The BINDER KBW series
Climate chamber with illumination
The BINDER climate chamber with illumination of the KBW series achieves homogeneous light distribution with its natural illumination. That allows this climate chamber to create constant lighting and temperature conditions.
Advantages:
- Homogeneous light distribution
- Natural growth conditions
- Temperature & light in one unit
Areas of application:
-
Cosmetics Industry -
Plant / Insect Growth -
Packaging Industry
| Features | Customer benefits | Characteristics |
|---|---|---|
| APT.line™ climate technology |
|
APT.line™
|
| Light |
|
|
| Cooling system |
|
|
| Standard equipment |
|
Comprehensive standards
|
| Unit design |
|
|
| Accessories and Services |
|
Comprehensive product portfolio
|
- Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results
- Temperature range 0 °C to 70 °C (without illumination)
- Temperature range 5 °C to 60 °C (with illumination)
- 3 variable positioned illumination cassettes with 5 daylight illumination tubes, each, in steps switchable
- The patented illumination system assures an unique homogeneous light distribution
-
MP controller with 2 programs with 10 sections each, alternatively switchable to 1 program with 20 sections
- Integrated week program timer with real time function
- Digital temperature setting with an accuracy of a tenth of a degree
- Adjustable ramp functions via program editor
- Adjustable fan speed
- Elapsed time indicator
- Access port Ø 30 mm, left side
- Inner glass door
- Independent adjustable temperature safety device class 3.1, providing full protection against chamber over-temperature, with visual and audible temperature alarm
- RS 422 interface for communication software APT-COM™ DataControlSystem
- Adjustable intervals for printer
- 3 stainless steel racks
- BINDER test confirmation
| KBW 720 (E5.1) | ||
|---|---|---|
|
|
Exterior dimensions | |
| Width (mm) | 1255 | |
| Height (incl. casters) (mm) | 1925 | |
| Depth (incl. door handle, I-triangle, connection 80 mm) (mm) | 890 | |
| Wall clearance, rear (mm) | 100 | |
| Wall clearance, side (mm) | 100 | |
| Steam space volume (l) | 918 | |
| Number of doors (ea.) | 2 | |
| Inner glass door(s) (ea.) | 2 | |
|
|
Interior dimensions | |
| Width (mm) | 970 | |
| Height (mm) | 1250 | |
| Depth (mm) | 576 | |
| Interior volume (l) | 698 | |
| Racks (number standard/max.) | 3 / 12 | |
| Load per rack (kg) | 45 | |
| Permitted total load (kg) | 150 | |
| Weight (empty) (kg) | 377 | |
| Variable position illumination cassettes | 3 | |
|
|
Temperature data (without illumination) | |
| Temperature range (°C) | 0 - 70 | |
| Temperature variation (± K) | 0,5 | |
| Temperature fluctuation (± K) | 0,1 | |
| Max. heat compensation up to 40 °C (W) | 750 | |
|
|
Temperature data (with 100 % illumination) | |
| Temperature range (°C) | 5 - 60 | |
| Temperature variation (± K) | 1 | |
| Temperature fluctuation (± K) | 0,1 | |
| Max. heat compensation up to 40 °C (W) | 500 | |
| Illumination data (per illumination cassette) | ||
| Daylight tubes (Lux) / (W/m2) | 11.450 / 33 | |
| Fluora® growth lamps (Lux) / (W/m2) | 10.500 / 31 | |
| Arabidopsis tubes (Lux) / (W/m2) | 14.400 / 38 | |
|
|
Electrical data | |
| IP protection class acc. to EN 60529 | IP 20 | |
| Voltage (± 10 %) 50 / 60 Hz (V) | 200 - 240, 1 N ~ | |
| Nominal power (kW) | 2,7 | |
| Energy consumption 1) | ||
| without illumination at 0 °C (W) | 455 | |
| with illumination at 4 °C (W) | 880 | |
| with illumination at 25 °C (W) | 820 | |
| with illumination at 37 °C (W) | 885 | |
| Noise level (dB (A)) | 53 |
1) These values can be used for dimensioning air condition systems.
All technical data are specified for units with standard equipment at an ambient temperature of 25 °C and a line voltage fluctuation of ±10%. The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. All figures are typical average values for series devices. We reserve the right to alter technical specifications at any time.
Variable illumination
The illumination selection is based on the application, light spectrum and intensity. For example: FLUORA® growth fluorescent tubes may be exchanged for daylight fluorescent tubes
BINDER Data Logger Kits
The new BINDER Data Logger Kits – Makes independent recording of temperature and humidity data in the BINDER device possible. The tailored product solution contains helpful accessories: from mounting the logger to the BINDER device to cable access assistance to the sensor mount
Access port
With silicone plugs for introducing external measuring instruments into the chamber, access ports with 30, 50 or 100 mm diameters
Analog output
Analog output for temperature 4 - 20 mA with 6-pin DIN socket (output not adjustable)
Calibration certificate & validation
BINDER can significantly reduce the workload in qualifying and validating devices. Nobody knows our devices as well and has as much experience in certifications as we do
| KBW 720 (E5.1) | |
|---|---|
| Access ports with silicone plug 30, 50, 100 mm | O |
| Securing elements for additional fastening of racks (1 set of 4 pieces) | O |
| Additional PT 100 temperature sensor, flexibly installed, with external connection, including DIN connector (6-pin) | O |
| Ethernet interface for communication software APTCOM™ DataControlSystem | O |
| Factory calibration certificate. Measurement in center of chamber at 37 °C or at specified testing temperature | O |
| Extension to factory calibration certificate. Each additional measurement at an additional measuring point or temperature | O |
| Temperature measurement acc. to DIN 12880 (27 measuring points) at 37 °C or at specified temperature with measuring protocol and certificate | O |
| Data Logger Kit T 220: For the continuous temperature recording of -90 °C to 220 °C. Kit includes 1 data logger, PT 100 sensor with 2 m Teflon extension cable and 1 fixture for the connection at the BINDER unit | O |
| Data Logger Software: Configuration und evaluation software for all BINDER Data Logger Kits, incl. data cable (RS 232) | O |
| Rack, stainless steel | O |
| Shelf, perforated, stainless steel | O |
| Reinforced rack, stainless steel, with 1 set of securing elements (1 set of 4 pieces), max. load 70 kg | O |
| Vibration compatible shelf / platform (positioned at bottom level) to be mounted inside the chamber for shaker / spinner / roller operation. Other positions available on request | O |
| Temperature safety device, class 3.3 (DIN 12880) with optical alarm | O |
| Analog temperature output, 4-20 mA, with 6-pin DIN socket (output not adjustable) | O |
| Zero-voltage relay outputs accessible via 6-pin DIN socket. Additional module for controlling 2 relay outputs via 2 of the programmable controller’s controller contacts. Outputs can be switched on and off either automatically, or also manually | O |
| Waterproof interior socket 230 V AC (max. 200 W), IP65 protected, with corresponding plug (IP66 protected) Max. allowed operating temperature 50 °C | O |
| FLUORA® growth lamps set as replacement for the standard tubes for 230 V | O |
| Arabidopsis fluorescent tubes set as replacement for the standard tubes for 230 V | O |
| Replacement set of standard daylight illumination tubes for one illumination cassette | O |
| Replacement set of FLUORA® illumination tubes for one illumination cassette | O |
| Replacement set of Arabidopsis illumination tubes for one illumination cassette | O |
| Locking door handle with key | O |