The BINDER incubator - BF series
Incubator with mechanical convection
The BINDER incubator of the BD series with mechanical convection is suitable for all gentle incubation applications, particularly under a full load and large batch throughputs. This incubator is extremely homogeneous and has quick recovery times.
Advantages:
- Safe and reproducible incubation
- Disinfection routine
- Broad range of products and applications
Areas of application:
-
Plastics Industry -
Microbiology
| Features | Customer benefits | Characteristics |
|---|---|---|
| APT.line™ |
|
APT.line™ for maximum precision
|
| Disinfection and cleaning |
|
Disinfection routine at 100 °C
|
| Inner chamber concept |
|
High energy efficiency
|
| Control and Documentation |
|
|
| Accessories and Service |
|
Wide range of products
|
- Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results
- Temperature range from 5 °C above room temperature to 100 °C
- Digital temperature setting with an accuracy of a tenth of a degree
- MS controller with several timer functions
- Controller time functions:
-
0 to 99 h 59 min delayed ON
- 0 to 99 h 59 min delayed OFF
- 0 to 99 h 59 min temperature-dependent
- delayed OFF
- Adjustable fan speed
- Adjustable ventilation by means of front ventilation flap slide and rear exhaust Ø 50 mm
- Inner glass door
- Independent temperature safety device class 3.1 (DIN 12880) with optical alarm
- RS 422 interface for use with APT-COM™ DataControlSystem communication software or switch over to printer output with RS 232 / RS 422 interface converter
- Adjustable intervals for printer
- Units up to 115 liters are stackable
- 2 chrome-plated racks included
- BINDER test confirmation
| BF 720 | ||
|---|---|---|
|
|
Exterior dimensions | |
| Width (mm) | 1235 | |
| Height (incl. casters) (mm) | 1530 | |
| Depth (mm) | 865 | |
| Plus door handle and connection (mm) | 90 | |
| Wall clearance, rear (mm) | 100 | |
| Wall clearance, side (mm) | 160 | |
| Exhaust duct (outer Ø mm) | 52 | |
| Steam space volume (l) | 808 | |
| Number of doors (ea.) | 2 | |
|
|
Interior dimensions | |
| Width (mm) | 1000 | |
| Height (mm) | 1200 | |
| Depth (mm) | 600 | |
| Interior volume (l) | 720 | |
| Racks (number standard/max.) | 2 / 15 | |
| Load per rack (kg) | 45 | |
| Permitted total load (kg) | 120 | |
| Weight (empty) (kg) | 180 | |
|
|
Temperature data | |
| Temperature range approx. 5 °C above ambient temperature to (°C) | 100 | |
| Temperature variation | ||
| at 37 °C (± K) | 0,4 | |
| at 50 °C (± K) | 0,6 | |
| Temperature fluctuation | ||
| at 37 °C (± K) | 0,1 | |
| at 50 °C (± K) | 0,2 | |
| Warm-up time 1) | ||
| to 37 °C (min.) | 21 | |
| to 50 °C (min.) | 24 | |
| Recovery time after doors were open for 30 sec. 1) | ||
| at 37 °C (min.) | 1 | |
| at 50 °C (min.) | 4 | |
| Ventilation at 70 °C (x/h) | 11 | |
|
|
Electrical data | |
| IP protection class acc. to EN 60529 | IP 20 | |
| Voltage (± 10%) 50 / 60 Hz (V) | 230 | |
| Nominal power (kW) | 1,25 | |
| Energy consumption at 37 °C (W) | 80 |
1) To 98% of the set value
All technical data are specified for units with standard equipment at an ambient temperature of 25 °C and a line voltage fluctuation of ±10%. The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. All figures are typical average values for series devices. We reserve the right to alter technical specifications at any time.
BINDER Data Logger Kits
The new BINDER Data Logger Kits – Makes independent recording of temperature data in the BINDER device possible. The tailored product solution contains helpful accessories: from mounting the logger to the BINDER device to cable access assistance to the sensor mount.
Additional PT 100 temperature sensor
Fixed or flexible installation with external connection for accurate temperature measurement within the sample material.
Watertight inner chamber socket
Switched via main switch. For connecting external devices within the chamber.
Door lock
Prevents unauthorized access to the process sequences in the chamber.
Numerous access ports
With silicone plugs for introducing external measuring instruments into the chamber, access ports with 10, 30, 50, 100 mm diameters.
| BF 720 | |
|---|---|
| Access ports with silicone plug, 10, 30, 50, 100 mm | O |
| Switchable audible alarm for over temperature. Adjustable limit on the independent temperature safety device | O |
| Switchable waterproof interior socket 230 V AC (max. 500 W), protection class IP 65, with corresponding plug. Max. permissible operating temperature 50 °C | O |
| Additional PT 100 temperature sensor, flexibly installed with external connection, includes LEMO connector (3-pin) | O |
| Temperature measurement acc. to DIN 12880 (27 measuring points) at 37 °C or at specified temperature with measuring protocol and certificate | O |
| Calibration certificate, measurement in center of chamber at 37 °C or at specified testing temperature | O |
| Extension to calibration certificate. Each additional measurement at additional measuring point or testing temperature | O |
| Anti-slip rubber pads for safe stacking (1 set of 4 pieces) | O |
| Data Logger Kit T 220: For continuous temperature recording of -90 °C to 220 °C. Kit includes 1 data logger, PT 100 sensor with 2 m Teflon extension cable and 1 fixture for mounting to the BINDER unit | O |
| Data Logger Software: Configuration and evaluation software for all BINDER Data Logger Kits, incl. data cable | O |
| Rack, chrome-plated | O |
| Rack, stainless steel | O |
| Perforated shelf, stainless steel | O |
| Door lock | O |