2.4 68 / 2.7
Usable space per rack (width x depth) (mm/inch) 234 x 280 / 9.2 x 11 349 x 320 / 13.7 x 12.6 455 x 440 / 17.9 x 17.3
Load per rack (kg/lbs.) 20 / 44 20 / 44 20 / 44
Permitted total load (kg/lbs.) 35 / 77 45 / 99 65 / 143
Weight of the unit (empty) (kg/lbs.) 76 / 167.6 94 / 207.2 165 / 363.8
    
Temperature data   
Temperature range, 15 °C (27 °F) above ambient up to (°C/°F) 200 / 392 200 / 392 200 / 392
Temperature variation 3)   
at 100 °C (212 °F) (± °C) 2 2,5 4
at 200 °C (392 °F) (± °C) 4 4,5 7
Temperature fluctuation 3) (± °C) 0.4 0.4 0.4
Heating up time 2)   
to 100 °C (212 °F) (Min.) 80 70 80
to 200 °C (392 °F) (Min.) 110 130 150
Vacuum connection with small flange (DN mm/inch) 16 / 0.63 16 / 0.63 16 / 0.63
Measuring access port with small flange (DN mm/in) 16 / 0.63 16 / 0.63 16 / 0.63
Inert gas connection with flow limiter - thread (RP") 3 / 8 3 / 8 3 / 8
Inert gas connection with flow limiter Adapter with hose olive (DN mm/inch) 8 / 0.31 8 / 0.31 8 / 0.31
Permitted end vacuum (mbar / torr) 0.01 / 0.008 0.01 / 0.008 0.01 / 0.008
Leak rate (max. bar 1/h / torr 1/h) 0.01 / 0.008 0.01 / 0.008 0.01 / 0.008
Compressed air connect. f. pressure-encapsulation (Ø mm/in.) 8 / 0.32 8 / 0.32 8 / 0.32
Compressed air use (l/Min.) 50 50 50
    
Electrical data   
Housing protection acc. to EN 60529 IP 54 IP 54 IP 54
Nominal voltage (±10 %) 50/60 Hz (V) 230 230 230
Nominal power (W) 800 1200 1900
Energy consumption   
at 100 °C (212 °F) (W) 105 150 250
at 200 °C (392 °F) (W) 280 445 785
    

1) value with aluminium racks
2) up to 98 % of the set value

All specified technical data apply for standard equipment operating at an ambient temperature of + 25 °C (77 °F), with line voltage fluctuations of ± 10%. These typical average values for series equipment were determined in accordance with the BINDER factory standard. They are based upon the recommended distances from the inner chamber walls, namely 10% of height, width and depth. Differing ambient temperatures or variances in the design of individual equipment may produce different performance data.

We therefore recommend that equipment be calibrated and/or validated on a case-by-case basis when working at the extremes of the permissible ambient temperature range.

Vacuum drying oven plus vacuum module with vacuum pump


  • Vacuummodul

    BINDER’s one-stop solution consists of:


    • A matching single-source solution consisting of the vacuum drying oven with a vacuum pump
    • Space-saving, low-noise modular vacuum pump unit
    • Easy and user-friendly condenser access
    • A technically proven and approved solution, easy to start up and operate
    • Time-saving, programmable intermediate air release and 2-stage drying
    • Full compliance with ATEX Directives
VP4 VP5
Vacuum module with integrated
chemical membrane pump
Vacuum module with chemical membrane pump,
Ex-proof
Vacuum module with chemical membrane pump,
Ex-proof
Type VP 4 VP 5
 
Nominal airflow 1.9 m³/h (67 cu.ft./h) 3.7 m³/h (131 cu.ft./h)
Final pressure (mbar / torr) 12 / 9 3 / 2
ATEX approved according to ATEX Directive 94/9/EC yes yes
VDL 23 O -
VDL 53 O -
VDL 115 O O
Application details Approved ATEX-compliant diaphragm pump for VDL 23, VDL 53 and VDL 115 series Approved ATEX-compliant diaphragm pump with a reduced final vacuum for the VDL 115 series

O option     - not available

Technical specifications subject to change

As of July 1, 2003, only ATEX-compliant equipment may be used in explosion hazard areas.
All BINDER VDL series vacuum pumps are ATEX certified.