The BINDER M series
Material test chambers with individual programming
With a maximum temperature of 300 °C and comprehensive programming options, the M series from BINDER is ideally suited for material and aging testing. The high-performance fan ensures fast heating.
Advantages:
- The specialist for demanding heating profiles
- Adjustable high air exchange rate
- "Made in Germany" quality
Areas of application:
-
Buildings Materials Industry -
Metal Industry / Engineering -
Surface Technology
| Features | Customer benefits | Characteristics |
|---|---|---|
| APT.line™ temperature technology |
|
APT.line for maximum precision ™
|
| Performance characteristics |
|
|
| Inner chamber concept |
|
|
| Controller |
|
|
| Quality |
|
|
| Accessories and Services |
|
Comprehensive product portfolio
|
- Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results
- Temperature range from 5 °C above ambient temperature to 300 °C
- MCS controller with 25 storable programs of 100 sections each for a maximum of 500 program segments
-
User-friendly LCD screen
- Easy-to-read menu guide
- Integrated electronic chart recorder
- Variety of options for the graphic display of process parameters
- Real-time clock
- Adjustable ramp function via program editor
- Program-controlled ventilation flap
- High air-exchange rate through high-performance fan
- Adjustable fan speed
- Exhaust duct Ø 50 mm
- Independent adjustable temperature safety device class 2 (DIN 12880), with visual temperature alarm
- RS 422 interface for communication software APT-COM™ DataControlSystem
- Units up to 115 liters are stackable
- 2 chrome-plated racks included
- BINDER test certificate
| M 53 | ||
|---|---|---|
|
|
Exterior dimensions | |
| Width (mm) | 634 | |
| Height (incl. feet) (mm) | 779 | |
| Depth (mm) | 575 | |
| Plus door handle and connection (mm) | 150 | |
| Wall clearance, rear (mm) | 100 | |
| Wall clearance, side (mm) | 160 | |
| Exhaust duct (outer Ø mm) | 52 | |
| Steam space volume (l) | 77 | |
| Number of doors (ea.) | 1 | |
|
|
Interior dimensions | |
| Width (mm) | 400 | |
| Height (mm) | 400 | |
| Depth (mm) | 330 | |
| Interior volume (l) | 53 | |
| Racks (number standard/max.) | 2 / 5 | |
| Load per rack (kg) | 15 | |
| Permitted total load (kg) | 40 | |
| Weight (empty) (kg) | 61 | |
|
|
Temperature data | |
| Temperature range approx. 5 °C above ambient temperature to (°C) | 300 | |
| Temperature variation | ||
| at 70 °C (± K) | 0,5 | |
| at 150 °C (± K) | 1,3 | |
| at 300 °C (± K) | 2,8 | |
| Temperature fluctuation (± K) | 0,3 | |
| Warm-up time 1) | ||
| to 70 °C (min.) | 5 | |
| to 150 °C (min.) | 15 | |
| to 250 °C (min.) | 35 | |
| Recovery time after doors were open for 30 sec. 1) | ||
| at 70 °C (min.) | 1 | |
| at 150 °C (min.) | 3 | |
| at 300 °C (min.) | 5 | |
|
|
Electrical data | |
| IP protection class acc. to EN 60529 | IP 20 | |
| Voltage (± 10%) 50 / 60 Hz (V) | 230 / 1N~ | |
| Nominal power (kW) | 1,2 | |
| at 70 °C (W) | 145 | |
| at 150 °C (W) | 300 | |
| at 300 °C (W) | 720 |
1) To 98% of the set value
All technical data are specified for units with standard equipment at an ambient temperature of 25 °C and a line voltage fluctuation of ±10%. The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. All data was determined at 100% of fan speed. All figures are typical average values for series devices. We reserve the right to alter technical specifications at any time.
Access port
With silicone plugs for introducing external measuring instruments into the chamber. Access ports with 10, 30, 50 mm diameters.
Door lock
Prevents unauthorized access to the process sequences in the chamber.
Calibration certificate & validation
BINDER can significantly reduce the workload in qualifying devices. Nobody knows our devices as well and has as much experience in certifications as we do.
| M 53 | |
|---|---|
| Access ports with silicone plug, 10, 30, 50 mm | O |
| Fasteners for additional security for racks (1 set of 4) | O |
| Anti-slip rubber pads for safe stacking (1 set of 4 pieces) | O |
| Lockable controller keyboard | O |
| HEPA fresh-air filter, Class H 14 (according to EN 1822, min. 99.995% for 0.1-0.3 µm particles) | O |
| Measurement of air exchange rate according to ASTM D5374, with definition and protocol according to ambient temperature (measurement in factory) | O |
| Analog output for temperature 4 - 20 mA with 6-pin DIN socket (output not adjustable) | O |
| Additional PT 100 temperature sensor, flexibly installed with external connection, includes LEMO connector (3-pin) | O |
| Additional measuring channel for digital display of specimen temperature with flexible PT 100 temperature sensor, measured data recorded via RS 422 interface | O |
| Temperature measurement acc. to DIN 12880 (27 measuring points) at 150 °C or at specified temperature with measuring protocol and certificate | O |
| Calibration certificate, measurement in center of chamber at 150 °C or at specified testing temperature | O |
| Extension to calibration certificate. Each additional measurement at additional measuring point or testing temperature | O |
| Data Logger Kit T 350: For continuous temperature recording of 0 °C to 350 °C. Kit includes 1 data logger, PT 100 sensor with 2 m Teflon extension cable and 1 fixture for mounting to the BINDER unit | O |
| Data Logger Software: Configuration and evaluation software for all BINDER Data Logger Kits, incl. data cable | O |
| Rack, chrome-plated | O |
| Rack, stainless steel | O |
| Perforated shelf, stainless steel | O |
| Door lock | O |
| Door gasket, FKM (Viton) | O |
| Door with window 230 x 230 mm and interior lighting (15 W) | O |
| Inert gas connection (gas inlet and outlet; Ø 10 mm).We also recommend the mostly gas-tight version option | O |
| Mostly gas-tight version | O |