M series:
Temperature test chambers with forced convection and individual programming
Because of its individual programming options and ability to operate at maximum temperatures up to 300 °C, the M series is ideally suited for materials testing and aging tests. The heavy-duty air turbine and a programmable exhaust ventilation flap provide rapid heating-up and ensure that the test temperature is maintained absolutely precise at all levels, with minimal spatial fluctuations; performance as never before.
- Electronically controlled APT.line™ preheating chamber technology
- Temperature range of 5 °C (9 °F) above ambient temperature up to 300 °C (572 °F)
-
MCS controller with 25 storable programs of 100 sections each for a maximum of 500 program segments
- User-friendly LCD screen
- Easy-to-read menu guide
- Integrated electronic chart recorder
- Variety of options for the graphic display of process parameters
- Real-time clock
- Adjustable ramp function via program editor
- Program-controlled ventilation flap
- High air-exchange rate through high-performance fan (+ approx. 280 %)
- Adjustable fan speed (0 to 100 %)
- Rear exhaust duct Ø 50 mm (2 inch) diameter
- Independent adjustable temperature safety device, Class 2 (DIN 12880), with optical temperature alarm
- RS 422 interface for communication software APT-COM™ DataControlSystem
- 2 chrome-plated shelves
- BINDER test certificate
| M 53 | M 115 | M 240 | M 400 | M 720 | |
| Exterior dimensions | |||||
| Width (mm/inch) | 634 / 25.0 | 834 / 32.8 | 1034 / 40.7 | 1234 / 48.6 | 1234 / 48.6 |
| Height (inclusive feet/castors) (mm/inch) | 779 / 30.7 | 863 / 34.0 | 984 / 38.7 | 1184 / 46.6 | 1692 / 66.6 |
| Depth (mm/inch) | 575 / 22.6 | 645 / 25.4 | 745 / 29.3 | 765 / 30.1 | 865 / 34.1 |
| Plus door handle (mm/inch) | 150 / 5.9 | 150 / 5.9 | 150 / 5.9 | 150 / 5.9 | 150 / 5.9 |
| Wall clearance rear (mm/inch) | 100 / 3.9 | 100 / 3.9 | 100 / 3.9 | 100 / 3.9 | 100 / 3.9 |
| Wall clearance side (mm/inch) | 160 / 6.3 | 160 / 6.3 | 160 / 6.3 | 160 / 6.3 | 160 / 6.3 |
| Exhaust duct outer- Ø (mm/inch) | 52 / 2.1 | 52 / 2.1 | 52 / 2.1 | 52 / 2.1 | 52 / 2.1 |
| Steam space volume (l/cu.ft.) | 77 / 2.7 | 158 / 5.6 | 308 / 10.9 | 498 / 17.6 | 869 / 30.7 |
| Number of doors | 1 | 1 | 2 | 2 | 2 |
| Interior dimensions | |||||
| Width (mm/inch) | 400 / 15.8 | 600 / 23.6 | 800 / 31.5 | 1000 / 39.4 | 1000 / 39.4 |
| Height (mm/inch) | 400 / 15.8 | 480 / 18.9 | 600 / 23.6 | 800 / 31.5 | 1200 / 47.2 |
| Depth (mm/inch) | 330 / 13.0 | 400 / 15.8 | 500 / 19.7 | 500 / 19.7 | 600 / 23.6 |
| Interior volume (l/cu.ft.) | 53 / 1.9 | 115 / 4.1 | 240 / 8.6 | 400 / 14.3 | 720 / 25.7 |
| Shelves, chrome-plated (number standard/max.) | 2/5 | 2/6 | 2/8 | 2/10 | 2/16 |
| Load per rack (kg/lbs.) | 15 / 33 | 20 / 44 | 30 / 66 | 35 / 77 | 45 / 99 |
| Permitted total load (kg/lbs.) | 40 / 88 | 50 / 110 | 70 / 155 | 90 / 199 | 120 / 265 |
| Weight of the unit (empty) (kg/lbs.) | 61 / 135 | 89 / 196 | 131 / 289 | 173 / 382 | 203 / 448 |
| Temperature data | |||||
| Temperature range, 5°C (9°F) above ambient up to (°C/°F) | 300 / 572 | 300 / 572 | 300 / 572 | 300 / 572 | 300 / 572 |
| Temperature variation 1) | |||||
| at 70 °C (158 °F) (± °C) | 0,5 | 0,6 | 0,8 | 0,7 | 0,7 |
| at 150 °C (302 °F) (± °C) | 1,3 | 1,5 | 1,5 | 1,5 | 1,9 |
| at 300 °C (572 °F) (± °C) | 2,8 | 2,8 | 2,8 | 5 | 4,6 |
| Temperature fluctuation (± °C) | 0,3 | 0,3 | 0,3 | 0,3 | 0,3 |
| Heating-up time 2) | |||||
| to 70 °C (158 °F) (Min.) | 5 | 5 | 6 | 6 | 7 |
| to 150 °C (302 °F) (Min.) | 15 | 16 | 19 | 18 | 21 |
| to 250 °C (482 °F) (Min.) | 35 | 36 | 42 | 44 | 51 |
| Recov. time after door was opened for 30 sec. 2) | |||||
| at 70 °C (158 °F) (Min.) | 1 | 1 | 1 | 1 | 1 |
| at 150 °C (302 °F) (Min.) | 3 | 3 | 3 | 3 | 3 |
| at 300 °C (572 °F) (Min.) | 5 | 5 | 5 | 5 | 5 |
| Air change 3) | |||||
| at 150 °C (302 °F) (x/h) | 192 | 96 | 60 | 54 | 36 |
| Electrical data | |||||
| Housing protection acc. to EN 60529 | IP 20 | IP 20 | IP 20 | IP 20 | IP 20 |
| Nominal voltage (±10 %) 50/60 Hz (V) | 230 | 230 | 230 | 400 3/N~ | 400 3/N~ |
| Nominal power (W) | 1200 | 1600 | 2700 | 3400 | 5000 |
| Energy consumption | |||||
| at 70 °C (W) | 145 | 230 | 370 | 520 | 570 |
| at 150 °C (W) | 300 | 544 | 850 | 1200 | 1320 |
| at 300 °C (W) | 720 | 1100 | 1400 | 2340 | 2600 |
1) value without window
2) up to 98 % of the set value
3) The air change depends on the inner chamber- and ambient temperature. Further measurement of air change rate in acc. to ASTM D5374 with definition and protocol are optional.
All specified technical data apply for standard equipment operating at an ambient temperature of + 25 °C (77 °F), with line voltage fluctuations of ± 10%. These typical average values for series equipment were determined in accordance with the BINDER factory standard. They are based upon the recommended distances from the inner chamber walls, namely 10% of height, width and depth. Differing ambient temperatures or variances in the design of individual equipment may produce different performance data.
We therefore recommend that equipment be calibrated and/or validated on a case-by-case basis when working at the extremes of the permissible ambient temperature range.