CO2 incubators from the APT.line™

C 150

The new STANDARD in CO2 incubators

The C 150 sets new standards for cell cultivation: contamination-free as a result of hot air sterilization at 180 °C, stable pH thanks to a drift-free FPI infrared measurement system, superior temperature precision with excellent dynamics, and an interior that is condensation-free even at high humidity levels to ensure best growth results. Used stand alone or as a stack the C 150 is easily accessible.

Modell C 150
  • Temperature range 7 °C (13 °F) above ambient up to 50 °C (122 °F)
  • Air jacket system assuring temperature accuracy and reproducible results
  • Water pan with built-in condensation control maintains dry interior walls up to 95 % RH
  • Standard compliant 180 °C (356 °F) hot air sterilization using an overnight cycletime (acc to. DIN 58947, Pharmacopoeias, ANSI)
  • Weldless deep-drawn inner chamber made of stainless steel with integrated shelf support system
  • Drift free infrared CO2 measurement system
  • Microprocessor with LED display for temperature and CO2-concentration
  • Various alarm and status displays
  • Automatic diagnostic system with optical and audible alarm, as well as zero-voltage relay contact for central monitoring
  • Lockable controller keyboard via 3 digit password
  • Independent adjustable temperature safety device class 3.1, providing full protection against chamber over-temperature, with visual and audible temperature alarm
  • Gas mixing head
  • Tightly-fitting inner glass door
  • Lockable door
  • Door hinged left or right, factory fitted
  • Units are stackable with stacking adapter
  • 3 perforated shelves made of stainless steel
  • BINDER test certificate
  • Two BINDER CO2-incubators, stacked:
  • Two incubators C 150 series, incl. stacking adapter, ensuring direct thermal decoupled stacking
 
  C 150
Exterior dimensions
Width (mm/inch) 680 / 26.8
Height (mm/inch) 819 / 32.2
Depth (mm/inch) 815 / 32.1
plus door handle, I-triangle (mm/inch) 54 / 2.1
plus mains- and gas connection (mm/inch) 60 / 2.4
Wall clearance rear (mm/inch) 100 / 3.9
Wall clearance side (mm/inch) 50 / 2.0
 
Interior dimensions
Width (mm/inch) 500 / 19.7
Height (mm/inch) 600 / 23.6
Depth (mm/inch) 500 / 19.7
Interior volume (l/cu.ft.) 150 / 5.4
Perforated shelves, stainless steel (number standard/max.) 3 / 6
Dimensions of perforated shelves, width x depth (mm) 491 x 442
width x depth (inch) 19.3 x 17.4
Weight of the unit (kg/lbs.) 95 / 209
 
Temperature / CO2 data
Temperature range, 7 °C (44,6 °F) above ambient up to °C / °F 50
Temperature variation at 37 °C (98.6 °F) (± °C) 0.4
Temperature fluctiation (± °C) 0.1
Recovery time after door was opened for 30 sec. at 37°C (98.6°F) (Min) 1) 5
 
CO2 data
CO2-range (Vol.-% CO2) 0 - 20
Setting accuracy (Vol.-% CO2) 0.1
Recovery time after door was opened for 30 sec. up to 5 vol. % (Min) 1) 2) 7
CO2 measurement IR
Hose connectors for CO2 (mm/inch) 6
 
Humidity data
Humidity (average values) (±2 % RH) 95 (±2)
 
Electrical data
Housing protection acc. to EN 60529 IP 20
Nominal voltage (±10 %) 50 / 60 Hz (V) 230
Nominal power (kW) 1.4
Energy consumption 2) at 37°C (W) 110

1) up to 98 % of the set value
2) The recovery times of the gas concentration inside the chamber following a door opening refer to an inlet pressure of 2.0 bars (psi).

All technical data is specified for units with standard equipment at an ambient temperature of  25 °C (77 °F) and a mains voltage fluctuation of +5%/-10%. The temperature data is determined in accordance to factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. Technical data refers to 100 % fan speed. All indications are average values, typical for units produced in series. We reserve the right to alter technical  specifications at all times.

  • Cleaning kit

    Halogenfree detergents for gentle cleaning of interior chamber without causing corrosion:

    - Mild cleaner for stainless steel to provide long -therm-protection
    - Clinically approved disinfectant
    - Cleaning cloth

  • Calibration certificate for temperature and CO2

    Factory calibration certificate for temperature and CO2. Temperature measurement in center / CO2 measurement performed using analyzed test gas at 37 °C (98.6 °F) and 5 % CO2.

C 150
Perforated shelf, stainless steel O
Base on castors O
Stacking adapter for direct thermal decoupled stacking of a CB / C 150 combination O
Stacking adapter for direct thermal decoupled stacking of two C 150 models O
Factory calibration certificate for temperature and CO2. Temperature measurement in center / CO2 measurement performed using analyzed test gas at 37 °C (98.6 °F) and 5 % CO2. O
Gas Supply Control Unit, for the connection of 2 gastank, either CO2 or N2, with acoustic and visible alarm. (Available 2. quarter 2009) O
Temperature measurement acc. to DIN 12880 (27 measuring points) at 37 °C (98.6 °F) or at specified temperature with measuring protocol and certificate O
Gastank connection kit for CO2, consisting of a gastank pressure regulator, max. pressure 10 bar, with connection parts, and a 5 m (16.4 ft.) hose O
Manual for Primary Human Cell Culture, in English O
Cleaning kit for equipment maintenance and disinfection, consisting of pH - neutral detergent (1000 g / 2.21 lbs., concentrate), spray disinfectant (500 ml / 0.13 gal. ready - for - use solution) and lint - free cleansing tissues O