The BINDER E series
Size ranges of E series
Heating oven with mechanical control
With a BINDER E series heating oven, you obtain a reliable and powerful unit in a solid configuration at a very attractive price. This heating oven is preferred for tasks in basic research, as well as in human and veterinary medicine.Advantages:
- Reliable powerful drying oven
- Very attractive price
- Quality "Made in Germany"
Areas of application:
-
Basic Research / Research Institutes -
Human / Veterinary Medicine -
Clinics / University Hospitals
| Features | Customer benefits | Characteristics |
|---|---|---|
| Reliable powerful drying oven |
|
|
- Temperature range 60 °C (140 °F) to 230 °C (446 °F)
- Robust, space-saving small unit with hydraulic-mechanical control and adjustable ventilation
- Hydro-mechanical thermostat
- Timer 0 to 120 min
- Optional independent temperature safety device class 1 (DIN 12880)
- 2 chrome-plated racks included
- Units are stackable
- BINDER test confirmation
| E 28 | ||
|---|---|---|
|
|
Exterior dimensions | |
| Width (mm) | 580 | |
| Height (incl. feet) (mm) | 405 | |
| Depth (mm) | 425 | |
| Plus door handle, I-panel, connection (mm) | 50 | |
| Wall clearance, rear (mm) | 100 | |
| Wall clearance, side (mm) | 160 | |
| Number of doors (ea.) | 1 | |
|
|
Interior dimensions | |
| Width (mm) | 400 | |
| Height (mm) | 280 | |
| Depth (mm) | 250 | |
| Interior volume (l) | 28 | |
| Racks (number standard/max.) | 2 / 4 | |
| Load per rack (kg) | 10 | |
| Permitted total load (kg) | 25 | |
| Weight (empty) (kg) | 22 | |
|
|
Temperature data | |
| Temperature range approx. 5 °C above ambient temperature to (°C) | 60 - 230 | |
| Temperature variation at 70 °C (± K) | 3 | |
| Temperature fluctuation at 70 °C (± K) | 1,5 | |
| Warm-up time 1) | ||
| at 70 °C (min.) | 28 | |
| at 150 °C (min.) | 36 | |
| Recovery time after doors were open for 30 sec. 1) | ||
| at 70 °C (min.) | 11 | |
| at 150 °C (min.) | 19 | |
| Ventilation data | ||
| Ventilation at 70 °C (x/h) | 30 | |
|
|
Electrical data | |
| IP protection class acc. to EN 60529 | IP 20 | |
| Voltage (± 10%) 50 / 60 Hz (V) | 230 | |
| Nominal power (kW) | 0,8 |
1) To 98% of the set value
All technical data are specified for units with standard equipment at an ambient temperature of 25 °C and a line voltage fluctuation of ±10%. The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. All figures are typical average values for series devices.
We reserve the right to alter technical specifications at any time.
| E 28 | |
|---|---|
| Rack, chrome-plated | O |
| Shelf, perforated, stainless steel | O |