Plant growth chambers from the APT.line™
KBWF series:
Plant growth chambers with optimal climatic conditions
The closest thing to natural conditions. Making use of the multifaceted programming options, we achieve perfect interaction between heat or cold, humidity and light. This wide climatic range can simulate any climatic condition precisely and constant over extended periods of time, including natural lighting conditions and day-night simulation.
- Electronically controlled APT.line™ preheating chamber technology with DCT™ cooling system
- Temperature range: -5 °C to 100 °C (23 °F to 212 °F) without humidity and illumination
- Temperature range: 20 °C to 90 °C (68 °F to 194 °F) with humidity respective illumination
- MCS controller for temperature, humidity, and illumination with 25 storable programs of 100 sections each for a maximum of 500 program segments, for programming day/night cycles
- Integrated electronic chart recorder
- Variety of options for the graphic display of process parameters
- Real-time clock
-
Illumination system with daylight fluorescent tubes in the doors, thermally isolated from the
inner chamber and ventilated (optimized illumination)
- KBWF 240 - 10 count × 18 W
- KBWF 720 - 10 count × 36 W
- Electronically controlled humidification and dehumidification system for use with tap water
- Capacitve humidity sensor with SPH technology
- Automatic defrosting device
- Independent adjustable temperature safety device, Class 3.1 (DIN 12880) with visual and acoustic temperature alarm
- 2 access ports, Ø 30 mm (1.2 inch) each
- Environmentally friendly refrigerant R 134a
- Inner glass door
- Collecting pan for condensate on the door
- Complete safety connection kit for water supply incl. water hose as well as for water drain (total length 6 m / 19.7 ft.)
- RS 422 interface for communication software APT-COM™ DataControlSystem
- 2 stainless steel shelves
- BINDER test certificate
| KBWF 240 | KBWF 720 | |
| Exterior dimensions | ||
| Width (mm/inch) | 1034 / 40.7 | 1234 / 48.6 |
| Height (inclusive feet/castors) (mm/inch) | 1142 / 45.0 | 1816 / 71.5 |
| Depth (mm/inch) | 746 / 29.4 | 867 / 34.1 |
| Plus door handle, I-panel, connection (mm/inch) | 100 / 3.9 | 100 / 3.9 |
| Wall clearance (mm/inch) | 100 / 3.9 | 100 / 3.9 |
| Wall clearance with open door(s) (mm/inch) | 160 / 6.3 | 160 / 6.3 |
| Steam space volume (l/cu.ft.) | 308 / 12.1 | 855 / 33.7 |
| Height of water connections (± 3 mm / ± 0.12 inch) in mm/inch | 84 / 3.3 | 190 / 7.5 |
| Number of doors | 2 | 2 |
| Number of inner glass doors | 2 | 2 |
| Interior dimensions | ||
| Width (mm/inch) | 800 / 31.5 | 1000 / 39.4 |
| Height (mm/inch) | 600 / 23.6 | 1168 / 46.0 |
| Depth (mm/inch) | 500 / 19.7 | 600 / 23.6 |
| Interior volume (l/cu.ft.) | 240 / 8.5 | 700 / 24.7 |
| Shelves (number standard/max.) | 2/7 | 2/14 |
| Load per rack (kg/lbs.) | 30 / 66 | 45 / 99 |
| Permitted total load (kg/lbs.) | 70 / 155 | 120 / 265 |
| Weight of the unit (empty) (kg/lbs.) | 184 / 406 | 345 / 762 |
| Temperature data | ||
| Temperature range | ||
| without humidity / without illumination (°C/°F) | -5-100/23-212 | -5-100/23-212 |
| with humidity / with illumination (°C/°F) | 5–100 /41–212 | 5-100 /41-212 |
| with humidity / without illumination (°C/°F) | 20-90/68-194 | 20-90/68-194 |
| with humidity / with illumination (illumination in the door) (°C/°F) | 20-90/68-194 | 20-90/68-194 |
| Temperature variation without humidity | ||
| at 10 °C (50 °F) (± °C) | 0,4 | 0,4 |
| at 37 °C (98.6 °F) (± °C) | 0,4 | 0,4 |
| Temperature variation with humidity (± °C) | 1,0 | 1,0 |
| Temperature fluctuation during heating operation (± °C) | 0,1 | 0,1 |
| Heating up time 1), 2) to 37 °C (98.6 °F) (Min.) | 30 | 28 |
| Cooling down time from room temperature 1), 2) to 10 °C (50 °F) (Min.) | 35 | 35 |
| Recovery time after door was open for 30 sec 1), 2) | ||
| at 37 °C (98.6 °F) (Min.) | 5 | 5 |
| at 50 °C (122 °F) (Min.) | 4 | 4 |
| Max. illumination / growth lamps | ||
| (Lux) / μE/(s x m2)* | 3200 / 100 | 3200 / 100 |
| Max. illumination/daylight lamps | ||
| (Lux) / μE/(s x m2)* | 5000 / 120 | 5000 / 120 |
| Max. illum. with option high intensity illum. under the chamber celling (Lux) | 7000 | 7000 |
| Humidity fluctuation 1), 2), 3) (± % r.H.) | 1.5 | 1.5 |
| Programming of light intensity 20 % / 40 % / at 60 % / at 100 % | true | true |
| Electrical data | ||
| Housing protection acc. to EN 60529 | IP 20 | IP 20 |
| Nominal voltage (±10 %) 50/60 Hz (V) | 230 / 115 | 230 / 115 |
| Nominal power (W) | 2420 | 2950 |
| Energy consumption 4) at 37 °C (98.6 °F) (W) | 730 | 970 |
1) up to 98 % of the set value
2) value without illumination
3) upon door opening or water exchange in humidity cylinder: > ± 1.5 % r.H., recovery time approx. 20 min.
4) these energy consumption values can be used upon calculation of air conditioning systems
5) maximal value, measured in center of usable volume
* Light measurement using Quantum sensors with cosinus correction positioned vertically upside.
All specified technical data apply for standard equipment operating at an ambient temperature of + 20 °C (68 °F), with line voltage fluctuations of ± 10%. These typical average values for series equipment were determined in accordance with the BINDER factory standard. They are based upon the recommended distances from the inner chamber walls, namely 10% of height, width and depth. Differing ambient temperatures or variances in the design of individual equipment may produce different performance data.
We therefore recommend that equipment be calibrated and/or validated on a case-by-case basis when working at the extremes of the permissible ambient temperature range.