Size ranges of KBF series

Constant climate chambers

The BINDER KBF is the specialist for unconditionally reliable stability testing and precise maintenance of constant climatic conditions. From programming to documentation, it meets all critical requirements.


Advantages:

  • Long-term stable test conditions
  • Independent of the water supply
  • Complete solution: Temperature/humidity simulation in one chamber

Areas of application:

  • Cosmetics Industry
    Cosmetics Industry
  • Food / Beverage
    Food / Beverage
  • Pharmaceuticals Industry
    Pharmaceuticals Industry
  • Packaging Industry
    Packaging Industry
Features Customer benefits Characteristics
APT.line™ climate technology
  • Same test conditions throughout the chamber interior
  • Independent of specimen size and quantity

APT.line™

  • Constant and gentle circulation of air through large-surface side walls even under a full load
  • Homogeneous climate conditions throughout test specimens
Water supply
  • Flexible solution independent of water supply and installation site
  • Easy plug-and-play handling

Suitable for any water quality

  • Direct connection to the domestic water system
  • Convenient water treatment with BINDER PURE AQUA SERVICE
  • Sewage pump for discharges up to 1 m in height
Humidification system
  • Very accurately measured values
  • Minimal maintenance requirements
  • FDA-compliant test results
  • Vapor pressure humidification for fast response times
  • Drift-free, capacitive humidity sensor
  • Short recovery time after door opening
Cooling system
  • Reliable test results even under extreme climate and site conditions
  • Stable, long-term tests according to GLP/GMP
  • Large power reserves, e.g. for lighting applications
  • Powerful compressor cooling system for ambient temperatures up to 32 °C
  • No freezing thanks to double evaporator plate
Standard equipment
  • No extra costs
  • Chamber is easily moved on casters
  • Condensation-free glass door for clear viewing and constant climate

Comprehensive standards

  • Ethernet interface
  • Access port diameter 30 mm
  • Casters from 240 l volume
  • Inner glass door with seal and double outer door seal
  • Door heated against condensation
Controller
  • All measured values read at a glance
  • Convenient documentation
  • Easy operation
  • Convenient HMI (Human Machine Interface)

MCS screen controller

  • 25 program memory
  • Digital continuous-line recorder for monitoring limits
  • Alarm function
Qualification
  • FDA- and GLP/GMP-compliant documentation
  • Maximum process stability
  • FDA approved for several years

More than 10 years of experience with:

  • GLP/GMP, IQ, OQ, PQ, DQ
  • FDA-compliant software APT-COM™
  • Years of proven and recognized validation and documentation materials
Accessories and Services
  • Complete solution
  • Everything from one source
  • BINDER Service is always nearby
  • Various options: BINDER Data Logger Kits, GLP compliance, access ports in various sizes and positions, object temperature measurement
  • Worldwide service network
  • Electronically controlled APT.line™ preheating chamber with cooling system assuring temperature accuracy and reproducible results
  • Temperature range 0 °C to 70 °C (without humidity)
  • Temperature range 10 °C to 70 °C (with humidity)
  • Humidity range 10% to 80% RH
  • MCS controller with 25 storable programs of 100 sections each for a maximum of 500 program segments
  • User-friendly LCD screen
    • Easy-to-read menu guide
    • Integrated electronic chart recorder
    • Variety of options for the graphic display of process parameters
    • Real-time clock
  • Electronically controlled humidification and dehumidification system with capacitive humidity sensor
  • Suitable for stability tests according to ICH guideline Q1A (R2)
  • Inner glass door with seal
  • Independent temperature safety device class 3.1 (DIN 12880) with optical and audible temperature alarm
  • Access port with silicone plug Ø 30 mm, left side
  • Complete safety connection kit for water supply and drainage, including water hose, (total length 6 m)
  • Ethernet interface for communication software APT-COM™ DataControlSystem
  • 2 shelves, stainless steel
  • BINDER test certificate
KBF 240 (E5.2)
Exterior dimensions
Width (mm) 925
Height (incl. casters) (mm) 1460
Depth (incl. door handle, I-triangle, connection 50 mm) (mm) 800
Wall clearance, rear (mm) 100
Wall clearance, side (mm) 100
Steam space volume (l) 348
Number of doors (ea.) 1
Inner glass door(s) (ea.) 1
     
Interior dimensions
Width (mm) 650
Height (mm) 785
Depth (mm) 485
Interior volume (l) 247
Racks (number standard/max.) 2 / 9
Load per rack (kg) 30
Permitted total load (kg) 100
Weight (empty) (kg) 184
     
Temperature data (without humidity)
Temperature range (°C) 1) 0 - 70
Temperature variation at 25 °C (± K) 0,2
Temperature variation at 40 °C (± K) 0,3
Temperature fluctuation (± K) 0,1
Max. heat compensation up to 40 °C (W) 300
     
