The BINDER C series
The STANDARD for CO2 incubators
The BINDER C is the fitting partner for routine applications for cell cultivation: contamination-free due to hot-air sterilization at 180 °C, reliable pH-stable thanks to the drift-free FPI infrared measuring system, high temperature precision with excellent dynamic and no risk of condensation even in high humidity.
Advantages:
- Hot air sterilization at 180 °C
- Seamless, deep-drawn inner chamber made of stainless steel
- Unique BINDER technology (patented air jacket system, etc.)
Areas of application:
-
Biotechnology -
Basic Research / Research Institutes -
Clinics / University Hospitals
| Features | Customer benefits | Characteristics |
|---|---|---|
| Sterilization |
|
180 °C hot air sterilization
|
| APT.line™ heating technology |
|
Preheating chamber with VENTAIR™ air jacket
|
| Gas distribution |
|
Gas mixing head with venturi effect
|
| Inner chamber concept |
|
Seamless, deep-drawn inner chamber
|
| CO2 measurement |
|
Single-beam infrared sensor
|
| Operating costs |
|
|
- Air jacket system assuring temperature accuracy and reproducible results
- Temperature range from 7 °C above ambient up to 50 °C
- Water pan with built-in condensation control maintains dry interior walls
- Standard-compliant hot air sterilization at 180 °C (DIN 58947)
- Seamless deep-drawn inner chamber made of stainless steel with integrated shelf support system
- Drift-free infrared CO2 measurement system
- Microprocessor with LED display for temperature and CO2 concentration
- Various alarm and status displays
- Automatic diagnostic system with visual and audible alarm, as well as zero-voltage relay contact for central monitoring
- Lockable controller keyboard via 3-digit password
- Independent temperature safety device class 3.1 (DIN 12880) with optical and audible temperature alarm
- Gas mixing head
- Tightly-fitted inner glass door
- Door lock
- Door hinged right or left
- Units are stackable with stacking adapter
- 3 perforated shelves, stainless steel
- BINDER test certificate
- 2 CO2 incubators, stacked
- 2 incubators C 150 series, incl. stacking adapter, ensuring direct thermal decoupled stacking
| C 150 | ||
|---|---|---|
|
|
Exterior dimensions | |
| Width (mm) | 680 | |
| Height (incl. feet) (mm) | 819 | |
| Depth (mm) | 815 | |
| Plus door handle, I-triangle (mm) | 54 | |
| Plus connections (mm) | 60 | |
| Wall clearance, rear (mm) | 100 | |
| Wall clearance, side (mm) | 50 | |
|
|
Interior dimensions | |
| Width (mm) | 500 | |
| Height (mm) | 600 | |
| Depth (mm) | 500 | |
| Interior volume (l) | 150 | |
| Stainless steel shelf (number standard/max.) | 3 / 6 | |
| Dimensions of perforated shelves, width x depth (mm) | 491 x 442 | |
| Dimensions (external dimensions) shelves width x depth (mm) | 495 x 444 | |
| Weight (kg) | 95 | |
|
|
Temperature data | |
| Temperature range from 7 °C above ambient temperature to (°C) | 50 | |
| Temperature variation at 37 °C (± K) | 0,4 | |
| Temperature fluctuation (± K) | 0,1 | |
| Recovery time after door was opened for 30 sec. at 37 °C (min) 1) | 5 | |
|
|
CO2 data | |
| CO2 range (Vol.-% CO2) | 0-20 | |
| Setting accuracy (Vol.-% CO2) | 0,1 | |
| Recovery time after door was opened for 30 sec. 5 vol. % (min.) 1) 2) | 7 | |
| CO2 measurement | IR | |
| Connection hose nozzle for CO2 DN 6 for hose with internal diameter (mm) | 6 | |
|
|
Humidity data | |
| Humidity (avg. value) (±2 % RH) | ≥ 95 | |
|
|
Electrical data | |
| IP protection class acc. to EN 60529 | IP 20 | |
| Voltage (± 10%) 50/60 Hz (V) | 230 | |
| Nominal power (kW) | 1,4 | |
| Energy consumption 2) at 37 °C (W) | 110 |
1) To 98% of the set value
2) The specified recovery times of the gas concentration in the inner chamber after the door was opened is based on a connection pressure of 2.0 bar.
All technical data are specified for units with standard equipment at an ambient temperature of 25 °C and a line voltage fluctuation of ±10%. The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. All figures are typical average values for series devices. We reserve the right to alter technical specifications at any time.
Cleaning kit
Special cleaning set:
- Gentle stainless steel cleaner with long-term protection
- Clinically approved disinfectant
- Cleaning cloths
Calibration certificate for temperature and CO2
Calibration certificate for temperature and CO2 (temperature measurement in center / CO2 measurement performed using analyzed test gas at 37 °C and 5% CO2).
| C 150 | |
|---|---|
| Silicone access port with two silicone plugs, 30 mm, left side | O |
| Silicone access port with two silicone plugs, 30 mm, right side | O |
| Perforated shelf, stainless steel | O |
| Base on casters | O |
| Stacking adapter C 150 / CB 150, for direct thermal decoupled stacking of two CB 150 / C 150 unit combinations | O |
| Stacking adapter for direct thermal decoupled stacking of two C 150 CO2 incubators | O |
| Calibration certificate for temperature and CO2 (temperature measurement in center / CO2 measurement performed using analyzed test gas at 37 °C and 5% CO2) | O |
| Temperature measurement acc. to DIN 12880 (27 measuring points) at 37 °C or at specified temperature with measuring protocol and certificate | O |
| BINDER Gas Supply Service external gas tank replacement for connecting 2 gas tanks, either CO2 or N2, with audible and visual alarms, as well as potential-free alarm output | O |
| Gas tank connection set for CO2 consisting of a gas tank pressure regulator (max. pressure 10 bar) with connection parts and 5 m hose | O |
| 4 - 20 mA analog outputs for temperature and CO2 measurements (e.g. chart recorder connection) with 6-pin DIN socket (output not adjustable) | O |
| Cleaning kit for equipment maintenance and disinfection, consisting of pH-neutral detergent (1000 g concentrate), spray disinfectant (500 ml / ready-to-use solution) and lint-free cleansing tissues | O |
| Manual for Primary Human Cell Culture, in English | O |