Climate data (with humidity)
Temperature range (°C) 1) 10 - 70
Temperature variation
at 25 °C / 60% RH (± K) 0,3
at 40 °C / 75% RH (± K) 0,3
Temperature fluctuation
at 25 °C / 60% RH (± K) 0,1
at 40 °C / 75% RH (± K) 0,1
Humidity range (% RH) 10 - 80
Humidity fluctuation
at 25 °C / 60% RH (± % RH) 1,5
at 40 °C / 75% RH (± % RH) 1,5
Recovery time after door was opened for 30 sec
at 25 °C / 60% RH (min.) 4
at 40 °C / 75% RH (min.) 5
     
Electrical data
Voltage (± 10%) 50 / 60 Hz (V) 200-240 1N~
Nominal power (kW) 2,1
Energy consumption at 40 °C / 75 % RH (kW) 1) 0,65
Noise level approx. (dB (A)) 52

Temperature-humidity chart

A: Standard Climate range / B: Discontinuous range

Heat compensation

1) These values can be used for dimensioning air condition systems.

All technical data are specified for units with standard equipment at an ambient temperature of 25 °C and a line voltage fluctuation of ±10%. The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. All figures are typical average values for series devices. We reserve the right to alter technical specifications at any time.

BINDER DataLogger Kits

The new BINDER Data Logger Kits – Makes independent recording of temperature and humidity data in the BINDER device possible. The tailored product solution contains helpful accessories: from mounting the logger to the BINDER device to cable access assistance to the sensor mount.

BINDER Pure Aqua Service

This convenient and flexible water treatment system extends maintenance intervals and is easily implemented independent of water quality.
The specific advantage: Point-of-use system with water quality display and replaceable filter cartridge.

External water supply set

External water supply set from fresh and waste water container, wiring and pump

Temperature measurement of the specimen

Additional PT 100 temperature sensor for accurate temperature measurement of the specimen and digital display of measured values. Recording of measured data via Ethernet or RS 422 interface optional.

KBF 240 (E5.2)
Access ports with silicone plug, 30, 50, 100 mm O
Fasteners for additional security for racks (1 set of 4) O
Lockable controller keyboard O
Additional PT 100 temperature sensor, flexibly installed with external connection, includes LEMO connector (3-pin) O
RS 422 interface for communication software APT-COM™ DataControlSystem O
External water supply set including fresh and waste water containers, cabling, and pump O
BINDER PURE AQUA SERVICE containing single-use cartridge, cabling and measuring device O
Calibration certificate for temperature and humidity. Measurement in center of chamber at 25 °C / 60% RH or at specified values O
Extension to calibration certificate for temperature and humidity. Each additional measurement at an additional measuring point or set of values O
Temperature precision measurement according to DIN 12880 and 9-point humidity measurement / factory standard with measurement log and certificate, measured at 25 °C / 60% RH or at specified values O
Data Logger Kit TH 70: For continuous temperature and humidity recording of -40 °C to 70 °C / 0% to 100% RH. Kit includes 1 data logger, 1 attachable combined humidity/temperature sensor, 2 m extension cable and 1 fixture for mounting to the BINDER unit O
Data Logger Kit TH 70/70: With two attachable combined sensors. One for the continuous temperature and humidity recording of -40 °C to 70 °C / 0% to 100 % RH. Second one at the data logger for recording the environmental conditions. Kit includes 1 data logger, 2 attachable combined humidity/temperature sensors, 2 m extension cable and 1 fixture for mounting to the BINDER unit O
Data Logger Kit T 220: For continuous temperature recording of -90 °C to 220 °C. Kit includes 1 data logger, PT 100 sensor with 2 m Teflon extension cable and 1 fixture for mounting to the BINDER unit O
Data Logger Software: Configuration and evaluation software for all BINDER Data Logger Kits, incl. data cable O
Rack, stainless steel O
Reinforced rack, stainless steel, with 1 set of fasteners (4 pieces), max. load 70 kg O
Perforated shelf, stainless steel O
Independent temperature safety device class 3.3 (DIN 12880) with optical alarm O
4 - 20 mA analog outputs for temperature and humidity measurements (e.g. chart recorder connection) with 6-pin DIN socket. Outputs are adjusted automatically when adjusting the controller. O
Zero-voltage relay alarm outputs for temperature (±2 °C) and humidity (±5% RH), accessible via 6-pin DIN socket, with audible signal that can be switched off (maximum power rating 24 V AC/DC, 2.5 A) O
Switchable waterproof interior socket 230 V AC (max. 500 W), protection class IP 65, with corresponding plug. Max. permissible operating temperature 50 °C O
Door lock O
Interior lighting (15 W) O